Potential Doses to Individuals Screened with Whole-Body Security Screening Systems: OSL Dosimeter Measurements

Potential Doses to Individuals Screened with Whole-Body Security Screening Systems: OSL Dosimeter Measurements. Craig R. Jones, CHP Fran Szrom, CHP. June 2011. UNCLASSIFIED. Disclaimer.

forrest
Download Presentation

Potential Doses to Individuals Screened with Whole-Body Security Screening Systems: OSL Dosimeter Measurements

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Potential Doses to Individuals Screened with Whole-Body Security Screening Systems: OSL Dosimeter Measurements Craig R. Jones, CHP Fran Szrom, CHP June 2011 UNCLASSIFIED

  2. Disclaimer The appearance of a manufacturer’s equipment does not imply endorsement of the equipment by the Army, DOD, or the Federal Government. UNCLASSIFIED

  3. Introduction to Personnel Security Screening Systems • Individual Screened • General-Use: 0.25 μSv (25 μrem) per screening • Limited-Use: 0.01 mSv (1 mrem) per screening • 0.25 mSv (25 mrem) per year • Operator’s/Bystanders Annual Limit • 1 mSv (100 mrem) per year • Inspection zone – operators and bystanders outside inspection zone during screening UNCLASSIFIED 3

  4. Background • Purpose: Evaluate potential dose to the individual being screened • Field measurements • ANSI N43.17-2009 methodology (Reference Effective Dose) • Energy correction, Dose conversion factors • No estimate of eye or skin doses with survey instruments • Dosimeter study requires • Very large number of screenings • Automated method for initiating screenings • Using DDE as a surrogate for Reference Effective Dose UNCLASSIFIED

  5. Test Setup – Repetitive Screening • System 1: • Robotic solution • 93,105 screenings • ~ 2 weeks • System 2: • Software solution • 8,627 screenings • < 1 week UNCLASSIFIED

  6. Test Setup - Phantoms UNCLASSIFIED

  7. System 1: Deep Dose Equivalent • Maximum DDE per screening: • 4,200 μSv/93,105 screenings = 0.045 μSv (4.5 μrem) • Average DDE per screening: • 3,559 μSv/93,105 screenings = 0.038 μSv (3.8 μrem) • Minimum DDE per screening: • 2,560 μSv/93,105 screenings = 0.027 μSv (2.7 μrem) UNCLASSIFIED

  8. System 1: Lens Dose Equivalent • Maximum LDE per screening: • 8,450 μSv/93,105 screenings = 0.091 μSv (9.1 μrem) • Average LDE per screening: • 6,226 μSv/93,105 screenings = 0.067 μSv (6.7 μrem) • Minimum LDE per screening: • 4,710 μSv/93,105 screenings = 0.051 μSv (5.1 μrem) UNCLASSIFIED

  9. System 1: Shallow Dose Equivalent • Maximum SDE per screening: • 15,050 μSv/93,105 screenings = 0.162 μSv (16.2 μrem) • Average SDE per screening: • 10,552 μSv/93,105 screenings = 0.113 μSv (11.3 μrem) • Minimum SDE per screening: • 7,420 μSv/93,105 screenings = 0.080 μSv (8.0 μrem) UNCLASSIFIED

  10. System 2 UNCLASSIFIED 10

  11. System 2: Deep Dose Equivalent • Maximum DDE per screening: • 540 μSv/8,627 screenings = 0.063 μSv (6.3 μrem) • Average DDE per screening: • 416 μSv/8,627 screenings = 0.048 μSv (4.8 μrem) • Minimum DDE per screening: • 250 μSv/8,627 screenings = 0.029 μSv (2.9 μrem) UNCLASSIFIED

  12. System 2: Lens Dose Equivalent • Maximum LDE per screening: • 650 μSv/8,627 screenings = 0.075 μSv (7.5 μrem) • Average LDE per screening: • 516 μSv/8,627 screenings = 0.060 μSv (6.0 μrem) • Minimum LDE per screening: • 280 μSv/8,627 screenings = 0.033 μSv (3.3 μrem) UNCLASSIFIED

  13. System 2: Shallow Dose Equivalent • Maximum SDE per screening: • 780 μSv/8,627 screenings = 0.090 μSv (9.0 μrem) • Average SDE per screening: • 577 μSv/8,627 screenings = 0.067 μSv (6.7 μrem) • Minimum SDE per screening: • 280 μSv/8,627 screenings = 0.033 μSv (3.3 μrem) UNCLASSIFIED

  14. Summary • Both systems evaluated comply with the ANSI/HPS N43.17-2009 dose per screening limit for general-use • DDE is comparable to the Reference Effective Dose (ANSI/HPS N43.17-2009) • The annual number of screenings per individual is limited by the DDE/Reference Effective Dose, not LDE or SDE. UNCLASSIFIED