Pediatric Ultrasound guided blocks “ block under vision “. Dr. Amr Abdulfatah Sayed (M.D.) Associate prof. of Anesthesia, Chronic Pain Management Ain Shams Univ. , Cairo , EGYPT Oct. 2012. Imagine . NO clicks No pops No paresthesia No trans-arterial . Under vision in real time

cosima
Download Presentation

Pediatric Ultrasound guided blocks “ block under vision “

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Pediatric Ultrasound guided blocks“block under vision “ Dr. Amr Abdulfatah Sayed (M.D.) Associate prof. of Anesthesia, Chronic Pain Management Ain Shams Univ. , Cairo , EGYPT Oct. 2012

  2. Imagine • NO clicks • No pops • No paresthesia • No trans-arterial • Under vision in real time • How L.A. behaves • How catheter lodge • Reinjection with inconsistent block

  3. Basic Principles • Mechanical sound energy. • Sinusoidal. • Pulse longitudinalwavealternating compression (high pressure) & rarefaction (low pressure) • P = pressure • T = wave length • F= frequency • distance one peak to other peak is a wavelength one peak to other peak is a wavelength

  4. GENERATION OF AN ULTRASOUND WAVES • electric field is applied to a piezoelectric crystals. • mechanical distortion of the crystals • sound waves (i.e. mechanical energy)

  5. GENERATION OF AN ULTRASOUND IMAGE

  6. Ultrasound & Tissue Interaction • U/S beam travels through different tissues • Subjected to attenuation(Energy Loss). 1) absorption.2) reflection.3) scattering. • Factors affecting attenuation : • Frequency ( high highatten ) ( low lowatten) • Travel distance • Tissue nature

  7. Fact The Ultrasound Transducer Source of Energy and Image Energy progressively degraded (attenuate)as it enters deeper tissues lateral & axial resolution improved with higher frequency transducers,decrease with increasing tissue depth

  8. Facts • Structures typically seen as hyperechoic or echogenic include bone, tendons, pleura, and nerves below the clavicles. • In contrast, blood, fluids, and nerves above the clavicles are hypoechoic.

  9. Facts • Generally speaking, a high frequency wave is subjected to high attenuation thus limiting tissue penetration • low frequency wave is associated with low tissue attenuation and deep tissue penetration.

  10. GAIN & TIME GAIN COMPENSATION (TGC) • receiver amplification is called the Gain. • Gain increases overall brightness of the entire image, including the background noise. • (TGC) selectively amplify the weak returning (attenuated) signals from deeper structures. TGC Gain

  11. TISSUE ECHOGENICITY • Tissue impedance • resistance of a tissue to US passage • Strong wave reflection = hyperechoic (white) • Weak reflection= hypoechoic (greyish) • No reflection = anechoic. (black)

  12. IsoechoicHypoechoicHyperechoic Anechoic

  13. TISSUE ECHOGENICITY

  14. TISSUE ECHOGENICITY

  15. TISSUE ECHOGENICITY • Bone : High tissue impedance • Strong reflection • ++ hyperechoic lines with a hypoechoic shadow underneath

  16. Neural Tissue Echogenicity fascicular or honeycomb appearance

  17. Reflection is high for air • Air has extremely low acoustic impedance (0.0004) • acoustic coupling medium on the transducer surface to eliminate any air pockets • Otherwise ultrasound waves will be reflected limiting tissue penetration. large dropout artifact.

  18. IMAGE ARTIFACTS

  19. Acoustic Shadow Artifact • deep to hyperechoic bone outline is a beam attenuation • Ultrasound beams subjected to attenuation by bone, penetration is severely impeded.

  20. Probes

  21. Linear array probe • High frequency( > 6MHz) • Superficial structures • Depth max. 6 cm • High clarity • Curved probe • Low frequency (2-5 MHz) • Deep structures > 6cm • Less resolution Hokey stick 25mm

  22. Probe Orientation • Transducer marker

  23. Practical Recommendations • Needle probe orientation • Handling of probe resting hand on pt. body • Non dominant hand • More steep angle of needle = difficult visualization • IN PLANE (IP) • OUT OF PLANE (OOP)

  24. The more perpendicular the needle is to ultrasoundbeam, the stronger (brighter and solid) the needle will appear on the monitor

  25. Needle Selection • Large bore needles (e.g. 17 G) • Better visualization • easier to direct. • Preferred for deep blocks (e.g. infraclavicular block, sciatic ) when needle insertion is steep (> 45 degrees) • Smaller bore needles (e.g. 22 G) • easily visualized for more superficial blocks e.g., the axillary block, when the angle of needle insertion is shallow.

