BINP level meter for cold liquids

BINP level meter for cold liquids. The method of measurements: change of the capacitance as function of a level of a liquid Originally designed for LN 2 at the CMD-2 detector Used for LXe calorimeter now

mckile
Download Presentation

BINP level meter for cold liquids

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. BINP level meter for cold liquids The method of measurements:change of the capacitance as function of a level of a liquid Originally designed for LN2 at the CMD-2 detector Used for LXe calorimeter now Sensor: wire 1.2mm at axis of tube 3mm (inner) capacitance 117 pf (0cm)  212 pf (160cm) 0.6 pf/cm Material: stainless steel and teflon  clean! Electronics: conversion C  fresonance  V

  2. The lever meter for the LXe calorimeter 1- LXe calorimeter 2- short sensor (145 mm) 3- long sensor (1600 mm) 4- teflon insulator (10 cm step) 5- 1.2 mm stainless steel wire 6- stainless steel tube with inner diameter 3 mm

  3. LN2 level meter at the CMD-2 10 cm

  4. LXe level meter at the calorimeter