An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Top 10 Medical Device Citations Regina A. Barrell Food and Drug Administration Office of Regulatory Affairs Office of Enforcement and Import Operations Division of Enforcement

  2. QSIT Approach to Inspections • Level 1 – Abbreviated – • 2 subsystems; • Corrective and Preventive Actions (CAPA) plus Production and Process Controls (P & PC) or Design Controls.

  3. QSIT Approach to Inspections • Level 2 – Comprehensive – • 4 major subsystems; • Management Controls, Design Controls, CAPA and P & PC.

  4. QSIT Approach to Inspections • Level 3 – Compliance Follow-Up – • Special – For Cause – • Special – Risk Based Work Plan - As directed by inspectional guidance and elements of QSIT.

  5. QSIT Approach to Inspections • Compliance Follow-Up, For Cause and Risk Based Work Plan are dictated by the previous FDA-483 findings and other regulatory information and may differ from the typical QSIT approach.

  6. 2012 • 820.100(a) – Corrective and Preventive Action (Procedures) • 820.198(a) – Complaint procedures • 803.17 – Written MDR procedures • 820.100(b) – Corrective and Preventive Action (Documentation) • 820.50 – Purchasing Controls

  7. 2012 • 820.75(a) – Process validation • 820.90(a) – Control of non-conforming product • 820.30 (i) – Design change procedures • 820.198 (c) – Complaint investigation • 820.181 – Device master record not maintained

  8. Total Turbo Citationsvs.2012 Citations

  9. Total Turbo Citationsvs. 2012 Citations

  10. TOP FIVE FDA 483 CITATIONS FOR MEDICAL DEVICE FIRMS Number One 21 CFR 820.100(a) Procedures for corrective and preventive action have not been [adequately] established.

  11. 21 CFR 820.100(a) For example: • Failure to analyze processes, service records, quality audit reports, complaints, returned product, and other sources of quality data to identify existing and potential causes of nonconforming product;

  12. 21 CFR 820.100(a) contd. • Cause of nonconformities relating to product, processes and the quality system are not investigated; • Actions needed to correct and prevent recurrence of nonconforming product or other quality problems are not identified.

  13. TOP FIVE FDA 483 CITATIONS FOR MEDICAL DEVICE FIRMS Number Two 21 CFR 820.198(a) Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been [adequately] established.

  14. 21 CFR 820.198(a) For example: • Complaints are not processed in a uniform or timely manner; • are not documented; or • have not been evaluated for MDR applicability.

  15. TOP FIVE FDA 483 CITATIONS FOR MEDICAL DEVICE FIRMS Number Three 21 CFR 803.17 Written MDR procedures have not been [developed] [maintained] [implemented].

  16. 21 CFR 803.17 For example: • Your firm’s Complaint Handling procedure states that complaints shall be reviewed for MDR reporting per 21 CFR 803, however there are no additional MDR procedures.

  17. 21 CFR 803.17 contd. For example: • Your procedure does not contain a standardized process for determining when an event meets the criteria for MDR. There are no instructions for how your firm will evaluate information about an event to make MDR reportability determinations in a timely manner.

  18. TOP FIVE FDA 483 CITATIONS FOR MEDICAL DEVICE FIRMS Number Four 21 CFR 820.100(b) Corrective and preventive action activities and/or results have not been [adequately] documented.

  19. TOP FIVE FDA 483 CITATIONS FOR MEDICAL DEVICE FIRMS Number Five 21 CFR 820.50 Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been [adequately] established.

  20. 21 CFR 820.50 For example: • Of 11 supplier records reviewed, 4 were not evaluated or included on your firm’s approved supplier list.

  21. 21 CFR 820.50 contd. For example: • Your supplier, which was used to fill ampoules, was disqualified due to quality issues but was left on your approved suppliers list because the supplier was still approved for other services.

  22. FDA Enforcement StatisticsSummaryFiscal Year 2012

  23. CDRH Enforcement StatisticsSummaryFiscal Year 2012

  24. CDRH Seizures By Fiscal Year

  25. CDRH InjunctionsBy Fiscal Year

  26. CDRH Warning LettersBy Fiscal Year

  27. Total Recall Events - CDRHFiscal Year 2007 - 2012

  28. Total Recall Products - CDRHFiscal Year 2007 - 2012

  29. CDRH Recalls – All ClassesFiscal Year 2007 - 2012

  30. Recalls: Definition of Class I • Class I recall: a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

  31. CDRH Class I Recalls – Fiscal Year 2007 - 2012

  32. Recalls: Definition of Class II • Class II recall: a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

  33. CDRH Class II Recalls – Fiscal Year 2007 - 2012

  34. Recalls: Definition of Class III • Class III recall: a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences.

  35. CDRH Class III Recalls – Fiscal Year 2007 - 2012

  36. Root-Cause CategoriesFY 2010 – FY 2012