1 / 16

Nuclear Medicine Image (NM) Integration Profile

Nuclear Medicine Image (NM) Integration Profile. Kevin O’Donnell IHE Radiology Technical Committee Member, Toshiba Medical Systems Company. NM – Actors and Transactions. Points of Interest. Appendix E, Volume 1 (Informative) New to Nuclear Medicine?

Download Presentation

Nuclear Medicine Image (NM) Integration Profile

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Nuclear Medicine Image (NM)Integration Profile Kevin O’Donnell IHE Radiology Technical Committee Member, Toshiba Medical Systems Company Participant Workshop, Oct 2004

  2. NM – Actors and Transactions Participant Workshop, Oct 2004

  3. Points of Interest • Appendix E, Volume 1 (Informative) • New to Nuclear Medicine? • Trouble matching NM to Scheduled Workflow? • E.2 NM Workflow – What happens in NM procedures • E.3 NM Worklists – How to schedule/model it • E.4 NM Data – What to expect in the DICOM • E.5 NM Display – How it should be presented • Lots of subtleties Participant Workshop, Oct 2004

  4. Points of Interest • Acquisition Modality • Modality Images Stored [8] • Create NM objects as specified • additional attribute & vector handling • e.g. Cardiac context codes, slice spacing, patient orientation • Study & Series UID creation clarification • e.g. After PPS complete, new images MUST be new series • e.g. Reconstructed images MUST be new series Participant Workshop, Oct 2004

  5. Points of Interest • Image Manager/Archive • Store/Retrieve NM Objects • Store/Retrieve Secondary Capture Objects • Refer to list in the spec. • “Level 2” storage requirement clarification • e.g. keep private tags • Support for Series Description as Query matching/return key Participant Workshop, Oct 2004

  6. Points of Interest • Evidence Creator • Creator Images Stored [18] • Ditto. (See Acquisition Modality) • Result Screen Export Option Participant Workshop, Oct 2004

  7. Result Screen Export Option:Evidence Creators • Render Static and Dynamic Result Screens • Based on the following DICOM Objects: • Secondary Capture • Multi-Frame Grayscale Byte Secondary Capture • Multi-Frame True Color Secondary Capture • Generate appropriate Secondary Capture objects from any Result Screens created. Participant Workshop, Oct 2004

  8. Example Result Screen Participant Workshop, Oct 2004

  9. Points of Interest – Image Display • Image Display • Retrieve Images [16] • Display NM and Secondary Capture Images • Note requirement for Color, Greyscale, Static & Dynamic Secondary Capture • Support specified NM essential display capabilities • Additional attribute handling • e.g. Patient State, Image Orientation, Detector Sequence, View Code Sequence • Review Option Participant Workshop, Oct 2004

  10. Points of Interest – Image Display • Handle multi-frame / multi-vector images • Multiple simultaneous cines • Grayscale and color result screens • Dynamic and static result screens • Original resolution and scaled • Separate adjustment of upper and lower window levels • Apply local color tables to grayscale screens • Add color tables to system • Change intensity of grayscale result screens • (strongly recommended) • Display Series Description tag Participant Workshop, Oct 2004

  11. Points of Interest – Image Display • Scale “intelligently” to avoid presenting postage-stamp size NM images • Just because it was acquired at 64x64 doesn’t mean they want to review it at 64x64 Participant Workshop, Oct 2004

  12. Points of Interest – Image Display • Scale “intelligently” to avoid presenting postage-stamp size NM images • Allow user to select different vectors/framesets of complex multi-frame images • Allow user to separately adjust intensity for different framesets (eg, phases) in the same “image” Participant Workshop, Oct 2004

  13. Review Option <More details> Adds requirements to the Image Display to allow Review of NM Images • Image Displays • Must additionally support MPR Display, more series, selected different series types together, etc. Participant Workshop, Oct 2004

  14. Display Formats • Grid display • One image set • Fit display • Multiple image sets • Comparison display • Multiple image sets Participant Workshop, Oct 2004

  15. More information…. • IHE Web sites: http://www.rsna.org/IHE • Technical Frameworks: • RAD V5.5 • Technical Framework Supplements - Trial Implementation • May 2004: Radiology Participant Workshop, Oct 2004

  16. Questions? Participant Workshop, Oct 2004

More Related