Next Generation of Dual Data Acquisition for High Sample Throughput of Dioxins April 2013 Heinz Mehlmann
Overview • Theoretical background of the Dual Data Option. • Improvement on the Hardware.
Background • Dual Data Acquisition is based on a double GC HRMS System
first peak (tcdf) last peak (ocdd) Background • The first and last peak of a chromatogram defines the time section of analytical intrest. section of analytical interest wasted time wasted time Chromatogram of a typical Dioxin run on a 60 meter column
Background • Sequence timescale in normal mode (no Dual Data) No benefit in terms of timesaving Dual GC-MS system Single GC-MS system sample 2 sample 3 sample 1 GC 1 GC 2
sample 1 sample 3 waiting Time waiting Time sample 2 waiting Time sample 1 sample 5 sample 3 GC 1 sample 6 sample 4 sample 2 GC 2 Background Sequence timescale in normal mode GC 1 GC 2 Sequence timescale with DUAL DATAAquisition.
Background • Continuous Dataflow
Background • The injection on GC 2 is performed during the aquisition of GC 1 and vice versa. GC1 GC2
Injector swiching valves to MS analytical column T-connector Background • Dynamic Flow Switching System • A staggert injection can only be realised with a hardware based modification in each GC.
The column-flow is directed into the MS. The column-flow is directed into the service vacuum and He compensates the flow into the ionsource. Background • Functioin of the Dynamic Flow Switching System
Background • Functioin of the Dynamic Flow Switching System
Background • Functioin of the Dynamic Flow Switching System
Background Time Timing GC 1 GC 2 GC 1 GC 1 GC 2
Dual Data 2nd Generation Focus regarding the Redesign of the Dual Data Option: • Robustness • Ease of use • Maintenance • Performance
Dual Data 2nd Generation Improvement on the Hardware • Optimization of all restrictions. • Purging the He line between regulator and switching valve. • Using a high quality 3-Way valve in the He makeup gas line. • Implementing a MCD wafer instead of T-pieces in the oven. • Reducing the column connections from four to two. • Using finger tight siltite ferrules® for column connection ® by SGE
He Injector Vacuum Transfereline T-Connector Analytical Column Dual Data 2nd Generation • First version of the Dual Data Hardware
Dual Data 2nd Generation • In case of micro leaks, Air can accumulate in the He Line when the line is closed for a longer time. • Closing the He line (with a leak) over night and reopen it again. He Vacuum N2 Dosing valve Injector O2 Switching valves Column Ar to MS
Dual Data 2nd Generation • Purging the He line during the Off position via a 3-Way switching valve doesn’t allow any air accumulation. • Closing the He line (with a leak) over night and then opened. He Vacuum N2 Dosing valve Injector 3-Way Purge Line O2 Switching valves Column Ar to MS
He Injector Vacuum Transfereline T-Connector Analytical Column Dual Data 2nd Generation • Redesign of the components in the GC oven
Dual Data 2nd Generation • All connections inside the oven are implemented in one single MCD (Multi Chanel Device)® Ø 25mm ® by SGE
Dual Data 2nd Generation • Both T-connections and the restriction in-between are inside the wafer. Vacuum He Column in Ø 25mm Column out
Dual Data 2nd Generation • Dead volume leads to tailing in the chromatography • No dead volume in the T-connection of the wafer • Reproducible conditions (restrictions) MCD Old Hardware
Dual Data 2nd Generation • Separation (CS3WT diluted): Chromatogram direct vs. via Wafer • Column direct into the MS • Column via the Wafer TCDD
Dual Data 2nd Generation • Only two column connection. He Vacuum Column in Column out MCD
Dual Data 2nd Generation • The ferrule is pre-mounted with a fingertite tool® that determines the distances between column and ferrule. • Columns with ID 0.25 mm and ID 0.32 mm are supported. ® by SGE
Dual Data 2nd Generation • The prepared column is then mounted to the MCD® again with a fingertite tool®. ® by SGE
Dual Data 2nd Generation • Data from Beta Test Site 5Cl-PCB
Dual Data 2nd Generation Customer feedback • 1.200 injections have been made (80 % real samples, 20 % standards and blanks) without any maintenance on the wafer. • There had been no problems neither with the wafers themselves nor with a leakage. • Dual Data with the wafers is now one of the best options of a DFS for a laboratory with a high throughput of similar samples.
Dual Data 2nd Generation • Customer Case Study http://pops.thermo-bremen.com
Dual Data 2nd Generation Dual Data Acquisition
Acknowledgement • KayteParlevliet and R&D team Australia • DetlefGrzelak Hamburg Germany