An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Practical Fundamentals of Biomedical Engineering Sensors

  2. Introduction • Importance: • Sensors are the main building blocks of diagnostic medical instrumentation • Research Motivation: • Lower healthcare costs • Optimising efficiency • Patient self-testing • Instantaneous diagnosis at point of care

  3. Sensors are used to measure • Electrolytes, enzymes and others in blood • Pressure, flow and gas concentration • Typically gases like O2 and CO2 • First design step • Asses in vitro: • Accuracy • Operating Range • Response Time • Sensitivity • Resolution • Reproducibility • In vivo testing

  4. Sensor Classification • Physical • Electrical • Chemical • Biosensors • Biological recognition element • Supporting structure to convert biochemical reaction to recognizable signal

  5. Sensor Packaging • Very important for in vivo applications • Long operational lifetime & biocompatibility • Host affecting sensor Protein absorption and cellular deposits leading to drift, sensitivity and stability loss • Sensor affecting host • Nontoxic • Nonthrombogenic • Covering to prevent toxin leaching

  6. Biopotential Measurements • Electrodes coupling ionic potentials in body to electronic instrument

  7. Electrolyte/Metal Electrode Interface • Metal in electrolyte solution creates charge distribution causing half-cell potential at interface

  8. Typically two similar electrodes of same metal used to measure biopotential • E.g. two similar electrodes taped to chest to measure electrical potentials generated by heart (electrocardiogram – ECG)

  9. ECG Electrodes • Mylar • Disposable snap-type Ag/AgCL

  10. EEG Electrodes • Electroencephalographic signals • Signals from the brain • Cup electrodes • Platinum or Tin cup • Subdermalneedles • Platinum or Stainless Steel needle electrode

  11. Microelectrodes • Ultra-fine tapered tip to be inserted into individual cells • E.g. recording action potentials in neurophysiological studies • Capillary glass microelectrode

  12. Electromagnetic flow transducer • Measure fluid flow , e.g. blood flow in Aorta • Potential difference V over vessel proportional to blood velocity • V = B x l x u • B = Uniform magnetic field • l = Diameter of vessel • u = Uniform velocity in vessel • Clips around vessel

  13. Potentiometer • Converts linear or angular displacement to voltage by sliding a contact along a surface of a resistive element • Elastic resistive transducer around chest

  14. Strain gauges • Measures strain of object as result of applied force • Gauge factor G indicator of sensitivity • G = ( ΔR / R ) / ( ΔL / L ) • Bonded • Unbonded

  15. Piezoelectric transducer • Contains piezoelectric crystal that generates a small electric potential if mechanically strained • Uses • Listening to heart sounds in cardiology • Automated blood pressure measurements • Generating ultrasonic waves for blood measuring

  16. Airflow Transducers • Fleisch pneumatachometer • Uses • Monitoring volume, flow and breathing of patients on mechanical ventilators

  17. Temperature Measurement • Temperature very important in healthcare • Measuring areas • Armpit • Bodily cavities • Thermistor • Small enough to ensure rapid response

  18. Non-contact Thermometer • Measures body core temperature inside auditory canal • Measures near tympanic membrane in ear canal

  19. Blood Gases and pH Sensors • Arterial blood gases (pO2 and pCO2) and pH frequently measured on critically ill patients under operation or in ICU. • Why? - Physicians use information to adjust mechanical ventilation systems / administering of pharmacological agents. - Information on respiratory and metabolic imbalances in the body. - Reflects adequacy of blood oxygenation and CO2 elimination.

  20. Traditionally: Blood withdrawn from peripheral artery and sample analyzed in laboratory. Presently: Continuous noninvasive blood gas monitoring - allows physicians to follow trends in patient’s condition and receive immediate feedback on therapeutic methods employed.

  21. Oxygen Measurement • Two ways to measure blood oxygen or oxygenation: 1. Polarographic pO2 Sensors 2. Optical Oximeter (oxygen saturation)

  22. Measures oxygen saturation, SO2, relative amount of oxygen carried by hemoglobin in erythrocytes. • Measurements of arterial blood SaO2 and venous blood SvO2 made in vivo or in vitro. • Method (oximetry) is based on light absorption properties of blood. • Specifically interested in relative concentration of Hband HbO2, since deoxygenated blood is blue and fully oxygenated blood is bright red. Optical Oximeters

  23. Measurement Method: • Measurement performed at two specific wavelengts: red wavelength (λ1) and near-infrared region of the spectrum (λ2). • λ1 – large difference between light absorbance of Hb and HbO2 – 660nm. • λ2 – either isobestic (around 805 nm where absorbencies of Hb and HbO2 are equal) or around 940 – 960 nm where absorbance of Hb is slightly smaller than HbO2.

  24. In vivo – Uses noninvasive pulse oximetercontaining • optical sensors for measuring SaO2. • Optical sensors consists of pair of small inexpensive • LED’s. • LED’s typically red (R) – 660 nm and infrared (IR) – 940 – 960 nm. • Highly sensitive silicon photodetector. • Components mounted inside reusable spring-loaded • clip or disposable adhesive wrap. • Electronic circuits inside oximeter switches LED’s • on and off in sequential manner and synchronously • measure photodetectoroutput.

  25. Pulse oximetry relies on detection of photoplethysmographic signal. • Signal caused by: • Changes in arterial blood volume with systolic heartbeat. • Magnitude of signal depends on: • Amount of blood ejected from heart into peripheral vascular bed. • Optical absorption of blood. • Composition and color of skin and underlying tissues. • Wavelengths used to illuminate blood.

  26. Optical Biosensors • Play important role in development of highly • sensitive and selective methods for • biochemical analysis. • Fundamental principle employed based on • change in optical properties of biological or • physical medium. • Change produced can be result of changes in • absorbance, reflectance, scattering, fluorescence, • polarization or refractive index of biological medium.

  27. Optical Fibers • Optical fibers used to transmit light from one location to another. • Fibers typically made from two concentric and transparent glass or plastic materials. • Center piece is known as the core and outer layer, serving as the coating, is called the cladding. • Core and cladding have different index of refraction, n.

  28. Case study: Fibre optic sensor • Measure the stress in the knee’s patellar tendon.

  29. Case study: Fibre optic sensor • The posterior region of the proximal patellar tendon is subjected to greater tendinous forces than the corresponding anterior region • Jumping and deep squat exercises expose the patellar tendon to very large forces • This should be taken into account during training and rehabilitation

  30. Sport technology – eccentric/concentric • Eccentric exercise is commonly recommended for the treatment of patellar tendinopathy • 8 subjects • Calibration • Squat exercise

  31. calibration

  32. concentric phase caused greater loads on the patellar tendon than the eccentric phase end