presented by. Handheld Flow Calibrator. Jonathan Ma, Yung Nguyen, Mario Zuniga . Project Advisor | Craig Corrigan Course Advisor | Jan Kleissl. Progress Report MAE 126B | Project E5 May 31 st , 2013. Project E5 | Handheld Flow Calibrator . Project Objectives & Progress Summary.

crwys
Download Presentation

Handheld Flow Calibrator

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. presented by Handheld Flow Calibrator Jonathan Ma, Yung Nguyen, Mario Zuniga Project Advisor | Craig Corrigan Course Advisor | Jan Kleissl Progress Report MAE 126B | Project E5 May 31st, 2013 Project E5 | Handheld Flow Calibrator

  2. Project Objectives & Progress Summary OBJECTIVES • Develop a cost-effective and highly accurate low flow calibrator that utilizes two separated methods to measure mass flow rate and volumetric flow rate. PROJECT DESCRIPTION “JYMeter” Project E5 | Handheld Flow Calibrator

  3. Project Tasks • Estimate the appropriate size materials • Understanding resolutions and analog signals communication Integrate DP Sensor, Mass flow, Humidity Integrate Temp & Abs. Pressure Sensor Oversampling Programming Conditioning the tubes Finding appropriate Length Put everything into photoboard • 1L/Min Flow • Perform Procedures 1.1 & 1.2 (Basic Calibration) • DP • Mass Flow • Temperature, Humidity • Abs. Pressure Project E5 | Handheld Flow Calibrator

  4. Research| Numerical Analysis VOLUMETRIC FLOW RATE : DIFFERENTIAL PRESSURE METHOD Laminar Flow Element Poiseuille Equation obtain calibration equation from calibration experiment = Source: www.alicat.com Project E5 | Handheld Flow Calibrator

  5. Design | Numerical Analysis Poiseuille Equation = VOLUMETRIC FLOW RATE : DIFFERENTIAL PRESSURE METHOD Low Flow Rate (1 L/min) Calculation – Single Tube Higher Flow Rate (10 L/min) Calculation – LFE Model Source: www.alicat.com Project E5 | Handheld Flow Calibrator

  6. Research |Numerical Analysis MASS FLOW METHOD: CONSTANT TEMPERATURE DIFFERENTIAL • Mass Flow Meter • No calculation • Mainly Calibrations } m* (mass flow rate) Source: www.absoluteastronomy.com Project E5 | Handheld Flow Calibrator

  7. Prototyping| Fabrication Progress From Midterm Report| Initial Concept Design Initial Concept Design Project E5 | Handheld Flow Calibrator

  8. Prototyping| Fabrication Progress Final Protoboard Design Project E5 | Handheld Flow Calibrator

  9. Prototyping| Fabrication Progress A Closer Look Project E5 | Handheld Flow Calibrator

  10. Prototyping | Fabrication Progress SOFTWARE : OVERSAMPLING • Displays: • Average of 50 readings for noise reduction • Average of entire run for data analysis • Using Method of Weighted Averages Project E5 | Handheld Flow Calibrator

  11. Experiment | Calibration Experimental Set-Up Project E5 | Handheld Flow Calibrator

  12. Experiment | Calibration Set-Up Project E5 | Handheld Flow Calibrator

  13. Experiment| Preliminary Tube Set-Up Project E5 | Handheld Flow Calibrator

  14. Experiment | Preliminary Calibrations Volumetric Flow Rate: Calibration For Differential Pressure Sensor Project E5 | Handheld Flow Calibrator Source: I&EC Fundamentals, Vol 6, No.4

  15. Experiment | Preliminary Calibrations Mass Flow Rate: Calibrating Mass Flow Sensor Project E5 | Handheld Flow Calibrator Source: I&EC Fundamentals, Vol 6, No.4

  16. Experiment | Preliminary Calibrations Mass Flow Rate: Calibrating Mass Flow Sensor • Summary of Findings During Calibration • Offset due to the pump, not as efficient. • Temperature of the air changes. Project E5 | Handheld Flow Calibrator Source: I&EC Fundamentals, Vol 6, No.4

  17. Project Management & Work Schedule Project E5 | Handheld Flow Calibrator

  18. Remaining Project Tasks • Estimate the appropriate size materials • Understanding resolutions and analog signals communication Integrate DP Sensor, Mass flow, Humidity Integrate Temp & Abs. Pressure Sensor Oversampling Programming Conditioning the tubes Finding appropriate Length Put everything into photoboard • 1L/Min Flow • Perform Procedures 1.1 & 1.2 (Basic Calibration) • DP • Mass Flow • Temperature, Humidity • Abs. Pressure Project E5 | Handheld Flow Calibrator

  19. THE END. Questions? Assignments Research • Hardware & Components - Jonathan • Mass Flow Methods - Mario • Volumetric Flow Methods – Yung Prototyping • Programming – Mario/Jonathan • Numerical Analysis – Yung • Circuit Set-Up – Mario/Jonathan • Fabrication – Team Calibration – Team • Gilibrator Operation • Data Analysis • Testing & Turning Project E5 | Handheld Flow Calibrator