Stain appeared after EP acid renewal

Stain appeared after EP acid renewal. Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes, irises and equators. Now we are investigating STF EP process using old MHI-#0 cavity (with simple beam pipe).

denton
Download Presentation

Stain appeared after EP acid renewal

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes, irises and equators. Now we are investigating STF EP process using old MHI-#0 cavity (with simple beam pipe). They seems to be Niobium-Oxide. However, the problem is not yet solved, not yet fixed. 07.17.2009 ??? H. Hayano, KEK

  2. Brown spot and traces; MHI-06: spot or traces on BP-#1、#1、#2、#3、#4、#8、#9、#9-BP (red indicate bad) MHI-07:BP-#1、#1、#2、#3、#4、#5、#6、#7、#8、#9、#9-BP Kyoto-camera picture examples using new LED illumination BL#6 1-cell equator, t = 306 deg. Downstream : Outside weld area BL#6 #9-BP, t = 241 deg. -2 BL#6 #9-BP, t = 241 deg. -1

  3. BL#6 9-cell equator, t = 034 deg. Upstream : Outside weld area BL#6 9-cell equator, t = 184 deg. Upstream : Outside weld area BL#6 3-cell equator, t = 291 deg. BL#6 3-4 iris, t = 055 deg. BL#6 8-9 iris, t = 053 deg. BL#6 9 equator, t = 037 deg.

  4. Typical EP parameter OLD_MHI-#0 cavity EP test2009/07/07 EP condition :4th EP after acid renewal removal:20 um cavity surface area: 9000 cm2 acid soaking area:5400 cm2 target average current:270A acid flow:15 litter / min cavity rotation:1 rpm cavity acid temp at start: 20.9℃ acid tank temp: 20.1℃ room temp : 31℃

  5. 2009/5/27, BL#6, 1st EP after acid renew 2009/6/10, BL#7, 2nd EP after acid renew 4div 4div 2009/7/01, BL#8, 3rd EP after acid renew 2div 4div 2009/4/08, BL#5, EP before acid renewal

  6. 2009/3/25, ERL-2cell #1, EP before acid renew 2008/12/10, BL#6, EP before acid renew 2div 2div 4div 4div 2009/7/07, OLD_MHI-#0 5th EP after acid renew 2009/7/07, OLD_MHI-#0 4th EP after acid renew

  7. After EP-20um, 5min rotate, HPR #1 equator, t = 248 deg. After EP-20um, 30min rotate, HPR #1 equator, t = 248 deg. after 4th EP after 5th EP After EP-20um, 5min rotate, HPR #1 equator, t = 304 deg. After EP-20um, 30min rotate, HPR #1 equator, t = 304 deg. after 4th EP after 5th EP

  8. After EP-20um, 5min rotate, HPR #1 equator, t = 349 deg. After EP-20um, 30min rotate, HPR #1 equator, t = 349 deg. after 4th EP after 5th EP stains went away by EP 20um.

  9. after 5th EP After EP-20um, 30min rotate, HPR #1 equator, t = 342 deg. After EP-20um, 30min rotate, HPR #1 equator, t = 328 deg. after 5th EP after 5th EP new stains comes up After EP-20um, 30min rotate, HPR #1 equator, t = 000 deg.

  10. MHI#0 cavity scrape out stain in beam pipe edge sample was glued by silver paste on Si substrate. Then, apply XPS analysis. results; Nb 11 at% O 48 at% C 27 at% Si 14 at% (substrate) No special material, just Niobium-oxide, maybe. scraped sample

  11. AES001 had also similar stains, observed as received. ( Jlab EP )

  12. Other spots that have no wall gradients Other spots that have no wall gradients Appendix back side of EL AES001 #2-3 iris 212° to Equator of #3 cell iris to Equator of #2 cell arc scar? back side of EL The result of SI measurements shows that this spot is flat. This spots looks like a sign of Field Emission or Arcing. Many spots like this were observed. 1mm

  13. AES001 #4-5 iris 233° Appendix iris back side of EL back side of EL bright belt to Equator of #4 cell to Equator of #5 cell All the Iris regions are yellowish. This spot is locally bright. It may be caused by EP? 1mm

  14. AES001 #4-5 iris 136° Appendix iris back side of EL back side of EL to Equator of #4 cell to Equator of #5 cell Dropped coffee smear? 1mm

  15. AES001 #7-8 iris 279° Appendix to Equator of #7 cell to Equator of #8 cell Dropped coffee smear? scar? back side of EL back side of EL iris 1mm