An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Sysmex XE-2100 NRBC Detection

  2. NRBC Counts - Methodology • From semiconductor laser flow cytometry. • STROMATOLYSER-NR (diluent) • STROMATOLYSER-NR dye (stain) • Passed through the laser flow cytometer for analysis. • Counted using side fluorescence and forward light scatter.

  3. Flow Cytometry • Forward light scatter - cell size • Side fluorescence - RNA/DNA information (nucleus)

  4. NRBC Channel Explained

  5. Therefore the NRBC Scattergram…

  6. Resolution of NRBC and WBC • Clear & reliable separation of NRBC and WBC • Wbc take upp more dye than NRBC - so higher fluorescence • Smaller volume of the NRBC nucleus gives reduced forward light scatter.

  7. CORRECTING WBC COUNT

  8. The NRBC Count Algorithm NRBC NRBC% = NRBC# x 100 NRBC# + WBC# (NRBC/ 100 WBC) WBC# NRBC#

  9. Eo Ly Ba NRBC Mo Ne The Corrected WBC Count WBC NRBC

  10. NRBC # Revised Algorithm WBC region in WBC/Baso and Diff scattergram include NRBC if NRBC appear

  11. WBC & DIFF channels: Correction for NRBC’s Correct WBC Minus Correct WBC Differential Minus

  12. NRBC FlagsSwiching Algorithm NRBC-CH is measured “NRBC present” flag is monitored (WBC abnormal flag category) Detection from “NRBC%” NRBC-CH is not measured “NRBC?” flag is monitored (WBC suspect flag category) Detection from DIFF scattergram