Test tool for assessing lead equivalence in protective lead apparels

This study introduces a method to assess lead-equivalence in protective lead apparels using digital X-ray imaging. Results indicate the effectiveness of the test tool in evaluating lead attenuation and identifying defects in lead aprons. The tool aims to ensure radiation workers' safety by providing an affordable and simple screening solution.

latonyal
Download Presentation

Test tool for assessing lead equivalence in protective lead apparels

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Test tool for assessing lead equivalence in protective lead apparels Roshan S Livingstone and Anna Varghese Department of Radiology Christian Medical College, Vellore, S India Vision statement: The Christian Medical college, Vellore seeks to be a witness to the healing ministry of Christ through excellence in education, service and research

  2. Introduction • Protective lead‐equivalent (Pb-eq) aprons provide necessary shielding from secondary radiation (scatter). • Pb-eq aprons have vinyl/rubber with lead impregnation or made of composite materials (non lead) • Use of high tube potentials (kVp) lead to more scatter radiation. • Knowledge on Pb-eq is important for radiation protection

  3. Why this is of concern? C CT and fluoroscopic images showing defects in lead apparels Arrow showing a dark line indicating fragility. Nonuniformity of material. Manufacturing defect in thyroid collar. Different combination of lead equivalent material D A B

  4. Checking integrity of apron • Aprons are screened using CT/fluoro/radiography to check for integrity before use and as part of the annual quality control procedure • Lead equivalence can be assessed using dosimeter • Expensive and available only in few centers • Requires qualified personnel to interpret • A simple method to identify the Pbeq is required for screening purposes

  5. Aim • Create awareness to the radiation worker on integrity of radiation protective apparels • For new purchases • For adequate radiation safety. • Develop a simple test tool to test the Pb-eq of radiation protective apparels

  6. Materials and methods • Digital images of 0.25 mm and 0.5 mm lead and lead‐free aprons from 6 manufacturers were assessed using a calibrated digital X‐ray unit. • Standard protocol non-grid technique • Percentage attenuation values of the aprons were determined at 100 kVp, 32 mAs, 100cm using an ionization chamber

  7. Evaluating pixel values • Region of interest (ROI) from radiographic images of lead apron, copper step wedge, and 2 mm thick lead (reference). • GE Centricity PACS workstation (Milwaukee, USA) • The mean pixel value from the apron was matched with corresponding step (thickness of copper) from step wedge to monitor Pbeq • The pixel value ratio (PVR) was calculated from mean pixel intensity of the apron and lead block and lead block versus step wedge.

  8. Results • Percentage attenuation at 100 kV, 32 mAs • 90 % attenuation for 0.25 mm Pbeq • 97 % attenuation for 0.5 mm lead or lead‐free aprons. • Pixel values – images acquired at 100 kV, 5 mAs • 0.25 mm Pbeq apron correspond to 0.8–1.2 mm Cu (4 – 6 steps) • 0.5 mm Pbeq aprons correspond to 2.0 –2.8 mm Cu (12 – 14 steps) • Pixel value ratio (lead block/copper step) was constant for 0.25mm Pbeq apron (0.5) and for 0.5mm (0.7) • Shows consistency of output

  9. Discussion • Defects in the aprons include cracks, changes in homogeneity of pbeq, combination of 0.25 mm, and 0.5 mm Pbeq materials. • Deterioration of the lead‐impregnated vinyl is manifested as cracks or holes and may not be visible to the naked eye • Technique of measuring the attenuation using transmission ionization chamber is standard, but expensive. An alternative method is discussed in this study – copper step wedge • Both lead and lead free (composite bilayer: bismuth-antimony-barium) gives similar attenuation (96 – 98%)

  10. Discussion • Pixel values increased with increase in the thickness of copper step wedge indicating a corresponding increase in Pbeq in aprons. • The simple test tool is affordable and is validated in this study. • Calibrated radiographic unit and standard protocols for acquiring images should be selected

  11. Conclusion • It is suggestive that aprons be screened for its integrity from the time of purchase using computed tomography (CT), fluoroscopy, or radiography. • It is recommended that this simple test tool could be used for checking lead equivalence as part of periodic quality control (or) if any variation in contrast is seen in the image during screening. Livingstone et al Ijri 28:2; 2018

  12. Disclosure – No funding involved Conflict of interest - NIL