Isolated Bad-CAC Strip at UCSB

Isolated Bad-CAC Strip at UCSB. For first 57 modules produced at UCSB, there was a slight misunderstanding of bonding rules Bonded isolated Bad CAC channel that were not also marked as a bad IDIEL channel 23 of 57 modules had at least one such channel 42 channels total

Download Presentation

Isolated Bad-CAC Strip at UCSB

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Isolated Bad-CAC Strip at UCSB • For first 57 modules produced at UCSB, there was a slight misunderstanding of bonding rules • Bonded isolated Bad CAC channel that were not also marked as a bad IDIEL channel • 23 of 57 modules had at least one such channel • 42 channels total • Plucking channels would result in a few more Grade B modules • Not a large concern

  2. Testing Results on Isolated Bad CAC Strips • All modules have been fully tested with ARCS and LT system • Both pulsed and continuous LED tests taken • ALL of the 42 strips had regular test results (passed) • No problem seen with these channels CAC@ 100 Hz measured (in pF) at QTC Yes, that’s negative capacitance!!!