390 likes | 504 Views
Patient Interactions. Photoelectric Classic Coherent Scatter Compton Scattering Pair Production Photodisintegration. Interaction in The body begin at the atomic level Atoms Molecules Cells Tissues Organ structures. X-ray photons can change cells.
E N D
Patient Interactions • Photoelectric • Classic Coherent Scatter • Compton Scattering • Pair Production • Photodisintegration
Interaction in The body begin at the atomic level Atoms Molecules Cells Tissues Organ structures
Some radiations are energetic enough to rearrange atoms in materials through which they pass, and can therefore he hazardous to living tissue. 1913
EM Interactions with Matter • General interactions with matter include • scatter (w or w/o partial absorption) • absorption (full attenuation)
No interaction: X-ray passes completely and get to film Complete absorption: no x-rays get to film Partial absorption with scatter Interactions of X-rays with matter
Photoelectric effect • Low energy (low kVp) x-ray photon ejects inner shell electron (energy absorbed) • Leaving an orbital vacancy. As vacancy is filled a photon is produced • More likely to occur in absorbers of high atomic number (eg, bone, positive contrast media) • Contributes significantly to patient dose, • As all the photon energy is absorbed by the patient (and for the latter reason, is responsible for the production of short-scale contrast).
PHOTOELECTRIC ABSORBTION IN THE PATIENT (CASCADE OF ELECTRONS)
PHOTOELECTRIC ABSORBTION IS WHAT GIVES US THE CONTRAST ON THE FILM
8 p+ + 8e- = neutral atom CLASSICAL SCATTER IN PATIENT INCOMING PHOTONS FROM TUBE Pass by the ELECTRONS IN THE PATIENT Do not interact with e– Causes them to VIBRATE – RELEASING SMALL AMOUNTS OF HEAT
Classical (Coherent) Scattering • Excitation of the total complement of atomic electrons occurs as a result of interaction with the incident photon • No ionization takes place • Electrons in shells “vibrate” • Small heat is released • The photon is scattered in different directions • Energies below 10K keV
Compton scatter • High energy (high kVp) x-ray photon ejects an outer shell electron. • Energy is divided between scattered photon and the compton electron (ejected e-) • Scattered photon has sufficient energy to exit body. • Since the scattered photon exits the body, it does not pose a radiation hazard to the patient. • Can increase film fog (reduces contrast) • Radiation hazard to personnel
COMPTON SCATTERING – OUTER SHELL ELECTRON IN BODY – INTERACTS WITH X-RAY PHOTON FROM TUBE
During Fluoro – the patient is the largest scattering object
Differential Absorbtion • Results from the differences between xrays being abosorbed and those transmitted to the image receptor • Compton Scattering • Photoelectric Effect • X-rays transmitted with no interaction
Compton and Differential Absorbtion • Provides no useful info to the image • Produces image fog, a generalized dulling of the image by optical densities not representing diagnostic information • At high energies
Photoelectric and Differential Absorbtion • Provides diagnostic information • X-rays do not reach film because they are absorbed • Low energies (more differential absorbtion) • Gives us the contrast on our image
No interactions with Image Receptor and Differential Absorbtion • No interaction • Usually high kVp • Goes through body • Hits image receptor • Usually represents areas of radiolucency (low atomic numbers) • Results in dark areas on the film
The probability of radiation interaction is a function of tissue electron density, tissue thickness, and X-ray energy (kVp). • Dense material like bone and contrast dye attenuates more X-rays from the beam than less dense material (muscle, fat, air). • The differential rate of attenuation provides the contrast necessary to form an image.
Remember….When reviewing diagrams What is coming in (e or photon? Where is it occurring (the tube or body?) Keep practicing – you will get it