1 / 7

Fluorescence Spectroscopy

Fluorescence Spectroscopy. Quantum Yield f = ________k f ______________ k f + k i + k ec + k ic + k pd + k d k f = rate constant for fluorescence k i = “ “ for intersystem crossing k ec = “ “ for external conversion k ic = “ “ for internal conversion

liam
Download Presentation

Fluorescence Spectroscopy

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Fluorescence Spectroscopy

  2. Quantum Yield f = ________kf______________ kf + ki + kec + kic + kpd + kd kf = rate constant for fluorescence ki = “ “ for intersystem crossing kec = “ “ for external conversion kic = “ “ for internal conversion kpd = “ “ for pre-dissociation kd = “ “ for dissociation What factors control whether a molecule will fluoresce?

  3. What factors control whether a molecule will fluoresce? B. Structure: 1.) Aromatic 2.) Rigid structures exhibit more 3.) Heavy atoms will decrease fluorescence 4.) Fluorescence will increase when molecule is adhered to surface

  4. What factors control whether a molecule will fluoresce? C. Temperature and Solvent Effects 1.) Lower temperature increases fluorescence 2.) Solvent contains heavy atoms will decrease fluorescence but increase phosphorescence

  5. Effect of Concentration on Fluorescence • Effect of concentration on Fluorescence • F = k’Po2.303ebc F = Fluorescence k’ = constant depending on quantum efficiency Po = incident power For high concentrations linear relationship no longer holds due to self quenching What is self quenching? Excited molecules collide and radiationless transfer of energy occurs

  6. Instrument Design for Fluorimeters Important considerations for fluorimeters 1. Must have intense source, D2 & W not intense enough so use Xe or Hg • Fluorescence is not as strong as absorbtion so need more sensitive detectors like PMT • 3. Cells transparent on 4 sides detection usually at right angles of source 4. Phosphorescence measurements may be made on fluorimeters w/ 2 differences • Irradiation + time delay + measurement • Cooled samples @ 77K to prevent collisional deactivation

  7. Applications of Fluorescence • Fluorescence and phosphorescence are one of the most sensitive techniques available!

More Related