980 likes | 2.66k Views
Corneal topography orbscan. S.A.A MORTAZAVI MD Associate professor of ophthalmology Feiz hospital 1390. ORBSCAN SYSTEM. Use the principle of projection
E N D
Corneal topographyorbscan S.A.A MORTAZAVI MD Associate professor of ophthalmology Feiz hospital 1390
ORBSCAN SYSTEM • Use the principle of projection • Forty scanning slit beams (20 from the left and 20 from the right with up to 240 data points per slit ) to scan the cornea and measure independently the X,Y & Z locations
Orbscan imaging • Forty slit images are acquired in two 0.7 second periods • Each of the 40 slit images triangulates one slice of ocular surface • Distance between data slices average 250 microns
ORBSCAN • Orbscan I only slit scan topography • Orbscan II the placidodisc added in orbscan I
ORBSCAN • The images used to construct the anterior corneal surface,posterior corneal surface,anterior iris and anterior lens surfaces • Data regarding the corneal pachymetry and anterior chamber depth
Elevation • Orbscan measure elevation • Elevation is important the only complete scaler measure of surface shape • Both slope & curvature can be mathematically derived from a single elevation map
BEST FIT SPHERE (BFS) • The computer calculates a hypothetical sphere that matches as close as possible to the actual corneal shape being measured • Compares the real surface to the hypothetical sphere showing areas above the surface of the sphere in warm colours and areas below the surface in cool colours
Topography quad map • The upper left : anterior float • The upper right : posterior float • The lower left : keratometric pattern • The lower right : pachymetry map
NORMAL BAND SCALE • Highlights the abnormal areas in the cornea in orange to red colors • The normal areas are all shown in green • Helpful in generalized screening in preoperative examination
AXIAL MAP • Provides detailed keratometric information across the diameter of the cornea • K readings are between certain values the cornea must be neither too steep nor too flat
AXIAL MAP • To create a good quality corneal flap in LASIK if either extremes (too steep or too flat) is the case, this can lead to surgical flap complications • K readings of more than 48 D are an indication of potential keratoconus
Middle box • Keratometric readings • White to white distance in mm • Angle kappa readings • The thinnest point of cornea • irregularity within the central 3 mm & 5 mm
PACHYMETRY MAP • The orbscan measures thickness from the tear film layer to descemet’s membrane and is thicker than that obtained with ultrasound • Adjustment factor (acoustic factor) ,the default setting is 92% • Provides a reading showing the thinnest point of the cornea that may not necessarily be the central reading
PACHYMETRY MAP • Thinnest point <470 micron • In pathological corneas, thinnest point is often displaced inferotemporal • Difference of >100 microns from the thinnest point to the values at 7mm optical zone
ELEVATION MAP • The green colour is referred as refrence sphere (at sea level ) • The warmer colours are above this level and the cooler colours are below
ANTERIOR ELEVATION MAP • Looking at a proper scale in the cornea , can see height differences • Compare the height of the actual cornea to a best fit sphere
Posterior map • The highest elevation value as a keratoconus indicator or at least as a screen for patients may be at risk of developing keratectasia • 55 D elevation as an absolute cut off
ELEVATION DATA • The difference between the highest and lowest points is a potential keratoconus indicator if over 100 microns (Rousch criteria)
DIAGNOSTIC CRITERIA • Power map changes • Posterior elevation maps • Pachymetry • Composite/integrated topography information
POWER MAP • Mean corneal power >45D • In addition to steep corneal curvatures-the bowtie or broken bowtie appearance indicative of early keratoconus • Central corneal asymetry a change within central 3mm optical zone of the cornea of more than 3D
Irregularity in central cornea • Greater than 1.5 D in 3 mm zone and • greater than 2.0 D in 5 mm zone is considered abnormal and cause for concern
POSTERIOR ELEVATION MAP • Many surgeons think the first sign of keratoconus appears on the posterior surface of the cornea • 3.13% of population screened for laser surgery had posterior ectasia criteria by orbscan , despite having axial topography classified as normal
POSTERIOR ELEVATION MAP • The most common reference surface for viewing elevation maps is the best fit sphere • A best fit sphere (BFS) >55D on the posterior profile , indicative of posterior ectasia
Posterior float difference • Greater than 50 micron generally accepted as abnormal • In corneas thinner than normal over 40 as abnormal
Posterior elevation map • The location of the steepest part of the posterior float should be relatively central , but is a more concern it be located away from the center and in an area of corneal thinning • Posterior float difference;40 to 50 microns seems to be the maximum difference
Correlation of signs of the highest point • Highest point on the posterior elevation coincides with the highest point of anterior elevation , the thinnest on pachymetry and the point of steepest curvature on the power map
Although high posterior elevation and ratio between two elevation maps rarely used as exclusion criteria alone , but by considering these together , more conclusive information can be obtained
Risks of ectasia indices • Number of abnormal maps • Posterior float difference >0.050 • 3mm & 5mm irregularity • Peripheral thickness changes • Astigmatism variance between eyes • Steep k’s –mean power map
Three step rule • One abnormal map ; perform with caution • Two abnormal map ; with concern • Three abnormal map ;contraindicated
Composite/integrated information • Similarly between anterior & posterior profiles a forward bending of areas shown above the BFS and association with the thinnest point on the cornea • Inferotemporal displacement of the highest point