1 / 14

PMT response at high rates (and a final surprise)

PMT response at high rates (and a final surprise). A. M. Baldini INFN Pisa. photon charge and LED dependence on beam intensity. What we saw during the test:. UV. Beam on. off. on. BLUE. Beam on.

nickan
Download Presentation

PMT response at high rates (and a final surprise)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PMT response at high rates (and a final surprise) A. M. Baldini INFN Pisa

  2. photon charge and LED dependence on beam intensity What we saw during the test: UV Beam on off on BLUE Beam on

  3. One LED simulates crowding (50 KHz of 5 MeV pulses) due to neutrons and another one is read Neutrons rate during normal runs Ia 0.6 A, Ib 70 A   6 minutes Start crowd. Stop

  4. Cryogenics test facility at Pisa

  5. Method of the tests • PMT tested in Ar gas at several temperatures (30 -120°C) • Test LED (blue) pulse usually at 10-50 Hz, 300 mV, 20 ns wide • Crowding LED (blue) 300 mV, 10-40 KHz, 20 ns wide (Ia=1,10 A) • Crowding turned on and then off after some time • One old kind PMT (6041) and two new ones tested

  6. 30 10 50 Time (s) 60 20 120 Time (s) OLD (6041) PMT at –105 °C I crowding= 0.8 A  = 66 sec

  7. OLD (6041) PMT at 25 °C Same crowding as previous slide

  8. 200 100 300 Time (s) New (9288) PMT: TB0171 at several crow. Currents at T=-110°C TB0171 at 4,8, 12, 16 A Two effects present: cold + gain variation (non linearity)

  9. Gain variation as a function of photocathode illumination Taken from the photonis PMTs handbook

  10. 200 100 300 Time (s) TB0171 at room temperature (gain variation only) Icrowd=4A

  11. Effect vs temperature for the three tested PMTs TB0204 4A TB0171 4A TB0171 8A OLD 0.8A

  12.  vs temperature

  13. Crowding considerations in the LP + final detector • From the measured thermal neutron rate in the exp. Hall (TN022) one can estimate I 0.5 A in the LP and 5-6 times bigger in the final detector • If this current has to be lower than 1% of the base current (70  A) we have to screen the final detector from thermal neutrons • The corresponding current from low energy photons from the muon radiative decy turns out to be lower than for neutrons: I 04 A in the final detector

  14. Final surptise: last measurements... • Surprise !! after letting few days TB0171 in cold NO EFFECT is seen anymore in this PMT !!! • PMT stabilization? • Temperature offset? • Instrumental error(s) ? • ?? • Check on more PMTs

More Related