Chapter 23. An Introduction to Analytical Separations. Partition Between Two Phases A. Separatory Funnel two immiscible liquids -- solutes soluble in both. Partition Coefficient , K K = CU / CL after equilibration. K depends upon: solute -- polarity, ionization, etc.

Download Presentation

Chapter 23

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chapter 23 An Introduction toAnalytical Separations

  2. Partition Between Two Phases • A. Separatory Funnel • two immiscible liquids -- solutes soluble in both

  3. Partition Coefficient, KK = CU / CL after equilibration • K depends upon: • solute -- polarity, ionization, etc. • 2 liquid phases -- polarity etc. • If K >> 1 or K << 1 for a particular solute ==> can purify the solute pretty well with a single extraction (or relatively few extractions). • But what if K is approximately equal to 1?

  4. Let's look at two cases: (a) K = 1 (b) K = 3 • if K > 1 ==> more material in right most vessels --> peak moves to right quicklyif K < 1 == > more material in leftmost vessels --> peak moves to right more slowly(In the next diagram, lower phase stays in place while the upper phase is transferred to the next vessel to the right with fresh lower phase.)

  5. General Principle - Differential Hold-up • Method - Column • - Thin layer • - Paper • Columns can be of various sizes on the benchtop, or smaller with smaller sized packing and used in an instrument. • Generally there is a MOVING PHASE and a STATIONARY PHASE

  6. Principle (reason for hold-up) • Adsorption • Partition • Ion Exchange • Size Exclusion • Affinity

  7. Liquid chromatography • Liquid - liquid(adsorbed or bonded on solid) • Liquid - solid • SFC - liquid (adsorbed or bonded on solid) • Gas chromatography • Gas - liquid(adsorbed or bonded on solid • Gas - solid

  8. Chromatography is the separation - we must also have a detection method to follow the separation • Basic detectors - give a peak at a characteristic time for each component. • The most desirable, but more expensive detectors, in particular mass spectrometers, give a unique spectrum of each compound as it comes off, from which it can be identified.

  9. Ion Exchange Chromatography • In ion exchange chromatography, charged substances are separated via column materials that carry an opposite charge. • The ionic groups of exchanger columns are covalently bound to the gel matrix and are compensated for by small concentrations of counter ions, which are present in the buffer. • When a sample is added to the column, an exchange with the weakly bound counter ions takes place.

  10. Resin Two basic types of ion exchangers: • those for binding positively charged ions or cations, which display on their surface negatively charged groups • and those for binding negatively charged ions or anions, which display on their surface positively charged groups.

  11. Polystyrene-divinylbenzene resin (PS-DVB) • Most common polymer base for ion-exchange chromatography. Ionic groups are incorporated by chemical reactions. Porosity and mechanical stability are altered by varying the cross-linking through the variation of the DVB content. • Low crosslinking – resin swells a lot, changes size with different ions. • High crosslinking – solubility decreases - selectivity increases

  12. Anion exchangers • R-NR3+ OH- + A- R-NR3+ A- + OH-strong base pH 0-12 • R-NH3+ OH- + A- R-NH3+ A- + OH-weak base pH 0-9

  13. Apparatus • column • support • stationary phase • mobile phase • detector

  14. Column • copper tubing • stainless steel tubing • glass tubing • Polymer (plastic) Support • finely divided solids • ground firebrick • alumina, specially treated • walls of column for capillary columns

  15. Stationary Phase • stationary phase evenly dispersed on surface of support Mobile Phase • sample mixture carried through stationary phase by mobile phase • non-reactive gas in glc (gas-liquid chromatography, gc) • non-reactive liquid in llc (liquid-liquid chromatography, lc)