An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. An evaluation of the isotope ratio performance of an axial TOF - ICP - mass spectrometerF. Vanhaecke, L. Moens, R. Dams, L. Allen* and S. Georgitis*University of Ghent, Laboratory of Analytical Chemistry, Proeftuinstraat 86, B-9000 Gent, Belgium.* LECO Corporation, 3000 Lakeview Ave., St. Joseph, MI 49085 USA.

  2. 1. Instrumentation LECO Renaissance TOF-ICPMS instrument Detector Vacuum Stages Flight Tube 3 2 1 ICP Torch Ion Mirror Sampler Skimmer Extraction Acceleration

  3. 1. Instrumentation (2) simultaneous extraction of ions from ICP  automatic correction for instabilities in ion source & during ion sampling process on-axis acceleration up to 30000 full spectra / s modified ETP discrete dynode electron multiplier  use of analog detection mode to avoid pulse counting saturation  no correction for detector dead time required Meinhard type C concentric nebulizer & double pass cyclonic spray chamber (PTFE)  free aspiration of sample solution

  4. 2. Data processing Accurate background correction  application of 1 mV offset value

  5. 2. Data processing (2) use of signal intensity at m/z = 210 as internal reference (no ionic species present)  correction for signal drift & instrument instability  improvement of both precision & accuracy Isotope ratio precision expressed as RSD% for n = 10

  6. 3. Influence of acquisition time per replicate & analyte concentration on precision

  7. 3. Influence of acquisition time per replicate & analyte concentration on precision(2)

  8. 3. Influence of acquisition time per replicate & analyte concentration on precision (3) Despite use of analog mode, trends in agreement with Poisson counting statistics  average improvement 50  500 g/L = 3.5  10 = 3.2 (A)  average improvement 1  30 s = 5.2  30 = 5.5 (A)  average improvement 10  30 s = 1.8  3 = 1.7 (B) No further improvement on increasing acquisition time > 30 s !!

  9. 4.Typical isotope ratio precision at 30 s acquisition time / replicate (RSD% for n = 10)

  10. 5. Isotope ratio precision for transient signals (RSD% for n = 5) - FI signals, FWHM = 6 s Isotope ratio precision at 500 µg/L still comparable to the best values ever reported for quadrupole-based ICPMS

  11. 6. Mass discrimination 1% per mass unit at mid-mass (Cd)  comparable to quadrupole-based or sector field instrumentation 7. Accuracy The accuracy attainable was evaluated by comparing Pb isotopic results for a Merck Titrisol ('natural') Pb standard solution obtained by (1) TOF-ICPMS and (2) single collector TIMS. For TOF-ICPMS, 'NIST 981 - Common lead isotopic standard' was used as an external standard.

  12. 7. Accuracy (2) For TOF-ICPMS, the result is the average of 3 subsequent determinations,. Each determination consisted of 5 replicate measurements (30 s) of the 'sample' solution and 5 replicate measurements of the NIST standard solution. For TIMS, the result is the average of 5 determinations. For both techniques, the uncertainties indicated are 95% confidence intervals.

  13. 8. Conclusions At sufficiently high signal intensities and for a moderate acquisition time, axial TOF-ICPMS yields a typical isotope ratio precision  0.05% RSD. This precision can be obtained for many ratios simultaneously. The mass discrimination is comparable to that observed with quadrupole-based and sector field instruments and when appropriately corrected for, accurate results can be obtained.