Possible calibration methods for the final LXe calorimeter

Possible calibration methods for the final LXe calorimeter. A. Papa. 02/11/2004. 1. The motivations. (FWHM) implies:. An energy resolution of. a frequent and precise check of the calorimeter stability even during the normal data acquisition.

milica
Download Presentation

Possible calibration methods for the final LXe calorimeter

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Possible calibration methods for the final LXe calorimeter A. Papa 02/11/2004 1

  2. The motivations (FWHM) implies: An energy resolution of a frequent and precise check of the calorimeter stability even during the normal data acquisition Causes for gain instabilities: • Beam intensity variations • Variable background rates (photons and neutrons in the experimental hall) • Effects of the temperature T variation on the photocathode Q.E. and resistivity • Effects due to the capacitive coupling • Possible hysteresis phenomena as a function of T 2

  3. Precise calibrationrarely performed θ (degrees) • γ’s from decay (E(γ) ~ 54.9 MeV): use of a liquid hydrogen target recently tested with full success E (MeV) • Optional calorimeter calibration over range of γ energies: • γ’s from a tagged electron beam • (small magnet + MWPC’s) 3

  4. Frequent calibrations 1) α from Am source in detector (already used) 2) γ from Am/Be source out of the detector • E(γ) = 4.43 MeV • 60 % of Am decays 3) Thermal neutron capture Possible neutron sources: A) Am/Be (~ 10 KBq) B) Pulsed neutron generator Two neutron lines: at 4.5 or 14 MeV 0 12 MeV • Time separation of direct from • delayed reactions • Correlation: γ at 4.43 MeV and • n between 2-6 MeV 4

  5. Frequent calibration …again about pulsed neutron generator Commercially produced (Price ~ 10000$) Already used for the boron therapy, luggage screening etc. switchable on-off • D + 2H 3He + n Q = 3.27 MeV • D + 3H 4He + n Q = 17.59 MeV • Intensities from 106 n/s to 108 n/s • Typical pulse rate and pulse width 10 Hz and 1 μs • Time separation of direct from delayed reactions Moderator: ~ 10 cm of the polyethylene • 40% thermalized n • 10% n captured in moderator γ shield: ~ 3 cm of the tungsten or ~ 5 cm of the lead 5

  6. Caution in the use of n-source(n-activation) Results: 6 Neutron activation calculator:http://www.antenna.nl/wise/uranium/rnac.html

  7. Thermal neutron capture On Xe • Absorption length ~ 3 cm • Capture close to calorimeter walls • Multi γ,Σ E(γ)=9.3 MeV • Possible spill-out On Ni 9.0 8.534 8.122 7.698 9.0 MeV • Plate on calorimeter wall • Single γemission highly probable 52.7% • E(γ) =9.0 MeV • (used in Super Kamiokande) 52.7% 25.6% 4.65% 1.28% 7 0

  8. Neutrons in the Large Prototype recent measurement The neutron source was Am/Be (2 KBq) + diffused thermal neutron background in the experimental hall ( (?) note TN022 ) 4.43 MeV Without moderator γ energy spectrum 4.43 MeV γ energy spectrum n-edge n-edge and 9.3 MeV ADC With moderator (5 cm paraffin too thin! But space limitations) ADC 8

  9. Neutron calibration:other possibilities • Isotope activation in targets far from the detector with neutron • generator or intense neutron sources • No neutron on calorimeter (apart from hall background) • E(γ) = 6.13 MeV • Decay constant τ = 7.2 s or Possible reaction: Target: teflon disk 2) Nitrogen laser UV: emission line at ~ 300 nm; use of optical fibers and a small diffuser • Gain and relative QE measurements 175 nm is PMT independent? 14 nm (FWHM) 9

  10. γ γ NaI γ n LH2 Calibration from the calorimeter back Beam for calibration Normal beam Interesting possibility? π- μ+ • Liquid hydrogen target • permanently mounted close to • the Xenon calorimeter • π -/μ+ switching LXe Possibility of introducing also other particles (e-,e+ , π+,μ+ ) Locally same photocatode coverage as on the front face? 10

  11. Calibration and cryostat A choice must be made for the possible location of calibration ports before completing the cryostat final project 11

  12. Conclusion • Possible calibration methods were examined • Extremely important for calorimeter stability checks • Improvements studies depend of geometry, modera- • tors, sources, reactions, etc. under way 12