1 / 23

In the name of GOD

In the name of GOD. Chapter 11. Methods for measuring speed, attenuation, Absorption and Scattering. For a better understanding of the ultrasonic images we have to measure different ultrasonic properties of biological tissues. Most important parameters are: Velocity Attenuation Scattering

dgriswold
Download Presentation

In the name of GOD

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. In the name of GOD

  2. Chapter 11 Methods for measuring speed, attenuation, Absorption and Scattering

  3. For a better understanding of the ultrasonic images we have to measure different ultrasonic properties of biological tissues. Most important parameters are: • Velocity • Attenuation • Scattering • Absorption

  4. Velocity • In vitro Methods • Can be measured with continuous or pulse wave • Interferometeric method • Use a continuous wave • Accuracy can be of the order of 0.1% • A standing wave is set up between reflector and transducer • Wavelength can be measured by adjusting the position of the reflector

  5. Pulse echo method • Arrangement as above can be used for pulse excitation • TOF method is used (t=2x/c) • The accuracy depends on the sharpness of the pulse and the attenuation in the sample • Improvement in accuracy can be made using substitution method, in this case we have:

  6. To eliminate tissue thickness below set up can be used. • Tm and Tw are TOF with and without sample in the propagation path • T1 and t2 are TOF for pulse to travel from the transducer to the front and rare faces of the sample • cw and cm are sound velocity in the sample and water

  7. Velocity difference method • An extremely accurate (0.01%) method for fluid • A reference liquid and the sample separated by a thin membrane. • If a carriage move the T and R Transducers by Δx, it can be shown that Δt is given by: • In this equation all parameters are known except cm • CW can also be used. In this case phase change must be measured instead of TOF

  8. In vivo methods • Very few of the in-vitro methods can be applied to in-vivo conditions • One method for in-vivo measurements is registration method (Chen et al, 1987) • If a target D can be found by B-scan at a position, we have AB=c0Δt . Δt is the TOF. • At another position A’ we have A’B’=c0Δt’ • In triangle AA’D we have: • C1 is the true velocity and AB=(c0/c1)AD • We can write: • L can be measured physically and l estimated from the image • The method has several problems: • 1-Assume no refraction at the boundaries • 2-Target can be found

  9. Another approach used by Kondo et al (2000) is with a linear array • If the distance between two elements Δy and sinθ0 is known • T1AR1=t11 and t12=T2BR2 and t21=T2CR2 can be measured, the local sound speed c0 can be defined by ABC as: • It is assumed that θ0 =θ1

  10. Attenuation • In vitro methods • A) Transmission methods • Can be done with narrow or broadband pulse • Both can be done with fixed or variable path • Advantage of fix path is the reduction of the error due to diffraction • Setting for fixed path is the same as for velocity measurement • If attenuation coefficient in reference liquid and sample is α0 and α and the pressure p0 and p after the carriage movement we have: • This is not a suitable method for biological tissue

  11. A better method is substitution method • To reduce phase cancellation a small hydrophone used for receiving signal • If the scope has capability of FFT then: • Rt is the transfer function of the system • T=Transmission coefficient • α0 and α are attenuation coefficient in reference and sample • Dividing the two we have • If T≈1 then • Advantage: not necessary to no Rt • In attenuation measurement thickness measurement is not critical as in velocity measurement

  12. Transient thermoelectric method • As mentioned earlier attenuation consists of absorption+Scattering • In the medical range 2 to 15MHz scattering loss is minimal • More than 99% of absorbed energy change to heat • If we measure the temperature change we can calculate absorption by: • Cp, J and I are the specific at constant pressure of tissue, mechanical heat conversion (4.18)J/cal) and acoustic intensity • denotes the slope of temperature rise as a function of time at time t0

  13. The first phase of the curve can not be used because of the heat exchange • This method to be valid • The thermocouple wire and bead should be very thin (<75um) • The half power beam width grater than 3mm to minimize heat loss • Temperature rise <1 degree • The method is useful for high frequency and tightly focused beam

  14. Another approach is Parkers (1983) for pulse decay technique: • A Gaussian intensity profile beam is assumed. • Beam is scanned across an embedded thermocouple of 51um diameter • The intensity profile is • X0 is a measure of beam width • After Δt and assuming no significant heat conduction, temperature distribution along x is: • From main equation we have: • The temperature decay as a function of time found by Parker is • At x=0 or on beam axis: • Kd is the thermal diffusitivity of the medium surrounding the sensor and is equal to 1.5x10-3 cm2/s • Tm can be estimated from the temperature decay history on the beam axis • Other parameters of Im, Δt Cp and J are known

  15. In vivo methods • Methods can be classified into two categories of: • Amplitude loss methods • Frequency shift methods

  16. Amplitude loss methods • If amplitude loss in terms of thickness is measured the attenuation coefficient can be calculated • In this method diffraction compensation must be done

  17. Frequency shift method • The basics is shown in figure • Two spectral region at different depth is selected and the spectral of each region is obtained. • The two spectral at analog frequencies are divided • Slope of the logarithmic graph is the attenuation coefficient • The method should be repeated several time to obtain meaningful results

  18. Scattering • Scattering can be measured only for two extreme cases of Scatterer size>>wavelength and Scatterer size<<wavelength • If scatterer concentration is n is very small then: 2α=n(σs+σa) • If the scattering is the dominate (σs>>σa) then attenuation is equal to scattering • For the cases when scatterer size<<λ wavelength than from figure we have:

  19. If the transmitted ultrasound is a burst at frequency f, the scattered intensity from a region of tissue can be selected by time gating and we have: • η=nσb andσb is backscattering cross section for very small n ηV can be used for volume scattering For dense scatterer such as biological tissue, the power received by the transducer of aperture size A is (assuming R>>z0) By measuring S, P0, α, and α0 the backscattering coefficient can be calculated. The need for measuring P0 can be eliminated by using a reflected pulse from a reflector

  20. For a nonfocused transducer, we can assume, the transducer in the far field is a point source. • Applying mirror theory, the transducer as a receiver would appear as if it were located at a distance 2R from the source. • The received power from the reflector with aperture A is: • Г is reflectivity of the reflector and is about one • Dividing the above two equations we have: • This equation can be converted to dB as: • It can be seen that the backscattering coefficient can be calculated by measuring the difference in the scattering power and reflected power in dB if other parameters are known • This method not be used for focused beam as it result a large error

  21. The method can be used for wideband pulse using FFT for received pulse • Other methods can also be used such as comparative method

  22. In vivo methods • Gray level of a tissue in ultrasonic B-mode can be used to calculate scattering coefficient • Raw data also can be obtained and used to measure backscattering coefficient

More Related