An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Physiological Control of a Computer Mouse Presented by Danielle Smith, Shawn Purnell, & Chris Hawk

  2. Outline of Presentation • Project Introduction • Background/Theory • Block Diagram • Hardware Circuit Implementation • Software Implementation • Results/Discussion • Future Developments • Conclusion

  3. Introduction • Project Purpose • Acts as a proof of concept project • Simulate low resolution physiological directional control of an automated computer cursor • for paraplegic empowerment • for military strategic targeting systems

  4. Background on Electrodes • Silver/Silver-Chloride Electrodes • Contact/ Capacitive electrodes • Serve as a transducers to change an ionic current into an electronic current • Ag/AgCl Electrodes used for measurements • Electrodes used for EOG • Small with solid gel • Prevent motion artifact and contact with eye • Electrodes used for EMG • Large with liquid gel

  5. EMG and EOG Signals • EMG measurements - lie within muscle movement • Skeletal muscle organized on basis of motor unit • Motor unit - smallest unit that can be activated by volitional effort • Fibers of a given motor unit are interspersed with fibers of other motor units • At high levels of effort and contraction, many motor unit responses are superimposed • Gives Rise to the complicated interference pattern taken as the signal • EOG signal - based on dipole within the eye • There is a steady corneal-retinal potential from the back of the eye to the front of the eye • This steady dipole may be used to measure eye potential by placing surface electrodes around the eyes

  6. Facial Positioning of Electrodes

  7. Block Diagram

  8. Biopotential Differential Amplifier • Three stage gain • High input impedance • Variable resistor to minimize common mode gain • Low gain in dc-coupled stages to minimize offset potentials produced by electrodes • Bandpass filter in third stage

  9. EOG Circuit Design • EOG parameters • frequency range: 0 - 10 Hz • voltage amplitude: 50 - 3500 µVolt • EOG Amplifier theoretical values • Frequency response: .048 - 5.3 Hz • Gain of approximately 21 for dc-coupled stages • Overall gain: 14,000

  10. EOG Hardware Schematic • High Cutoff Frequency: 4.5 Hz • Low Cutoff Frequency: 0.05 Hz • Midband Frequency Range: 0.25 - 1.5 Hz • Midband Gain: 16,090

  11. EMG Circuit Design • EMG parameters • frequency range: dc - 10,000 Hz • voltage amplitude: 0.1 - 5 mVolt • EMG Amplifier theoretical values • Frequency response: 241 Hz - 5.3 kHz • Gain of approximately 21 for dc-coupled stages • EMG signal processed by diode and LPF • Overall gain: 15,000

  12. EMG Hardware Schematic • High Cutoff Frequency: 4.7 kHz • Low Cutoff Frequency: 250 Hz • Midband Frequency Range: 2.1 - 2.5 kHz • Midband Gain: 15,828

  13. Specification Testing • Common Mode Tests • Common mode output voltages less than 10% the signal • Minimize common mode gain • dc-Coupled Stage Gain Tests • dc-coupled stage gain of less than 25 • to prevent saturation of amplifiers

  14. Results and Discussion • Frequency Response Tests • Frequency responses were as predicted theoretically • Common Mode Tests • Common mode output voltage fall within 10% of their signals • dc-Coupled Stage Gain Tests • dc-coupled stage gain for all three amplifiers were approximately 25

  15. Goals of Software • Goals of Software • Acquire continuous analog signals • Create logic to process signals • click for EMG jaw clench • grid for EOG eye movement

  16. Data Acquisition • Scales analog signals into a one-dimensional array • Indexes the data by channel • Separates individual data within specified channel

  17. LabVIEW Logic Diagram Example

  18. Boolean Indicators

  19. Design Complications and Alterations • Complications • Improper installation of LabVIEW • Alterations • EOG amplifier high cutoff frequency reduced from 10 Hz to 4 Hz to minimize high frequency components in the signal • Summer circuits were found to be unnecessary • Comparator implemented into LabVIEW

  20. Future Development • Project acts as proof of concept • Software Development • Create more resolution in software implementation by creating a counter-based system • Apply to actual computer mouse connector • Hardware Development • Replace electrodes with high sensitivity magnetic sensors • Replace bioamplifiers with amplifiers designed for the magnetic sensor outputs • Develop marketable product

  21. Conclusion • Successful Overall Results in obtaining project goals • Project proves that directional control and clicking simulation can be achieved through facial physiological manipulation • Questions ???