110 likes | 247 Views
FETS Pepperpot. Simon Jolly 10 th February 2010. The Pepperpot. Beam segmented by intercepting screen with regular array of holes. Beamlets drift short distance before producing image on scintillator screen. Support block prevents beamlets from overlapping and provides cooling.
E N D
FETS Pepperpot Simon Jolly 10th February 2010
The Pepperpot • Beam segmented by intercepting screen with regular array of holes. • Beamlets drift short distance before producing image on scintillator screen. • Support block prevents beamlets from overlapping and provides cooling. • CCD camera records image of light spots. • Calculate emittance from spot distribution. Scintillator screen Support block Fast CCD Camera Incoming Beam Intercepting screen Beamlets Simon Jolly, Imperial College
FETS Pepperpot Design Beam profile head Tungsten mesh Pepperpot head Shutter Bellows Camera Moving rod Vacuum bellows Mounting flange Simon Jolly, Imperial College
Pepperpot and Slit-Slit Scanners Simon Jolly, Imperial College
Calibration System • Need to produce 4 light spots on scintillator screen for calibration. • Bring in 4 optical fibres through spare holes in copper mounting block. • Fibre stops a little way short of scintillator screen, mounted in centre of 2mm x 13mm glass dowel. • Glass dowel is frosted on scintillator screen end. • Light from fibre produces 1mm spot on frosted end of glass dowel ie. at scintillator screen. • Spots are imaged for calibration. Scintillator screen Fast CCD Camera To light source Copper mounting block Simon Jolly, Imperial College
Pepperpot Head FETS pepperpot uses tungsten intercepting screen with 50 micron holes on 3 mm pitch, 10 mm drift and quartz scintillator. Simon Jolly, Imperial College
Pepperpot Head: Mounting Holes FETS pepperpot uses tungsten intercepting screen with 50 micron holes on 3 mm pitch, 10 mm drift and quartz scintillator. Tungsten screen and copper mounting blocks are aligned with 4 2mm x 13mm metal dowels. Simon Jolly, Imperial College
Pepperpot Head: Calibration Holes Fibres come in from back side of head. FETS pepperpot uses tungsten intercepting screen with 50 micron holes on 3 mm pitch, 10 mm drift and quartz scintillator. 4 spare 2mm dowel holes to be used for fibre optic calibration Tungsten screen and copper mounting blocks are aligned with 4 2mm x 13mm metal dowels. Simon Jolly, Imperial College
Pepperpot Support & Main Flange Cooled feed through rod. The linear shift mechanism (LSML) from UHVDesign mounted to the main flange. KF-40 vacuum flanges used for feedthroughs Adjustable camera mount. Simon Jolly, Imperial College
Pepperpot Mounted to Vacuum Tank KF-40 flange to be used for fibre optic feedthrough. Simon Jolly, Imperial College