Konigsberg Nine AU Update. John Bolte IV, PhD The Ohio State University Injury Biomechanics Research Laboratory. Scientific Symposium on Injury Biomechanics Research. Konigsberg Bite Plate 9AU System. Laptop - NI Labview. Parallel Data Converter. Battery. Amplifier/ Encoder.
E N D
Konigsberg Nine AU Update John Bolte IV, PhD The Ohio State University Injury Biomechanics Research Laboratory Scientific Symposium on Injury Biomechanics Research
Konigsberg Bite Plate 9AU System Laptop - NI Labview Parallel Data Converter Battery Amplifier/ Encoder National Instruments – Digital I/O Card 9 AU Bite Plate
Calibration block X ^ ^ ^ +0.577 i – 0.577 j + 0.577 k q = 54.7o Z Y
Test Plan • Static test at four positions • -x, y, z, • 0.577 Gs • Shaker table test at two frequencies • (0.577 Gs) • 25 Hz • 1000 Hz • Rotation table test (future work) • Low angular acceleration • High angular acceleration
Static Calibration 5 seconds each position Position #1 -X Position #2 +Y Position #3 +Z Position #4 +0.577 i – 0.577 j + 0.577 k
Static Test Pos #1 Pos #2 Pos #3 Pos #4
Error – Static Test Channel #1 – X 6.8% Channel #2 – Y 9.2% Channel #3 – Z 13.0% Channel #4 – X 5.7% Channel #5 – Y 5.8% Channel #6 – Z 1.1% Channel #7 – X 3.9% Channel #8 – Y 2.2% Channel #9 – Z 10.6% average Gs 0.577 Gs % error =
Error – Shaker Test @25 Hz Correlation Amplitude Factor* – 0.577 0.577 E = % error = Channel #7 – X A = .581 E = 0.6% Channel #8 – Y A = .558 E = 3.4% Channel #9 – Z A = .637 E = 10.6% * Time series analysis
Trade-offs • Small size vs. Accuracy Sampling rate • Wireless vs. Skew interval • Low cost vs. Durability accelerometers
Alternatives? • Threshold data recording • Higher sampling rates • Head data ~ 1,000 Hz • DTS G5 data acquisition • More durable accelerometers
Follow-On Work? • Angular acceleration testing • 9 AU angular acceleration calculation check • PMHS head drops • PMHS “boxing punch” • IRB approval for human boxing use • Human use experiment development • Boxing test