1 / 17

Resistive Wall Wakes in the Undulator Beam Pipe – Theory, Measurement

This paper discusses the theory and measurement of resistive wall wakes in the undulator beam pipe, including the effects on energy variation within the beam bunch. The implications of these wakes on beam performance and possible mitigation strategies are also discussed.

gailjohnson
Download Presentation

Resistive Wall Wakes in the Undulator Beam Pipe – Theory, Measurement

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Resistive Wall Wakes in the Undulator Beam Pipe – Theory, Measurement Karl Bane FAC Meeting April 7, 2005 • Work done with G. Stupakov, Jiufeng Tu • see: • SLAC-PUB-10707/LCLS-TN-04-11 Revised, Oct. 2004 • SLAC-LCLS-TN-05-6, Feb. 2005

  2. Outline • Review ac resistive wall wake calculations given last time • Analyze reflectivity measurements performed by J. Tu at Brookhaven • Discuss implications

  3. Introduction • in the LCLS, the relative energy variation (within the bunch) induced within the undulator must be kept to a few times the Pierce parameter; if it becomes larger, part of the beam will not reach saturation • the largest contributor to energy change in the undulator is the (longitudinal) resistive wall wakefield • earlier calculations included only the so-called dcconductivity of the metal (beam pipe) wall; we showed last time that: • --if the ac conductivity is included the wake effect is larger • --that the anomalous skin effect is negligible, and can be ignored • --that the wake effect can be ameliorated by going to aluminum, and to a flat chamber.

  4. the free-electron model of conductivity is expected to be valid at low frequencies; it has two parameters: dc conductivity , and relaxation time  • at 295K: Cu--= 5.261017/s, = 2.5210-14 s; • Al --= 3.351017/s, = 0.7510-14 s • if it is valid, the wake in a round beam pipe can be approximated: • with the plasma frequency

  5. Point charge wake

  6. charge—1 nC, energy—14 GeV, tube radius—2.5 mm, tube length—130 m induced energy deviation for round chamber induced energy deviation for flat chamber

  7. Table I: Figure of merit, E (minimum total energy variation over 30 m stretch of beam), for different assumptions of beam pipe shape and material. Nominally, vertical aperture is 2a= 5 mm. Note that Cu-dc results are not physically realizable. --figure of merit only gives rough idea of effect in LCLS; better to use analytic model of Z. Huang and G. Stupakov or simulations

  8. FTIR Reflectivity Measurements (J. Tu talk) Kramers-Kronig relation: causality We can measure

  9. Measured conductivity compared with calculation Kramers Kronig analysis of measurements Calculation using Ashcroft-Mermin ,  • -1(0) is dc conductivity • -Kramer’s Kronig analysis of measurements gives dc conductivity factor 1/2.5 of expected value

  10. Behavior of R for free-electron model Frequency k vs. reflectivity R for a metallic conductor, assuming the free electron model (solid line). For this example =1.21016 /s and =5.410-15 s. Analytic guideposts are also given (dashes). --for LCLS bunch interested in : [10,100] m, or k: [0.06,0.6]m-1

  11. Brookhaven reflectivity data for Cu and evaporated Al range of interest Measurement results from Brookhaven: reflectivity R vs. frequency k for copper and evaporated aluminum. --Cu absorption band at 10 m-1 also seen in plot in Ashcroft-Mermin

  12. Al survey of measurements, Shiles, et al, 1980 range of interest 1evk= 5 m-1 --also slope in R(), but only ½ as steep --clear absorption resonance, only hinted at in Brookhaven results

  13. region of interest Al fit Aluminum reflectivity: comparison of measurements (blue) with calculations using nominal ,  (green); and fitted values: 0.63 nominal , 0.78 nominal  (red). The position of k= 1/c for the fit is also shown. --in literature no data below 0.2 m-1

  14. region of interest Cu fit Copper reflectivity: comparison of measurements (blue) with calculations using nominal ,  (green); and fitted values: 0.66 nominal , 0.67 nominal  (red). The position of k= 1/c for the fit is also shown. --in literature no data below 0.3 m-1

  15. For Al, how does the fit taken affect E for the LCLS? Consider (,): nom.– (3.351017/s, 0.7510-14 s) fit– (2.121017/s, 0.5810-14 s) model2– (1.681017/s, 0.4810-14 s)

  16. Conclusion and Discussion --The resistive wall wakefield in the beam pipe of the LCLS undulator will induce a significant energy variation within the bunch; it seems that this will inhibit much of the beam from lasing. --This effect can be ameliorated by going from a pipe made of Cu to one of Al (by ~2), and from one with a round to a flat cross-section (by ~30%). --To really quantify this, one should go to analytical calculation of Z. Huang and G. Stupakov, or simulations (S. Reichle and W. Fawley)

  17. Brookhaven Measurements --fits (/nom, /nom) are approx. (0.65, 0.80) for Al, and (0.65, 0.65) for Cu; fit is reasonably good, but does not cover desired frequency (k) range for Cu (only 40%) --At higher frequencies the measured R varies linearly with frequency, unlike the constant dependence expected by the free-electron model. This point should be understood before giving too much credence to our results --The deviation in R for Cu above 0.3 m-1 is small and may not have much effect. However, it is difficult to know precisely the implication for the wakefield of a copper pipe: reflectivity at normal incidence alone, without a model, is not enough information to make such a calculation. --Measurements of more Al samples are in the works.

More Related