1 / 7

Soft X-Ray Diagnostic System for HT-7 and EAST Tokamaks

HT-7. ASIPP. Soft X-Ray Diagnostic System for HT-7 and EAST Tokamaks. LIQUN HU. HT-7. ASIPP. Physical features of the diagnostic. I SXR  N e , T e , Z eff.

maya
Download Presentation

Soft X-Ray Diagnostic System for HT-7 and EAST Tokamaks

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. HT-7 ASIPP Soft X-Ray Diagnostic System for HT-7 and EAST Tokamaks LIQUN HU

  2. HT-7 ASIPP Physical features of the diagnostic ISXR Ne, Te, Zeff • A chord-integrated measurement—point measurement, high resolution in space and high temporal response in time to the variations happened inside of the plasma. • The intensity distribution and evolution and tomography of SXR radiation-------information on the plasma structure, MHD instability, sawtooth instability, phenomena occurring during major disruptions, heat pulse propagation transport et al.

  3. HT-7 Vertical array with 37 detectors of SBDs Au-Si detector Slot-aperture with a movable slider covered with Be foils Au-Si detector Horizontal array with 37 detectors of SBDs ASIPP Two multi-channel arrays, 37 detectors of rectangular Au-Si SBDs with active area of 320 mm2, a pinhole camera like SXR Array Diagnostic Features in HT-7 Sensitive energy region: 0.5-1.5keV Spatial resolution: 1.5cm Time resolution: 5s The outmost chord: 27cm Arc-shape SS pedestal with R=55cm (H) and R=60cm (V) Thin slot-apertures : 520 mm2 (V) 425 mm2 (H) A different adjustable mechanism for 4 thickness of Beryllium absorption foils : 5~6m, 12~13m, 25m, and full open

  4. HT-7 ASIPP SXR Array Diagnostic with SBDs in HT-7 Plug-in electronics and fast connectors for convenient repairing of the system

  5. HT-7 Au-Si detector Signal current transform to Voltage Voltage preamplifier with variable gain Sawtooth Heat Pulse Propagation for e Output DC Regulator Rejecter CAMAC Power supply ~220V VAX3300 Computer ASIPP Schematic drawing of electronics and data acquisition Differential Circuit

  6. HT-7 ASIPP Modification of HT-7 SXR detector HAMAMATSU: S4114-46N/35N Low terminal capacitor and high speed Down: 0.510mm slit, 1.6cm~0.7cm Below: 0.510mm slit, 1.6cm~0.7cm Above: 1.010mm slit, 1.5cm~0.9cm

  7. HT-7 ASIPP HT-7U SXR Pinhole Camera 11, each with 46 elements, 3 toroidal positions, 1-20keV, 1.0cm, cooling plate, adjustable mechanism for protection.