An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. WL3100 Dry Pressure Transducer • Temperature Compensated Measurement • LCD Display • SDI-12 interface • 4-20mA interface(at the same time) • Option : Extra SDI-12 By : Steve Laurence Hydrological Services P/L

  2. Using a Bubbler to Measure Pressure :

  3. Deriving the Relationship :

  4. Deriving the Relationship : We are taught that : “Force = mass x acceleration” the amount of force that a body of water exerts is dependent upon the acceleration due to gravity. The mass of an object is : “mass = volume x density” so now we have : Force = mass x acceleration due to gravity = (volume x density) x gravity = (area x depth x density) x gravity Relating back to pressure : “Pressure = Force / Area” (substitute the Force from above) = (area x depth x density) x gravity / area (area cancels) Pressure = depth x density x gravity Solve for depth : Depth = Pressure / ( density x gravity) Water Depth = Pressure x (1 / (density x gravity)) Therefore : User Factor = 1 / (density x gravity) NOTE : We have now shown that the User Factor is a function of water density and gravity !!! You might think they are both constants, but you would be wrong !!!!!!!!! The User Factor we use is 0.101972 m / kPa

  5. Water Salinity : • As water salinity increases, the density increases • The User Factor is proportional to 1 / density • Therefore the User Factor will decrease as salinity increases.

  6. LCD Menus : Navigate with the “Scroll” and “Select” buttons

  7. Setting Up 4-20mA : • The 4-20mA circuit is optically isolated from the rest of the electronics • Therefore it must be powered by the loop • Set the 4mA Level using the WL3100 LCD menu • Set the 20mA Level using the WL3100 LCD menu(You can use –ve levels) 4mA / 20mA Level 0.0 / 10.0 m To conserve power, the data logger switches power to the 4-20mA circuit only when it wants to make a measurement. ML420 Data Logger measuring the 4-20mA

  8. Sampling Modes : WL3100 S/W Rev 7.1 (and lower) • When “Continuous Mode” is set to “NO” the WL3100 only takes a sample when : • - The water level is displayed on the LCD or • - An SDI-12 Measure command is received or • - The 4-20mA loop power is detected. • When “Continuous Mode” is set to “YES” the WL3100 takes a sample : • Every 12 secs. (Continuously) • (This is useful if you use the SDI-12 command 0R0! as it will immediately return the most recent measurement made within the last 12 secs – which makes the WL3100 SDI-12 interface the same as the AD375A SDI-12 interface. It can also be useful when using the 4-20mA interface, as the 4-20mA loop current will be accurately updated 100mS after loop power is applied – without having to wait 12 secs for a measurement to be made !!)

  9. Sampling Modes : WL3100 S/W Rev 7.2 (and higher) • When “Sampling Mode” is set to “Single” the WL3100 only takes a sample when : • - The water level is displayed on the LCD or • - An SDI-12 Measure command is received or • - The 4-20mA loop power is detected. • When “Sampling Mode” is set to “Contin+Avg X” the WL3100 takes a sample : • -Every 12 secs. (Continuously)(When a sample is requested, via the display or via SDI-12 or 4-20mA, the last X samples are averaged. If it is known that the water level will change very slowly, it is advantageous to set X to a high number, so that the last say 20 samples will be averaged.)

  10. Sampling Modes : 5 Sample Average Example :

  11. Full Temperature Compensation : • Each WL3100 undergoes a full Temperature Compensation Calibration at our factory in Liverpool. • 10 units at a time in our Environmental Chamber • -20C, -10C, 0C, 10C, 20C, 30C, 40C, 50C, 60C, 70Cfor a 1 hour soak, for everything to stabilise • At each temperature we do a 7 point pressure calibration over the required range : 0%, 15%, 35%, 50%, 65%, 85%, 100% • Ruska 7250i pressure calibrator is accurate to 0.005%(for example 10.000m x 0.005% = 0.5mm) • Calibration of each unit takes over 24 hours

  12. Field Calibration Check HS40 : • Navigate to the WL3100 “View Pressure” display • Set River Line Select to “Close” position • Set Calibrate Valve to “Calibrate” position • Set the Pump Up calibrator to “Vent” • Connect Pump Up calibrator to the Bubble Rate Test PortCalibrator Pressure is now read on the WL3100 • Remove Pump Up calibrator from Bubble Rate Test Port • Set Calibrate Valve to “Open” • Set River Line Select to “Open”