Cryogen Free Desktop Vibrating Sample Magnetometer (VSM) to 5 Tesla. Background. Cryogenic Ltd are pioneers of Cryogen Free magnet systems. Prior to this development high magnetic fields were only available to those with access to liquid helium.

kollar
Download Presentation

Cryogen Free Desktop Vibrating Sample Magnetometer (VSM) to 5 Tesla

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Cryogen Free Desktop Vibrating Sample Magnetometer (VSM) to 5 Tesla

  2. Background • Cryogenic Ltd are pioneers of Cryogen Free magnet systems. • Prior to this development high magnetic fields were only available to those with access to liquid helium. • Alternatives such as water cooled electromagnets require very large power units. • The cryogen free technology allows magnets of a wide range of fields and dimensions to be built using the same method of refrigeration.

  3. 5 Tesla Desktop Cryogen Free VSM

  4. Cryogen Free VSM • The Basic system includes: • The superconducting magnet up to 5 Tesla • The VSM probe. • Cryocooler and compressor. • Electronics Rack housing the bipolar magnet power supply, temperature controller, thermometers, scanner and the measurement option electronics. • The computer (Pentium based with high resolution TFT screen), • including all software, I/O, IEEE and other necessary cards for measurements. • LabVIEW operating software.

  5. Desktop Cryogen Free VSM Magnetic field range: up to 5 Tesla Central field homogeneity: 0.1% over samples (0.01% optional) Persistent mode decay rate: 10ppm / hour (optional switch) Field Sweep rate: 0.5T/min Sample temperature range: 1) Room temperature only 2) 77K to 300K 3) up to 1000K with furnace Maximum sample size (in VSM): 8mm diameter VSM vibration amplitude: 0.5, 1, 2mm peak to peak (optically controlled) VSM vibration frequency: 15 to 25 Hz Typical Base noise (with 10 second averaging): 10-6 emu Accuracy and reproducibility: 0.5% Temperature setting period (if applicable): 5 minutes if dT is ~5K Temperature control (if applicable): 100mK @ 100K (with Lakeshore series unit)

  6. Model No mini-CFM-3T-51-H3 mini-CFM-5T-51-H3 mini-CFM-5T-51-H3 Magnetic field 2 Tesla Solenoid 3 Tesla Solenoid 5 Tesla solenoid Field direction to bore Axial Axial Axial Homogeneity in 10mm DSV 0.1% 0.1% 0.1% Persistency (optional switch) 1 x 10-5 per hour 1 x 10-5 per hour 1 x 10-5 per hour Room temperature bore 51mm bore 51mm bore 51mm bore Cooldown 300K to operating temperature 14 hours 15 hours 18 hours Ramp time 0 Tesla to full field 5 minutes 6 minutes 10 minutes Operating current <60 Amps <60 Amps 60 Amps Desktop Cryogen Free VSMModel Range

  7. The Advantages of • Cryogen Free Facility • Low Operating Costs as liquid helium is not required. • Ease of use as there are no cryogen transfers to make. • Independent and convenient operation • Fast Ramping capability • Automated operation • May be used in any orientation.

  8. Automation • The system is installed with the National Instruments LabVIEW operating software and a standard Cryogenic applications package, which allows the complete system to be run under computer control. • This allows:  • Control of magnetic field • Fail-safe shutdown even with power failure. • Better data presentation, real data and then curve fitted data. • Full development software licence allowing the user to develop special applications

  9. Optional ExtrasOption 2: High temperature oven insert To perform measurements at above room temperature, an oven insert is offered which increases the standard temperature range. In normal operation, the system has a continuous temperature range of 77K to 300K. With the oven insert this can complemented by a temperature range of 300K to 1000K.

  10. Other Applications Of Cryogen free magnets. • Magneto-optical effects, EPR, ESR and NMR • Industrial applications eg; process of industrial minerals including China Clay and Calcium Carbonate. • Synchrotron and Gyrotron magnets to generate X-rays and tuneable microwave. • Medical Applications • Magnetic Resonance Imaging