This compact bench-top hematology analyzer offers 18 parameters and intuitive software for efficient blood analysis. It utilizes innovative measurement principles and requires minimal maintenance, making it a reliable and user-friendly option for medical facilities. With multiple sampling modes and smart card compatibility, it ensures accurate and convenient data management. The analyzer meets various directives and standards for medical devices, ensuring quality performance.
E N D
Product specification • Measurement principles • Conclusion
Product specification • Measurement principles • Conclusion • External view • General specification • Reagents • Sampling mode • Control blood • MTBF • Customer maintenance • Performances • Directives and standards
External view Compact bench top system Size: 14 x 12.6 x 16.5 in Weight: 14 kg / 30 lb 42 cm 32 cm 36 cm
General specifications • 8 parameters / 2 histograms • WBC, RBC, Hgb, Hct, MCV, MCH, MCHC, Plt • 18 parameters / 3 histograms • WBC, RBC, Hgb, Hct, MCV, MCH, MCHC, Plt, RDW, MPV, Pct, PDW • Lym, Mon, Gra (% & #)
Reagent pack or bottles Intuitive software Optional smart card reader Minipack or bottles
Onboard reagents • 3 reagents • ABX Diluent • ABX Alphalyse or Lysebio • ABX Cleaner • 2 options • Minipack • Bottles Minipack containing waste and reagents
Open or close tube 2 sampling modes Closed tube Open tube Throughput OT: 60 samples per hour CT: 55 samples per hour
Sampling tubes Up to 13 alphanumeric characters External bar-code reader Sampling holder
Sampling volume • ABX patented the MDSS concept: • Whole blood micro-sampling • 10 µl • Minimum sample volume in tube • 50 µl • No sampling valve • No maintenance • No clogging
Controls and calibrator • Controls • ABX Minotrol 8 (Low, Normal, High) • ABX Minotrol 16 (Low, Normal, High) • Calibrator • ABX Minocal Multi-language package insert CE marking
Environmental protection Fluids Liquid waste 19 ml / sample 0,31 g / sample Solid waste Power Requirements • Power supply From 100 Vac to 240 Vac (50 HZ to 60 Hz) • Power consumption Maximum 150 VA Total number of active days for all active equipment MTBF = ---------------------------------------------------------------- = around 600 days Number of interventions for breakdown
Customer maintenance Automatic shut-down Stepper motors replace the old compressor concept -> No noise, no maintenance A double needle replace the old shear valve concept -> Small sampling volume without maintenance Liquid valve replace the old pinch valve concept -> No clogging, no maintenance
Directive and standards • Directives • 98/79/EC - IVD medical devices • 89/336/EEC-EMC • 72/23/EEC - Low voltage • Standards • IEC 1010-1: 1990 + A1: 1992 + A2: 1995 • EN 61326: 1997 + A1 • EN61000-3-2: 1995 + A1 + A2: 1998 + A14: 2000 • EN61000-3-3: 1995 • CSA - C22 - 2 No. 1010.1 -92 • UL 3101-01:1993
Product specification • Measurement principle • Conclusion • RBC / PLT • WBC • HGB • Flags and alarms • Data management
Method & technologies RBC/PLT/WBC - Impedance method HGB - Spectro-photometry HCT - Measured MCV, MCH, MCHC - Calculated
RBC / PLT count Impedance method Behind the aperture, a Diluent flow avoids the re-circulation of cells. RBC/PLT mobile threshold
RBC / PLT count - Methods & technologies: RBC/PLT distribution RBC: Distribution curves on 256 counting channels from 30fl to 300fl. PLT: Distribution curves on 128 channels from 2fl to a mobile threshold. This threshold moves according to the microcyte population present in the analysis area.
WBC count - Methods & technologies: WBC and differential measuring principles Histograms Distribution curves on 256 counting channels from 30fl to 450fl. Lymphocytes (30fL to 100fL) Monocytes (100fL to 150fL) Granulocytes (150fL to 450fL)
ICSH (International council for Standardization of Hematology) Cyanmethemoglobin is the reference method but contains Potassium cyanide. ABX customer can choose between 2 lyses. _ “Alphalyse” (Cyanmethemoglobin) _ “Lysebio” (cyanide free lyse) Principle: • Lyses all red blood cells • Photometry method
Flags and alarms Normal limits «H» : Result above upper set-up limit «L»:Result under lower set-up limit Default analysis “*” : Reject 3 counts were not correlated “$” : Reject 2 of 3 counts were correlated “!” : Reject HGB blank outside the limit
RBC/PLT Flags Morphology flag • “MIC” • Microcytes • “SCH” • Schizocytes or Platelet clumps • “SCL” • Small cells in the 2 and 3 fl area
WBC Flags Morphology flag • “G1” • Eosinophilia • Myelocytes • Hypo-segmented neutrophils • “G2” • Hyper-segmented neutrophils • Lyse problem • Old blood • “G3” • Metamyelocytes • Large immature cells • “L1” • Plt aggregates • NRBC • Atypical Lymphocytes • “M2” • Lymphoblast • Myelocytes • Abnormal Lymphocytes • Basophilia
Data management RS232 interface Smart card based memory (60 samples per card) • Quality control • With smart card : • Lowest and highest value • Mean • 2SD, CV% • Levey-Jennings charts
Product specification • Measurement principle • Conclusion
Strengths Easy to use Open tube or closed tube Optional reagent pack Sampling volume of 10 µl high reliability without routine maintenance Small reagent consumption Low power consumption Optional smart card reader for data storage and automated QC Alpha-numeric patient identification Automatic probe wipe More than 25 000 ABX MICROS are already installed world-wide