Assessment of radiotherapy set-up error for limb sarcoma using electronic portal imaging (EPI) Wendy Ella, Eleanor Gill, Anna Cassoni, Beatrice Seddon London Bone and Soft Tissue Tumour Service, University College Hospital, London UK CTOS 13th Annual Meeting, Seattle, USA
Purpose of study • To measure the set up variation in patients with limb sarcoma with reference to the standard departmental EPI protocol • To assess the reproducibility of current departmental limb immobilisation techniques • Calculation of margin to be applied to CTV to produce PTV, on basis of set-up variation
Radiotherapy process Mould Room & planning CT scan Volume definition Plan generation Treatment Treatment verification by EPI
Immobilisation • Immobilises limb in same position every day • Reduces day to day variation in limb position (potential source of error) • Impression of limb with patient in the optimum treatment position: • Baseboard with customised foam moulded under limb • sheet of thermoplastic (Orfit) moulded around limb, fixed onto baseboard
Gross tumour volume (GTV) Volume occupied by the gross visible tumour GTV
Clinical target volume (CTV) Volume surrounding macroscopic tumour to include local sub-clinical microscopic spread CTV
Planning target volume (PTV) Geometric concept of an additional margin to ensure that CTV receives prescribed dose, accounting for variations in set-up that inevitably occur day to day PTV
Treatment verification: EPI • Verification of accuracy of treatment • Take images of treatment portals regularly during treatment • Compare with planning images
Electronic portal imaging • Check isocentres on anterior and lateral portals • Match structures • Asymmetric fields • Include a joint • Match in 3 dimensions (left-right, superior-inferior & anterior-posterior) • Assess variation of EPI from planning images, and whether it is within defined departmental tolerance (≤5 mm)
ENSURE THAT ALL IMAGES INCLUDE A JOINT FOR REFERENCE PURPOSES ISO PVs DAY 1 WEEK 1 REVIEW BEFORE 2nd FRACTION * If unable to resolve consult doctor ? delay or continue ≤ 5mm iso deviation is acceptable ≥ 8mm > 5mm, < 8mm ≤ 5mm PRETREAT PV 2# RADIOGRAPHER APPROVAL DR TO SIGN OFF ON DAY 2 REPEAT IMAGE AND REVIEW BEFORENEXT # Error > 5mm < 8mm If ≤ 5mm If ≤ 5mm > 8mm WEEKLY ISO CHECKS Random Error > 5mm < 8mm PRETREAT PV 3RD # If < 5mm If > 5mm < 8mm systematic variation MOVE ON SET RECORD AND GET DR TO SIGN OFF If < 5mm PV # 4 * CONSULT WITH SIMULATOR, REVIEW SET UP If >5mm < 8mm
Methods • Patients receiving radical dose radiotherapy to a limb for a sarcoma under care of two consultants (AMC, BMS) • 49 patients treated between 15/11/05 – 14/11/06 over 12 months • All EPIs for all patients analysed in 3 dimensions • Each image match recorded greatest variation
Results • 614 EPIs were available for assessment • 24 EPIs could not be matched (for technical reasons) • 589 EPIs were analysed (mean 12 per patient) • Mean additional dose of EPIs: 12mu
Results • For 512 (90.5%) variation of the EPI from DRR was ≤ 5mm (i.e. within departmental tolerance) • 63% of patients had all EPIs ≤ 5mm tolerance • Mean number of EPIs within tolerance per patient: 10.39 (86%)
Cumulative percentage for set up variation 67% of all images were within 3mm
Cumulative percentage for set up variation 90.5% of images within 5mm
Cumulative percentage for set up variation 95.9% of images within 7mm
Error in radiotherapy planning and treatment • Error – any deviation between planned and executed treatment: • Systematic error – influences all fractions in identical way throughout treatment • Random error – influences fractions randomly, e.g. due to patient or organ motion
Quantification of error Mean of means = group systematic error RMS = group random error
CTV to PTV margin calculation • Margin = (2.5 X SD of group systematic error) + (0.7 X SD of random error) • Ensures that 90% of patients receive a minimum cumulative CTV dose of at least 95% of the prescribed dose Van Herk et al, IJROBP, 47: 1121-35, 2000 Van Herk et al, Seminars in Radiation Oncology, 14: 52-64, 2004
Systematic/Random errors Margin = (2.5 X SD of group systematic error) + (0.7 X SD of random error) - Van Herk, 2000, 2004
Conclusion • Current immobilisation techniques result in good reproducibility of set-up • >90% of EPIs were within departmental tolerance of ≤5mm • Current departmental EPI protocol results in acceptably accurate treatment • Calculated margin for CTV → PTV for our department = 5.3mm