An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SESAME Microtron ,TL1 and Booster Diagnostics Status Hussein AL-MOHAMMAD RF & DI Group SESAME 12th TAC

  2. SESAME Beam Diagnostic Systems • Mirotron DI • Transfer Line 1 DI • Booster DI

  3. SESAME Beam Diagnostic Systems • Mirotron DI • Deflection Tube, Gun Current, Magnetron : Forward & Reflected Power signals • Transfer Line 1 DI • Booster DI

  4. SESAME Microtron DI / Signals

  5. SESAME Microtron DI / Def. Tube

  6. SESAME Beam Diagnostic Systems • Mirotron DI • Deflection Tube, Gun Current, Magnetron, Forward & Reflected Power signals • Transfer Line 1 DI • Beam Current Monitor (BCM) • Beam Foil Monitor (FOM) • Fast Current Transformer (FCT) • Faraday Cup (FC) • Booster DI

  7. SESAME Transfer Line1 Diagnostic

  8. TL1 Diagnostic / BCM SESAME • BCM is installed in the TL1 with RF DC blocker. • The BCM not calibrated so it’s needs calibration in the Lab. • The signal from this cavity is mainly used to give us an indication for the beam passage . • Better solution for measuring the beam current is FCT ( Fast Current Transformer). • The new FCT will be used to calibrate BCM.

  9. TL1 Diagnostic / FCT SESAME • Recently we received an FCT from ANKA for temporary period to be used in SESAME TL1 to measure the current, and solve the problems with BCM. • The FCT is from Bergoz Instrumentation with BW 2GHz and 20 Turns. • RF shield for it were made in the lab. • The FCT is tested in the Lab and it’s already installed in TL1. • FCT was installed in TL1 between FOM and FC and this temporary till purchasing new in-air FCT FCT with Shield While testing in the Lab

  10. TL1 Diagnostic / FOM SESAME • The FOM is already installed in TL1. • We use florescent screen AF995R. • The problem with this type of materials is the long brightness time. • Right now we use old BESSY video camera and take the signal directly to TV monitor in control room.

  11. TL1 Diagnostic / FC SESAME • The Faraday Cup is already installed in TL1. • This FC can handle 35Mev up to 7.6 cm stop range. • The stopping range @22Mev is 4.71cm FC

  12. SESAME Beam Diagnostic Systems • Mirotron DI • Deflection Tube, Gun Current, Magnetron, Forward & Reflected Power signals • Transfer Line 1 DI • Beam Current Monitor (BCM) • Beam Foil Monitor (FOM) • Fast Current Transformer (FCT) • Faraday Cup (FC) • Booster DI • Beam Foil Monitor (FOM) • Beam Position Monitor(BPM) • Fast Current Transformer (FCT) • New Parametric Current Transformer (NPCT or DCCT) • Visible Beam Light • Tune Measurement

  13. Booster Diagnostic SESAME BPM BPM FOM DCCT FCT

  14. Booster Diagnostic/FOM SESAME • 3 FOMs in the booster • One at injection septum • One in section 1 and the third one in section 2 • Pneumatic system will be used for all FOMs . • Pneumatic system is tested and ready. • Most of the screens are good but the septum FOM must be replaced. • The cameras will be used good quality TV video camera.?????

  15. Booster Diagnostic/ CT SESAME • Old CT’s are available from BESSY. • New devices are purchased: • FCT from Bergoz • NPCT (DCCT) from Bergoz

  16. Booster Diagnostic/ FCT,NPCT SESAME • These new FCT and NPCT can give us the exact and accurate current. • There is enough space in booster sections 3 & 6, FCT will be installed in booster section 3 and the NPCT in booster section 6, we can swap between them. • These CT’s are in-air type. • The electronics for NPCT(DCCT) and cables provided from Bergoz.

  17. Booster Diagnostic/BPM SESAME • 6-BPM blocks are available. • 4-SLM. • 2-BP. • Phase matched cables are used. • All of these BPMs will be tested in the Lab using the instruments :Spectrum Analyzer ,Network Analyzer, Oscilloscope and design test bench.

  18. Booster Diagnostic/Libera SESAME • Libera electron had arrived to SESAME from Instrumentation technologies (Slovenia) • We have 7 Libera electron for the booster and 1 clock splitter. • Cable lengths are as follows: • BPM1,3 is 23m. • BPM4,6 is 29m. • BPM2 is 17m. • BPM5 is 35m. Libera clock splitter Libera electron

  19. Booster Diagnostic / Libera SESAME

  20. Booster Diagnostic/Libera SESAME

  21. Booster Diagnostic/Libera SESAME

  22. Booster Diagnostic/Libera SESAME

  23. Booster Diagnostic/Libera SESAME

  24. Booster Diagnostic/Libera Rack SESAME

  25. Booster Diagnostic/Tune Measurement SESAME • From last TAC no change in plan for tuning measurement. • Two 10W Amplifier will be purchased with frequency (0.5-525)MHz. Shaker Booster Amp NGy BPM4 Libera Detector Control System Amp Data NGx

  26. SESAME