An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Phakic posterior chamber IOLs • Cells involved in crystalline lens and capsular bag opacification • Evolution of designs of PPCIOLs • Relevant aspects of fixation and sizing • Sizing issue: New studies • Surgical implantation • Relationship between cataract and myopia • Cataractogenesis • Specimens analyzed in our Center • Mechanisms • Review of the literature • Classification of cataracts after phakic IOL implantation?

  2. Third generation: Staar ICL White-to-white + 0.5 mm (myopic eyes) White-to-white – 0.5 mm (hyperopic eyes)

  3. White-to-white?

  4. Pop M, Payette Y, Mansour M. Predicting sulcus size using ocular measurements. J Cataract Refract Surg 2001; 27:1033-1038. -43 eyes of 24 patients (34 ± 6 years) -Spherical equivalent: -25.50 to –8.75, and +4.00 to +7.50 -Measurements: white-to-white with surgical calipers; sulcus diameter with composites of UBM photographs (50 MHz) -Traditional estimation of sulcus size through limbal measurement is inadequate: limbus size alone cannot predict sulcus size

  5. 2 1 3 Werner L, Apple DJ, et al. Correlation between different measurements within the eye with relevance for phakic intraocular lens implantation. J Cataract Refract Surg 2002 (in preparation) 4

  6. No correlation: White-to-white and angle-to-angle (vertical meridian) N = 10 eyes

  7. No correlation: White-to-white and sulcus-to-sulcus (vertical meridian) N = 10 eyes

  8. No correlation: White-to-white and angle-to-angle (horizontal meridian) N = 12 eyes

  9. No correlation: White-to-white and sulcus-to-sulcus (horizontal meridian) N = 12 eyes

  10. Anterior segment B-scan with a high-frequency digital ultrasound system: “ARTEMIS” Developed at the Cornell University by J. Coleman, R. Silverman, and D.Z. Reinstein Manufactured by Ultralink G. Wilserman and S. Phillips

  11. Werner L, et al. Evaluation of a new ultrasound system for refractive surgery. • Poster presented at the XX Congress of the ESCRS, Nice, France, • September 2002 • 15 cadaver eyes (9 phakic and 6 pseudophakic) 1 2 3 4

  12. Phakic cadaver eye (horizontal meridian)