  26. Movement of Probe (ART) • A ( Alignment ) • R ( Rotation) • T ( Tilting)

  27. Proper Body Ergonomics • Proper handling the transducer and the needle, to view the screen, and to position the patient are essential for block success and to avoid operator fatigue and body injury.

  28. Proper Body Ergonomics Improper body position Proper body position

  29. Proper Body Ergonomics Proper operator and screen orientation Improper operator and screen orientation

  30. Proper Body Ergonomics Improper hand and arm positions Proper hand and arm position

  31. Local Anesthetic Injection • Aim to surround the neural structure • Doughnut sign. • Saline of D5% ( if PNS) Prior to L.A • Aspirate 1st

  32. Femoral Nerve • doughnut” sign. F.A F.A Doughnut sign.

  33. Glut. max Sc. N.

  34. Ultrasound With & Against proponents critics • improve success rates • simplify the technical challenges • decrease performance times • reduce complications • lack of controlled trials • High expense • equivocal nature of images • Requires sustained training.

  35. Pediatric USGRA Tips • Adults USGRA can be performed in children • Plan @ pre-op visit • Explain for partents ( legal guardian ) • Ped. USGRA performed under anesthetized • light sedation ( esp. > 8 yrs , diff. airway ,MH). • EMLA 1 hrs in advance • Small muscle bulk = high f probe (13-6 MHz, “hockey-stick,” 26-mm footprint)

  36. Pediatric USGRA Tips • Insulated needle • Epidural 22G for catheter ( long term post op). • In-plane > out off plane • Calculate full dose ( volume 0.3–0.5 mL/kg ) • < 5 yrsBupi. 0.25 %, Ropi. 0.2%, Lido 1% • 5 yrsBupi. 0.375 %, Ropi. 0.5% +epi. (Allison Ross et al. , A & A July 2000 vol. 91 no. 1 16-26)

  37. should we use nerve stimulation in addition to U/S Guidance?

  38. Any Questions ?

  39. Upper Limb Blocks • Interscalene • Supraclavicular • Infraclavicular • Axillary • Median N. • Ulnar N. • Radial N. 0.25 -0.5ml /kg 0.25% bupivacaine , 1% lidocaine , 0.25% ropivacaine

  40. Interscalene • Most cranial approach • Not popular ( Phrenic N.) • Indication : shoulder , upper arm, lateral clavicle. • Scanning : • Medial to lateral survey • Trace Back Method • Injectate “Perfect Block”= 0.15-0.3ml /kg front, in the sheath& behind

  41. Trace Back SCM Sc. m Sc. a medial ColourDoppler : to identify vertebral vs & branches of transverse cervical artery below C6

  42. Supraclavicular (Spinal Of UL)

  43. Supraclavicular (Spinal Of UL) • Clinical tips • The usual volume of L.A. 0.3-0.5ml/kg. • (Corner Pocket ) above the 1st rib, next to subclavian a. to anesthetize the lower trunk. • In plane approach is a must • All U.L • ???? Shoulder surgery • 2/3 L.A. @ Corner

  44. Infraclavicular (Cords) Local anesthetic injected posterior to the   axillary artery resulting in a U shape spread around the artery   is associated with complete blockade of the arm, forearm and hand.

  45. Axillary Approach (terminal branches of the brachial plexus ) • Below elbow surgeries AA MCN Inject : MCN +12 & 6 O’clock Inject : MCN + 10 , 2 , 6 O’clock In Plane

  46. Lower Limb

  47. Femoral C superficial to the Ilio-Psoas Muscle (IPM) Base touching CFA, extending lateral to it.

  48. Femoral Surgery on femur , knee