1 / 9

Introduction to Solid State NMR (2)

PSE797MR Weiguo Hu October 2013. Introduction to Solid State NMR (2). To remove orientation dependence of chemical shifts: Magic-angle spinning (MAS) Magic angle: To remove the strong 1H-13C dipolar coupling: High power decoupling Signal intensity improvement: Cross polarization (CP).

Download Presentation

Introduction to Solid State NMR (2)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PSE797MR Weiguo Hu October 2013 Introduction to Solid State NMR(2)

  2. To remove orientation dependence of chemical shifts: • Magic-angle spinning (MAS) • Magic angle: • To remove the strong 1H-13C dipolar coupling: • High power decoupling • Signal intensity improvement: • Cross polarization (CP) Techniques that deal with challenges in the solid state

  3. MAS 31P spectrum of DOPC Lipid centerband * * * * * * * Blue: non-spinning Red: MAS at 1090 Hz * : spinning sidebands Sideband position = N * spinning speed (Hz)

  4. Samples spins at high speed (>= 3kHz) during data acquisition to achieve high resolution • Direct Polarization (DP) • 1H to 13C Cross Polarization (CP): • Enhancement of signal 13C Magic Angle Spinning (MAS) NMR DP CP NMR Pulse Sequences

  5. 13C CP/MAS spectrum of a poly(urethane Urea) sample Spin speed = 5000 Hz

  6. Crystalline and amorphous signals are distinguished by different conformation and packing • Orthorhombic PE: 32.8 ppm • Monoclinic: 34.2 ppm • Amorphous: ca. 30 ppm • CP over-represents rigid phase • CP relies on strong dipolar coupling • DP over-represents mobile phase • Mobile phase usually has shorter 13C T1 • Quantitative when T1 relaxation is full CP vs. DP 13C NMR O A M DP CPx10 CP SSNMR of UHMWPE Fibers, Hu & Schmidt-Rohr, Polymer 2000

  7. 29Si in silica and zeolites usually has very long T1 • Why? • DP-MAS NMR often requires recycle delay of > 50s • 1H-29Si CP spectrum over-represents surface Si and underrepresent inside Si • Why? • CP pulse time (contact time) needs optimization • It needs time for 1H signal to transfer to 29Si • However, T1r relaxation happens during contact time • 1H-13C CP faces the same problem 29Si NMR

  8. Magnet safety • Will destroy watches and magnetic memories • Will strongly attract iron and steel objects • May harm embryos and stop heart pacemakers • Packing rotors • Stand on mat while packing; don’t drop rotors on the hard floor • Do not use damaged caps • Spin the rotor • Automatic vs manual mode Experiment setup - 1

  9. Create sample file from an existing standard file – dir; new • Tuning the probe – wobb • What if no dip is visible in window • Check and optimize parameters • Recycle delay • Calibration of 90-deg pulse • CP optimization • Pulsing safety • Never misuse pulse length unit: default for pulses are ms; delays are s Experiment setup - 2