XMUGXQ PFS0601 Principles of Fluorescence Spectroscopy Chemistry Department XMU
XMUGXQ PFS0601 Chapter Seven Measurement of Fluorescence Lifetime & Time-domain Fluorescence & Frequency-domain Fluorescence
XMUGXQ PFS0601 Content 7.1 introduction 7.2 pulse lifetime measurement 7.3 Phase and Modulation Measurements of Fluorescence Lifetime 7.4 Measurement of Time-resolved Decays of Fluorescence 7.5 Application of Time-resolved Fluorescence 7.6 Phase-sensitive Detection of Fluorescence 7.7 Application of PSDF
XMUGXQ PFS0601 7.1 Introduction Information given by fluorescence lifetime • The frequency of collisional encounters • The rate of energy transfer • The rate of excited state reaction • Information related to its environment And so on
XMUGXQ PFS0601 I0 I0 t t Methods for measuring fluorescence lifetime • Pulse method Light source • Harmonic or phase-modulation method Light source
XMUGXQ PFS0601 hvA hvF Log F(t) or log N(t) S1 S1 knr t S0 7.2 pulse lifetime measurement
XMUGXQ PFS0601 Average lifetime 1 ns, 1 2 ns, 1 3 ns, 1 For a large number of fluorophores and small time interval, this sum becomes
XMUGXQ PFS0601 5 ns 50 ns Multi-component Pre-exponential factor
XMUGXQ PFS0601 7.3 Phase and modulation measurements of fluorescence lifetime Phase angle () Demodulation factor (m)
XMUGXQ PFS0601 Phase-modulation method =2f an important factor For small lifetime, set large modulation frequency For large lifetime, set small modulation frequency Choosing modulation frequency, let m = 0.3 ~ 0.7, = 30 ~ 70º For commercial frequency-domain instrument, changing , measuring miand i, calculate average lifetime
XMUGXQ PFS0601 Phase-modulation method
XMUGXQ PFS0601 Phase-modulation method Single component multicomponent Average lifetime
XMUGXQ PFS0601 I0 t 7.4 Measurement of Time-resolved Decays of Fluorescence Pulse lifetime measurement • Pulse width Enough shorter compare to decay of fluorescence ps, fs • Photon counts Enough for accurate measurement Repeat excite
XMUGXQ PFS0601 Pulse sampling Method Photomultiplier Multichannel analyzer MAC
XMUGXQ PFS0601 Single photon counting method Time to amplitude converter TAC Multichannel pulse heigh analyzer MCPHA
XMUGXQ PFS0601 Streak Camera Simultaneous measurements of both wavelength and time resolved decays
XMUGXQ PFS0601 Log F(t) or log N(t) t Analysis of time-resolved decays of fluorescence intensity In principle, for single exponential decay
XMUGXQ PFS0601 Analysis of time-resolved decays of fluorescence intensity In practice, in consideration of the pulse width of lamp and multi-exponential decay Intensity profile of light, L(t) Intensity decay of fluorescence, F(t) Measured intensity decay, R(t)
XMUGXQ PFS0601 Analysis of time-resolved decays of fluorescence intensity In practicing measurement • Measuring the lamp profile, L(t), by using a solution which scatters light • Measuring the total intensity decay, R(t), by using the sample At ti, a large number of pulses with equal width ti, each induce an impulse response in the sample t - ti, emission delay compare to excitation
XMUGXQ PFS0601 Analysis of time-resolved decays of fluorescence intensity Total intensity decay, The purpose is to get F(t) Commercial software available
XMUGXQ PFS0601 Analysis of time-resolved decays of fluorescence intensity • Least –squares analysis of time-resolved decays Let the number of components to be n, Give initial values to i and i, and calculate, get L(t), was measured
XMUGXQ PFS0601 Analysis of time-resolved decays of fluorescence intensity The i and ivalues are varied until the best fit is observed. In this expression the sum extends over the number (n) of channels or data points used for a particular analysis. A minimum value of 2indicates the best fit.
XMUGXQ PFS0601 Analysis of time-resolved decays of fluorescence intensity
XMUGXQ PFS0601 Example Single exponential decay Double exponential decay
XMUGXQ PFS0601 Example Single exponential decay Double exponential decay
XMUGXQ PFS0601 I t 7.5 Application of time-resolved fluorescence • Deduct back ground Measure intensity ex/em Boxcar integrator Target component Sampling time Back ground Fast scan ex/ scan em Light Gated time
XMUGXQ PFS0601 Example 荧光素标记荧光免疫分析,常常受到血液样品中胆红素背景荧光干扰,采用时间分辨荧光免疫分析可以有效地消除干扰。 荧光素寿命 3.6±0.46 ns, 测定时间 6.0 ns 胆红素寿命 0.21±0.14 ns, 测定时间 信号为零
XMUGXQ PFS0601 I t Application • Multi-components measurement
XMUGXQ PFS0601 [Q] Application • Dynamic quenching
XMUGXQ PFS0601 Application • Time-resolved fluorescence immunoassay Example TBP-Eu3+ XL665 D A TBP-Eu3+ 链[霉]亲和素
XMUGXQ PFS0601 Example Interference: Background from media Emission from free acceptor
XMUGXQ PFS0601 background Application Free XL665 TBP-Eu3+(665nm/620nm) FRET(665nm/620nm)
XMUGXQ PFS0601 Expression for the Figs
XMUGXQ PFS0601 Reference
XMUGXQ PFS0601 Time-resolved emission spectra
XMUGXQ PFS0601 I(t) mL=b/a F(t) mA=B/A F(t) mB=B’/A’ 7.6 Phase-sensitive detection of fluorescence Principle A<B mL>mA>mB A<B
XMUGXQ PFS0601 principle For single component Excited with Emission Phase sensitive detection (lock-in amplifer ) D, detection phase
XMUGXQ PFS0601 F F Principle At i At i F change with cos(D-) F change with
XMUGXQ PFS0601 principle For two component Emission Phase sensitive detection Phase depression
XMUGXQ PFS0601 Phase sensitive detection fluorescence spectra
XMUGXQ PFS0601 Example
XMUGXQ PFS0601 Example
XMUGXQ PFS0601 Example
XMUGXQ PFS0601 Example