An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Anesthesiology Neuromuscular MonitoringSystem ANTS #25

  2. Group Information Greg Angelotti (EE) Otto Cruz (CpE) Kristin “Zaza” Soriano (CpE/EE) Sponsor: Dr. Thomas Looke (Anesthesiologist/EE)

  3. Current Methodology • Administer Neuromuscular Blocking Agents (NMBA) • Blocks Acetylcholine (ACh) • Trivial neuromuscular stimulating • Subjective muscle response monitoring (if any at all) • Pen and paper record keeping

  4. Current Methodology *Courtesy of nursingcrib.com

  5. Current Methodology * Research shows not as reliable as the first two sites

  6. Objectives • To safely stimulate targeted sites • To effectively obtain patient signals • To accurately record patient feedback • To create a more organized environment

  7. Requirements • Must consist of safe modular subsystems • Must obtain objective quantitative muscular responses • Must produce individual patient feedback and logging • Must reduce operating room clutter

  8. Overall Block Diagram

  9. Specifications

  10. Shocking Device Schematic

  11. Digital to Analog Converter • DAC 7800 Specification • Two 12 bit SPI controlled DAC. • 5 Volt Supply with a 10 Volt Reference • SPI Inputs CS, Din, and Clk. • Control inputs UPD_A, UPD_B, Clr. • Control OuputsOut_A, Out_B.

  12. Analog to Digital Converter • 24 input bits sent in three 8 bit pulses. • VDAC = (V–) + 4.75V – (31.6kΩ) (ILIM)/5000, Current and voltage control formula.

  13. OPA2336 Opamp • 5 Volt power supply • CMOS unity-gain to create a stable output. • Low offset voltage 125 micro Voltes max. • Low 20 micro Amps quiescent current.

  14. OPA547 Opamp • Single supply voltage range of 8 to 60 volts. • Peak output current of 750 mA and 500 mA continuous. • Current control signal ILIM of 0 to 150 micro Amps produces 0 to 750 mA output. • Voltage Control from 0.8 to 25 Volts output

  15. 5A200S DC-DC Converter • Supply voltage of 5 volts. • Output voltage of 250 volts peak and 200 volts continuous. • 0.1 microfarad capacitor. • Input voltage will be pulsed at selected waveform.

  16. TPS61093 DC-DC Converter • Supply voltage of 1.6 to 6 Volts. • Output voltage from 1.6 to 17 volts with a 1.1 Amp peak over current protection. • Three outputs 3.7, 5 , 10 volts, enable pins turn converters on. • R1 and R2 control output voltage.

  17. Overall Block Diagram

  18. Specifications

  19. Sensor Device

  20. Surface EMG • Choose targeted site • Prepare the site • Shave • Sanitize • Place Ag/Cl reference electrodes • Disc Electrodes [<10 mm] • Ready to read patient action potentials • 0-6 mV [Typical]

  21. sEMG Preconditioning

  22. Prototype sEMG Testing Differential Amplification (High Impedance) Differential Amplification Hand at rest

  23. Prototype sEMG Testing Full-Wave Rectification Smoothed Rectification

  24. Prototype sEMG Testing High Cut-Off and Final Gain Stage *Prototype Design Credit to Gundanium on instructables.com

  25. Sensor Device

  26. Tactile Pressure Sensor

  27. Pressure Sensor Conditioning Voltage Divider • Standard Resistance from sensor (FSR): • Appr. 1 MOhm • Vin comes from CR2032 • Appr. 3Volts • The more the surface is pressed, the less sensor resistance Vout = (Vin)*(R2/Rsensor+R2) R2 = 270 Ohms

  28. Sensor Device

  29. Flexible Potentiometer Sensor

  30. Bend Sensor Conditioning Voltage Divider • Standard Resistance from sensor (FSR): • Appr. 10 KOhm • Vin comes from CR2032 • Appr. 3Volts • The more the sensor bends the more the less impedance Vout = (Vin)*(R2/Rsensor+R2) R2 = 270 Ohms

  31. Overall Block Diagram

  32. Specifications

  33. Main Device

  34. Touchscreen

  35. Touchscreen Controller I2C Voltage-level Shifter GND SCL GND SDA SCL SDA 1.8 V 3.3 V VIO Pandaboard ES J6 Expansion Header 3.3 V Step-up

  36. Main Device

  37. TFT-LCD

  38. LCD Controller LED Backlight Control I/O Pandaboard ES J3/J4 30-pin LVDS Connection LCD I/O

  39. Main Device

  40. Pandboard ES

  41. Pandaboard ES

  42. MCU & Bluetooth

  43. Communication for BLE • Project specific • Medical Applications • Energy Efficiency • Protocol • Generic Attribute Profile (GAP/GATT) • Runs on top of Attribute Protocol • Server-Client Scheme • Characteristics hold values of readings

  44. GATT_Character_UUIDs

  45. Tablet Software • Ubuntu 12.10 armhf • Custom Kernel • Python • BlueZ stack • PyQtGraph • Plot real time response

  46. Monitoring Screen

  47. Bill of Materials (Tablet)

  48. Bill of Materials (Sensors)

  49. Bill of Materials (Total)