An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Transcutaneous Blood Gas MonitoringA Multi-Disciplinary Clinical Competency New England Medical Center Respiratory Care Programs Rev 1 03-01-01

  2. Transcutaneous Monitoring: Target Population Neonates, infants, and small children • Transcutaneous monitoring measures skin-surface PO2 and PCO2 • Estimates arterial partial pressure of oxygen and carbon dioxide (PaO2 and PaCO2) • Induces hyperperfusion by local heating of the skin • Measures partial pressure of oxygen and carbon dioxide electrochemically

  3. Transcutaneous Monitoring: Target Population Neonates, infants, and small children • Measures partial pressure of oxygen and carbon dioxide electrochemically

  4. Transcutaneous Monitoring: INDICATIONS • The need to monitor the adequacy of arterial oxygenation and/or ventilation • The need to quantitate the response to diagnostic and therapeutic interventions as evidenced by PtcO2 and/or PtcCO2 values

  5. Transcutaneous Monitoring: CONTRAINDICATIONS • Poor skin integrity • Adhesive allergy

  6. TCM Monitoring: HAZARDS/COMPLICATIONS False-negative and false-positive results may lead to inappropriate treatment of the patient • It is appropriate to correlate non-invasive values with arterialized blood values • Tissue injury may occur at the measuring site

  7. TCM : DEVICE LIMITATIONS/VALIDATION OF RESULTS: TCM monitoring is an indirect measurement • Does not reflect oxygen delivery or oxygen content. • Complete assessment of oxygen delivery requires knowledge of hemoglobin, saturation, and cardiac output • PtcCO2 accuracy is proportional to several factors: • Site placement • Probe temperature • Perfusion • Body temperature

  8. TCM : DEVICE LIMITATIONS/VALIDATION OF RESULTS: TCM monitoring is an indirect measurement • The following factors may increase the discrepancy between arterial and transcutaneous values-- • hyperoxemia (PaO2 > 100 torr) • a hypoperfused state (shock, acidosis) • Improper electrode placement or application • Vasoactive drugs • The nature of the patient's skin and subcutaneous tissue (skinfold thickness, edema)

  9. TCM Monitoring: Validation • Arterial blood gas values should be compared to transcutaneous readings taken at the time of arterial sampling in order to validate the transcutaneous values. • Validation should be performed initially and periodically as dictated by the patient's clinical state.

  10. TCM Monitoring: Validation • During validation studies in patients with functional shunts, electrode site and arterial sampling site should be on the same side of the shunt. • When disparity exists possible causes should be explored before results are reported.

  11. TCM Monitoring: Reducing disparity • Monitoring at alternate sites • Recalibration • Re-membrane TCM probe • Increasing TCM probe temperature • (Usually obligates clinician to increase site change frequency) • Substitution of instrument

  12. TCM ASSESSMENT OF NEED: • When direct measurement of arterial blood is not available/accessible • When lab turn-around time is excessive • PtcO2 and/or PtcCO2 measurements may temporarily suffice if the limitations of the data are appreciated

  13. TCM ASSESSMENT OF NEED: • Appropriate for continuous and prolonged monitoring (eg, during mechanical ventilation, CPAP, and supplemental oxygen administration) • PtcO2 diagnostic values: • assessment of functional shunts (eg, persistent pulmonary hypertension of the newborn, PPHN, or persistent fetal circulation • Useful to determine the response to oxygen challenge in the assessment of congenital heart disease

  14. TCM FREQUENCY: • Transcutaneous blood gas monitoring should be continuous for development of trending data. • Spot checks are not appropriate Reference: Adapted from AARC Clinical Practice Guideline Transcutaneous Blood Gas Monitoring for Neonatal & Pediatric Patients Respir Care 1994:39(12):1176-1179)