An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Medical Imaging X-Rays I

  2. Principle of X-ray A source of radiation

  3. Principle of X-ray A source of radiation A patient of non uniform substance

  4. Principle of X-ray A source of radiation A shadow A patient of non uniform substance

  5. Principle of X-ray A source of radiation

  6. X-ray tube • Working Principle: Accelerated charge causes EM radiation: • Cathode filament C is electrically heated (VC = ~10V / If = ~4 A) to boil off electrons • Electrons are accelerated toward the anode target (A) by applied high-voltage (Vtube = 40 – 150 kV); • Deceleration of electrons on target creates "Bremsstrahlung" evacuated gas envelope filament VC, If + A - C - +

  7. X-ray tube • Cathode Filament (-) • Coil of tungsten wire • High resistance in coil ->temperature rise to > 2200oC • Thermionic emission of electrons • Tube (vacuum) • Typical: Vtube = 40 – 150 kVp, Itube = 1-1000mA evacuated gas envelope filament VC, If - + - - - - - A - - - - C space charge stops further emission kVp, Itube - +

  8. X-ray tube • Anode • Tungsten (high atomic number Z=74) • Electrons striking the anode generate HEAT and X-Rays • In mammography ->Molybdenium (Z=42) and Rhodium (Z=45) • Stationary anode-> tungsten embedded in copper • Rotating anode (3000 to 10,000rpm) -> increase heat capacity, target area evacuated gas envelope filament VC, If - + - - - - - A - - - - C space charge kVp, Itube - +

  9. XRAY PRODUCTION

  10. X-RAY production • X-ray tube produces two forms of radiation • Bremsstrahlung radiation (white radiation) • Characteristic radiation

  11. White radiation, Bremsstrahlung (Brake) • Inelastic interaction with atoms nuclei • Loss of kinetic energy • Xray (E) = lost kinetic E X-Ray • High kinetic energy • Forward radiation • Emission  Z2 electron Coulombic interaction (Atomic number) # of protons

  12. White radiation, Bremsstrahlung -Smaller L produce larger X-ray -Broad range of emitted wavelengths X-Ray L

  13. How many wavelength will be emitted by a beam of electrons underegoing “Bremsstrahlung ”

  14. White radiation, Bremsstrahlung -Smaller L produce larger X-ray -Broad range of emitted wavelengths X-Ray L impact with nucleus maximum energy

  15. X-ray intensity -QUANTITY • Overall Bremsstrahlung intensity I : • 90% of electrical energy supplied goes to heat, 10% to X-ray production • X-ray production increases with increasing voltage V

  16. Bremsstrahlung spectrum relative output • Theoretically, bremsstrahlung from a thick target creates a continuous spectrum from E = 0 to Emax • Actual spectrum deviates from ideal form due to • Absorption in window / gas envelope material and absorption in anode • Multienergetic electron beam Peak voltage kVp

  17. Characteristic radiation relative output • Energy must be > binding energy • Discrete energy peaks due to electrons transitions • Ka transition L->K • Kb transition M,N,O->K Peak voltage kVp

  18. Characteristic radiation Incident electron

  19. Characteristic radiation l2 Incident electron Occurs only at discrete levels There is a possibility of forming Auger electrons

  20. Characteristic radiation • In Tungsten characteristic X-ray are formed only if V>69.5 kV because K shell binding energy is 69.5 keV • Molybdenum K-shell can be obtained at V> 20kV • L shell radiation is also produced but it’s low energy and often • absorbed by glass enclosure

  21. X-ray intensity -QUALITY • Effective photon energy produced • Effective = ability to penetrate the patient • Effective photon energy ~ 1/3 to ½ of energy produced • Higher energy better penetration • Beam filtration – beam hardening

  22. Beam Hardening Polyenergetic beam ------------------------------->monoenergetic beam

  23. X-ray tube construction

  24. Anode Most of the energy deposited on the anode transfers into heat

  25. Reduction of anode heating • Made of Tungsten, high melting point high atomic number Z = 74 Kinetic energy of incident electrons 100keV electron 6 MeV electron

  26. Anode • the target angle, 7 to 20 (average 12) Seffective = Sactual*sin() -----------> Line focusing principle

  27. Anode filament balance General radiography

  28. Heel effect - SID source to image distance - Heel effect is smaller at smaller SID Reduction of intensity on the anode side SID The reduction in intensity can be used to reduce patient exposure

  29. Beam collimation • Size and shape of the beam • Lead shutters • Dose reduction

  30. Reduction of anode heating • Anode angle of 7º…15º results in apparent or effective spot size Seffectivemuch smaller than the actual focal spot of the electron beam (by factor ~10) • Rotation speed ~ 3000 rpm • Decreases surface area for heat dissipation from by a factor of 18-35.

  31. Limitations of anode angle • Restricting target coverage for given source-to-image distance (SID) • "Heel effect" causes inhomogeneous x-ray exposure

  32. X-ray tube - space charge • Space charge cloud forms at low tube voltage • At low filament current a saturation voltage is achieved, rising tube voltage will not generate higher electron flow • At high filament current and low tube voltage, space charge limits tube current->space-charge limit

  33. Space charge limited • At high filament current and low tube voltage, space charge limits tube current->space-charge limit

  34. Generator • Single phase • Single phase input (220V, 50A) • Single pulse or double pulse->rectifier • Min exposure time 1/120 sec • Xray tube current non linear below 40kV • Three phase • Three phase wave, out of phase 120 deg • More efficient higher voltage • Better control on exposure

  35. Rectifier Protects cathode from anode thermionic emission

  36. Rectifier 1 phase 3 phase

  37. BREAK!

  38. Principle of X-ray A source of radiation A patient of non uniform substance

  39. N = Noe-mL Attenuation N True for monoenergetic x-ray Loss of photons by scattering or absorption L1 L No N m -> linear attenuation coefficient L1

  40. m linear attenuation coeff. • m = mr+ mph+ mc+ mp [cm-1] • rayleigh • photoelectric • Compton • pair

  41. m linear attenuation coeff. • m = mr+ mph+ mc+ mp [cm-1] • depends on tissue • soft tissue, hard tissue, metals • mdecreases when energy increase • soft tissue: • m = 0.35  0.16 cm-1 for E = 30  100keV • mdepends on density of material • mwat > mice> mvapor

  42. Mass attenuation coeff.

  43. Mass attenuation coeff. N = Noe- r(m/r)L rL = mass thickness

  44. I x Mass attenuation coeff. N = Noe- r(m/r)L rL = mass thickness

  45. Poly-energetic beam • Mass attenuation coefficient and linear attenuation coefficient are for mono-energetic beam • Half-value layer is for quantifying poly-energetic beams

  46. HVL half value layer • Thickness of material attenuating the beam of 50% - narrow beam geometry • HVL for soft tissue is 2.5  3.0 cm • at diagnostic energies

  47. HVL half value layer • Transmission of primary beam: • 10% chest radiography • 1% scull radiography • 0.5% abdomen radiography • Mammography (low energy HVL 1 cm)

  48. Mean free path 1/m • Average distance traveled before interaction MFP=1/m HVL mfp

  49. Principle of X-ray A source of radiation A shadow A patient of non uniform substance