Agreement of Artificial Intelligence Topographic Diagnosis From Placido and Scheimpflug Systems. Leonardo N. Pimentel, MD 1,2 ; Isaac Ramos, MD 1,2 ; Bruno Valbon, MD 1,2 Ana Laura C. Canedo, MD 1,2 Renato Ambrósio Jr., MD, PhD 1,2. 1 – Instituto de Olhos Renato Ambrósio (Brazil)

norman
Download Presentation

Dr. Ambrósio is consultant for Oculus Optikgeräte GmbH (Wetzlar, Germany)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Agreement of Artificial Intelligence Topographic Diagnosis From Placido and Scheimpflug Systems Leonardo N. Pimentel, MD1,2; Isaac Ramos, MD1,2 ; Bruno Valbon, MD1,2 Ana Laura C. Canedo, MD1,2 Renato Ambrósio Jr., MD, PhD1,2 • 1 – Instituto de Olhos Renato Ambrósio (Brazil) • 2 – Rio de Janeiro Corneal Tomography and Biomechanics Study Group (Brazil) Dr. Ambrósio is consultant for Oculus Optikgeräte GmbH (Wetzlar, Germany)

  2. Purpose • To compare artificial intelligence in the axial curvature maps (sagittal), obtained by the topographic survey of Placido reflection system and those obtained using the rotating Scheimpflug system.

  3. Methods • Retrospective study. • 449 patients (898 eyes) with axial topography by the Nidek Corneal Navigator System using the Magellan Placido’s Corneal Topographer and axial topography by the Pentacam using the Scheimpflug system.

  4. Placido’s Corneal Topographer • The corneal topography based on the Placido disk technology is a method that analyzes the anterior curvature assisted by a computer through the recording, capture and processing of the reflected rings that are illuminated and projected onto the cornea. • Analysis of geometric shape of the reflected rings provides twenty two thousands measuring points that allow the construction of a topographic map.

  5. Scheimpflug Corneal Topographer • The Scheimpflug scanning provides considerably larger area measurement of the cornea compared with Placido’s topography systems. • It is also not sensitive from irregularities on the corneal reflex and interference from the tear film. The exam generates curvature maps from the front and back corneal.

  6. Methods • To compare the artificial diagnoses, the corresponding numbers were used, then the percentage of agreement was observed. • Was used Artificial intelligence by Nidek Corneal Navigator .

  7. Methods • Was used TKC system by Pentacam (Oculus). • The pacients that made hypermetropic Lasik were excluded from the Study, because TKC system didn`t have program for this cases.

  8. Results • Observed agreement between diagnoses Artificial by Maguellan (Nidek Cornea Navigator) and TKC from Pentacam (Oculus) in 525 eyes (58%) and disagreement in 373 eyes (48%). The agreement between diagnoses Artificial by maguellan and TKC system by Pentacam was better in normal and keratoconus cases, while was worse in suspect cases. 525 Eyes

  9. Conclusions • The axial topography by the Pentacam using Scheimpflug system is not sensitive from irregularities on the corneal reflex and interference from the tear film. • Especially in suspicious cases, is necessary additional data from tomography or biomechanics study to enhance the sensitivity and specificity to detect ectasia.