Status of Far Detector Linearity Calibrations

Status of Far Detector Linearity Calibrations. Pat Ward University of Cambridge. History. Original 20-point Gain Curves had too few points in linear region Feb 06: 60-point GC used to make first set of FD linearity calibrations July 06: start of regular 40-point GC

Download Presentation

Status of Far Detector Linearity Calibrations

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Status of Far Detector Linearity Calibrations Pat Ward University of Cambridge C.P. Ward

  2. History • Original 20-point Gain Curves had too few points in linear region • Feb 06: 60-point GC used to make first set of FD linearity calibrations • July 06: start of regular 40-point GC • Second set of calibrations made from July 06 GC (in September 06) • This week: have added calibrations from Aug 06 and Sep 06 GC to offline database C.P. Ward

  3. Constants in Offline Database All have TimeEnd = 2038-01-01 C.P. Ward

  4. Stability:Jul06 – Sep 06 • 88% channels good fits both times and adcmax>12000 • Ratio of corrections at ADC=12000: mean = 1.003 sigma = 0.051 Bay 11 rack 4 tube 2; Hardware change?? C.P. Ward

  5. Have asked Howie to process remaining data – not yet done • Code in cvs: PulserCalibration/PulserGainFit PulserCalibration/macros/write_calpulserfits.C • Write-up: MINOS-doc-1886 • I will have to pass this on to Ryan now – good luck! C.P. Ward