430 likes | 447 Views
This detailed design review will cover the project background, customer needs, revised engineering specs, risk management, RITDAQ system, analog thermocouples, bill of materials, LabView VI interface, and test plan.
E N D
MSD 1 P12453Detailed Design Review Markus Holzenkamp Robin Leili Cody Anderson Detailed Design Review
Agenda • Introduction (1 min) • Project Background (2 min) • Customer Needs (2 min) • Revised Engineering Specs (5 min) • Risk Management (5 min) • RITDAQ (40 min) • Analog • Thermocouples • Bill of Materials (10 min) • Complete • Cost Overview • LabView VI Interface (10 min) • Test Plan (10 min) • Questions Detailed Design Review
Team Members TEAM: Project Manager: Markus Holzenkamp (ME) Project Engineers: Cody Anderson (ME) Robin Leili-Marrazzo (ME) CUSTOMER: Dr. Jason Kolodziej, ME Department, RIT Dr. Edward Hensel, ME Department, RIT SUPPORT: Project Guide: Bill Nowak (Xerox) Faculty Champion: Dr. Kolodziej (RIT) Sponsors: Scott Delmotte (D-R) James Sorokes (D-R) Britt Dinsdale (D-R) Detailed Design Review
Customer Needs Detailed Design Review
Project Scope • Understand the current state of the compressor • Current sensors, DAQ capabilities • Operation characteristics of reciprocating compressors • Install and run ENVISION Condition monitoring system to be donated by Dresser-Rand • Increase the current DAQ capabilities • End Goal: Design, develop, and install effective health monitoring capabilities Detailed Design Review
Sample Engineering Specs (Updated) Detailed Design Review
Concept Selection Detailed Design Review
Risk Assessment Detailed Design Review
Risk Assessment Detailed Design Review
Channel layout and sensor attachments RITDAQ Detailed Design Review
DAQ Channel Layout (Analog) Detailed Design Review
DAQ Channel Layout (Analog) Detailed Design Review
Flow meter How? • Insert in flex-hose section Why? • Easy to modify layout • Cost efficient Materials Needed: • hose clamps (x2) • Adapter (x2) Detailed Design Review
Tank Pressure Detailed Design Review
Tank Pressure How? • Use Rosemount Alphaline 1151 • Mounted on compressor base • Connected with pressure tubing Why? • Already have sensor • Easy to place T-Piece before analog gage Materials Needed: • T-Piece • Adapter • Thermocouple fitting Detailed Design Review
dP across orifice tank How? • Use Rosemount sensor • Mount in basement with tank Why? • No room constraints • Out of the way Materials Needed: • Hardware • Adapter (x2) • Elbow Detailed Design Review
Photocraft HS20 Encoder How? • Use included mounting bracket to attach to oil pump housing Why? • No modifications to existing structure Materials needed: - 7/16-14 Bolt 3in long - 7/16-14 nuts Detailed Design Review
DAQ Channel Layout (Thermocouples) Detailed Design Review
DAQ Channel Layout (Thermocouples) Detailed Design Review
Inlet Coolant Temperature How? • Attach T-Piece between flex hose and steel flange Why? • No modifications to existing structure • Inlet coolant is at room temperature, heat loss is negligible along steel pipe Materials Needed: • T-Piece • Hose Clamp • Adapter for Thermocouple • Thermocouple fitting Detailed Design Review
Cylinder Coolant Layout Detailed Design Review
Head Outlet Coolant Temperature Detailed Design Review
Head Outlet Coolant Temperature How? • Attach fitting to plugged hole and insert thermocouple directed at top outlet Why? • No modification of existing structure • Measures right at the outlet Materials Needed: • Thermocouple fitting • Adapter Detailed Design Review
Center Outlet Coolant Temperature How? • Attach fitting with thermocouple to plugged hole adjacent to outlet Why? • No modification to existing structure • No losses Materials Needed: • Adapter • Thermocouple fitting Detailed Design Review
Tail Coolant Out Temperature Detailed Design Review
Tail Outlet Coolant Temperature How? • Drill and tap hole in pipe. (4mm wall thickness) Why? • Close to outlet • No mixing with other coolant flows Materials Needed: - Thermocouple fitting Detailed Design Review
Tail Coolant Out Temperature • Test strength of threading schedule 80 pipe. Detailed Design Review
Combined Outlet Coolant Temperature How? • Drill and tap hole in bottom of flow sight Why? • Close after all 3 coolant outlets combine • Sufficient wall thickness to secure enough threads for fitting Materials Needed: - Fitting for Thermocouple Detailed Design Review
Combined Outlet Coolant Temperature Measuring the temperature of the cooling pipes at various locations Detailed Design Review
Coolant Temperature Before and After Chiller How? • Insert T-Piece between flex hose and Chiller inlet Why? • Easy to add piping at this place Materials Needed: • T-Piece (x2) • Adapter (x2) • Thermocouple fitting (x2) Detailed Design Review
Crankcase Oil Temperature How? • Attach fitting with thermocouple in plugged hole in crankcase Why? • No modification of existing structure Materials Needed: • Adapter • Thermocouple fitting Detailed Design Review
Crankshaft Bearing Temperature Detailed Design Review
Crankshaft Bearing Temperature How? • Drill into crankcase until just before bearing journal • Insert thermocouple to measure temperature of crankcase at that point • Assistance from Dresser-Rand Why? • Does not alter bearing function • Will show trends of bearing temperature accurately Materials Needed: - Thermocouple fitting (x2) Detailed Design Review
Bill of Materials Detailed Design Review
Cost Breakdown Detailed Design Review
LabView Interface Goals: Clearly display compressor data Time domain and Frequency domain P-v diagram for forward stroke and backstroke Detailed Design Review
Preliminary Test Plans Detailed Design Review
Sensor Tests Means of testing sensor functionality SigLab Existing USB DAQ RIT DAQ Detailed Design Review
RIT DAQ Cross referencing sensor outputs with existing DAQs and Signal Analyzers Use USB DAQ to validate signals Use SigLab to validate signals Reference expected values to ensure proper sensor readings Detailed Design Review
Envision System Utilize the functionality of the RIT DAQ Use the functioning RIT DAQ to ensure the Envision System is displaying proper outputs RIT DAQ will already be cross referenced with alternative Signal Processors and validated Detailed Design Review
Preliminary MSD II Schedule Detailed Design Review
Questions? Detailed Design Review