An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Medipix Collaboration Large scale x-ray images taken with the Medipix1 chip • Karl-Friedrich Pfeiffer • Physikalisches Institut, Abt. 4 • Universität Erlangen-Nürnberg • Workgroup SPOC* *Single PhOton Counting

  2. Overview • The Medipix1 chip • Experimental setup • Large Images 1: 'Move and Tile'Measurement and Simulation • Large Images 2: 'Tile and Move'Measurement and Simulation • DQE Measurements • Summary

  3. The Medipix1 chip • Pixelated solid state hybrid detector for single photon counting • 64 x 64 Pixel, size 170mm Þ active area 10.88 x 10.88 mm2 • Semiconductor sensor layer bump-bonded to the pixelated read-out electronics chip • Sensor layer: e.g. 300mm Si, but other materials (GaAs, Cd(Z)Te...) can be used as well • Developed within the framework of the Medipix1 collaboration

  4. Computer 1: main control Computer 2: data acquisition X-ray generator MUROS1 board & Medipix1 chip Translation stages Experimental setup • Complete imaging process controlled by computer 1 • Data acquisition done with Medisoft software (developed in the Medipix collaboration) • (Semi-)automated processing of acquired data

  5. Large Images 1: 'Move and Tile' • Make single exposure with single detector • Move detector to next position • Make another exposure • Repeat until the whole area is covered • Merge all single exposures to get one large tiled image

  6. Large Images 1: Measurement 5x5 single exposures, 10 pixels overlap

  7. Contrast ratio measurement -simulation: 0.996 0.978 Comparison Measurement - Simulation Object: „Siemensstern“ 0.05mm Pb in 2mm PMMA Ø 4.5cm • Contrast Pb -Air:0.965 • Contrast Pb-PMMA:0.941 • Contrast Pb -Air:0.969 • Contrast Pb-PMMA:0.962

  8. Large Images 2: 'Tile and Move' • Tile several detectors to make an array • Take large picture (with gaps) • Move array to next position • Take another picture • Repeat to cover all gaps • Merge pictures to get one large image without gaps and with high photon statistics

  9. 3 images, 4 x 4 tiles each Large Images 2: Measurement

  10. Contrast ratio measurement -simulation: 0.988 0.964 Comparison Measurement - Simulation Object: „Siemensstern“ 0.05mm Pb in 2mm PMMA Ø 4.5cm • Contrast Pb -Air:0.954 • Contrast Pb-PMMA:0.928 • Contrast Pb -Air:0.966 • Contrast Pb-PMMA:0.963

  11. Improved 'Tile and Move' If it is possible to tile several detecors without gaps (e.g. with the Medipix2 chip): Þ Tile detectors to get ’detector bars': two pictures are sufficient to cover all gaps Þ even faster & less dose!

  12. DQE Definition • DQE: Detective Quantum Efficiency • SNR: Signal-to-Noise-Ratio • in: SNR of incoming radiation field (~ N ) • out: SNR of the image • f = f(x,y): Spatial frequency

  13. Calculation of the DQE  : System gain factor MTF : Modulation Transfer Function NPS : Noise Power Spectrum <E> : Mean photon energy 0 : Incoming energy fluence

  14. 15.5cm DQE Measurement • MTF(f) measurement:line pair phantom • NPSout measurement:flat field images • ... some lenghty calculations ... Þ DQE(f)

  15. MTF: results

  16. First DQE results

  17. Summary • There are basically two ways to get large field images: 1) 'Move and Tile': single detector, flexible, but takes quite a long time and high dose Þ 'Lab Method' 2) 'Tile and Move': quite fast, but requires several detectors Þ 'Standard Imaging Method' • The MTF and DQE of a Medipix1 chip with a 300mm Si sensor layer was determined • We are looking forward to the Medipix2 chip

  18. Thanks to: • Prof. G. Anton1 • Ch. Bert1 & D. Niederlöhner1 • J. Giersch1 • M. Hoheisel2 & L. Bätz2 (for the DQE calculations) • Medipix Collaboration • B. Mikulec3 1Universität Erlangen 2Siemens 3CERN

  19. Phantoms used: Siemensstern Line pair phantom

  20. Background picture: blown fuse