Miysaka Lab. Keisuke Yamada

Alexandros Pertsinidis, Yunxiang Zhang & Steven Chu. Miysaka Lab. Keisuke Yamada. Contents. I. Introduction II. Experiment Sample & Setup III. Results & Discussion The separation between differently coloured fluorescent molecules IV. Coclusion. Optical detection. : resolution.

Download Presentation

Miysaka Lab. Keisuke Yamada

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Alexandros Pertsinidis, Yunxiang Zhang & Steven Chu Miysaka Lab. Keisuke Yamada

  2. Contents I. Introduction II. Experiment Sample & Setup III. Results & Discussion The separation between differently coloured fluorescent molecules IV. Coclusion

  3. Optical detection : resolution NA : numerical aperture λ : Wavelength Optical microscope A type of microscope which uses visible light and a system of lenses to magnify images of small samples ・ Advantage nondestructive, noncontact observe inside of a sample ・ Disadvantage diffraction-limited optics (回折限界) limiting the resolution of the microscopy (~200nm)

  4. Super-ResolvingMicroscopy The resolution power is higer than the optical microscope (≦100 nm) Example of the method ・ STimulatedEmissionDepletion (STED) Microscopy ・ Scanning NearFieldOpticalMicroscopy (SNOM) ・ Photoactivated LocalizationMicroscopy (PALM) ・STochastic Optical Reconstruction Microscopy(STORM) ・ To identify the structure of a biological molecule ・ Nanoscale property in materials can be evaluated.

  5. STED DM DM detector Excitation laser STED laser + = STED spot Excitation spot The diameter fluorescence spot is decreased and get better resolution power

  6. SNOM ~100 nm SNOM is a kind of Scanning Probe Microscope using optical fiber as probe.

  7. PALM/STORM 生物物理50(4),174-179 藤田克昌 Super-resolution with wide-field microscope

  8. Motivation To know about the structure of multisubunit biological complexes in biologically relevant environment The spatial resolution is limited by noise and systematic effects that include drift and vibration of the measurement apparatus To improve the spatial resolution utilizing closed-loop feedback system

  9. Sample biotin PEG PEG silan silan silan glass glass glass glass coated glass silanize coated with biotin coated with PEG BSA Streptavidin biotin-PEG beads DNA 100 m

  10. Setup Closed-Loop feedback system

  11. Result 320 nm 3.2 nm The feedback systems controll the position ・ By locking the pinhole position in each image, they avoid the effect of the apparatus’ drift ・ By locking a Cy3 molecule and kept same the position, they demonstrate 0.7 nm molecule to molecule reproducibitliy

  12. Sample Polystyrene bead measurement Locked by feedback system Laser Optical Tweezers Measuring the position of the Alexa 647 and using the mapping calibration

  13. Result a c b 0.77 nm spatial resolution is achieved

  14. Conclusion ・ The Closed-roop feedback system could avoid vibration of the measurement apparatusandthe combined effects of the relay imaging optics and CCD array. ・ The spatial resolution of 0.77 nm is attained for the optical microscope. ・ The closed-loop control and registration could become essential concepts in design of future sub-10-nm optical lithography tools