1 / 7

Measured baseline fluctuations and dark counts with Burle-Photonis MCP-PMT's

Measured baseline fluctuations and dark counts with Burle-Photonis MCP-PMT's. Jean-Francois Genat and Edward May Dec 2009 –Jan 2010. Experimental conditions. 10 and 25 m m 2” x 2” Burle-Photonis MCP tested - 25 m m MCP HV: 1.7-2.0 kV

moke
Download Presentation

Measured baseline fluctuations and dark counts with Burle-Photonis MCP-PMT's

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Measured baseline fluctuations and dark counts with Burle-Photonis MCP-PMT's Jean-Francois Genat and Edward May Dec 2009 –Jan 2010

  2. Experimental conditions 10 and 25 mm 2” x 2” Burle-Photonis MCP tested - 25 mm MCP HV: 1.7-2.0 kV Signals taken on one anode pad and fours neighbors grounded: - 10 mm MCP HV: 2.2-2.5 kV Signals taken on one anode pad and fours neighbors grounded: Use Ed’s calibrated test setup 408nm light set at 100 Photo-Electrons TDS 6154C 18GHz abw from Tek

  3. Signals 408nm laser 100 Photo-Electrons Conclusions: Gain is 40mV/100= 0.4mV/PE (25mm) at 2100 V 5mV/100= 50 mV/PE (10mm) at 2500V 10mm somewhat faster rise time, longer trailing edge, presumably due to 4 pads connected together. The rise time does NOT depend upon the amplitude

  4. Signals spectrum Fourier spectrum of a 2”x 2” MCP signal System noise 2 GHz Slope as steep as possible

  5. Baseline dark noise vs HV Conclusion: Baseline at 675 mV , no dependence with HV

  6. Impulse dark noise vs HV Conclusion: At full efficiency (25mm 2000V, 10mm 2400V), dark counts rates are: 25Hz (25mm) 20Hz (10mm)

  7. Conclusions MCP PMTs show signals, baseline fluctuations and dark counts quite similar to regular Photomultiplier tubes With: Faster signals (device is thinner, consequently better timing resolution) rise-time 250-500ps rise time compared to 500ps-1ns Less noise compared to “good” PM Tubes: dark counts 10-100 compared to 100-1000 Hz The rise time does NOT depend upon the amplitude