1 / 48

Electron Microscopes

Electron Microscopes. Scanning electron microscopes (SEM) Transmission electron microscopes (TEM) System components Theory of operation Sample Preparation Vendors. History of the electron microscope Types of electron microscopes SEM TEM Cryo Atmospheric Electron Microscope Images

pcarpenter
Download Presentation

Electron Microscopes

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Electron Microscopes Scanning electron microscopes (SEM) Transmission electron microscopes (TEM) System components Theory of operation Sample Preparation Vendors FNI 2C EM

  2. History of the electron microscope Types of electron microscopes SEM TEM Cryo Atmospheric Electron Microscope Images System components Theory of operation Electron beam specimen interaction Detectors Specimen preparation The electron gun The five axis stage The Electron Microscope FNI 2C EM

  3. Introductory SEM Text http://www.jeolusa.com/SERVICESUPPORT/ApplicationsResources/ElectronOptics/DocumentsDownloads/tabid/320/DMXModule/692/Command/Core_ViewDetails/Default.aspx?EntryId=257 FNI 2C EM

  4. Ernst Ruska • Invented the electron microscope in 1931. • The electron microscope obtains images by scanning the surface of a sample with a beam of electrons. http://nobelprize.org/physics/laureates/1986/ruska-autobio.html FNI 2C EM

  5. Older SEM control panel FNI 2C EM

  6. Modern SEM in operation FNI 2C EM

  7. My Ant Rosie with a Micromachined Part FNI 2C EM

  8. Microscopic Crystal Garden in Concrete Sample FNI 2C EM

  9. Computer Chip Cross Section FNI 2C EM

  10. Leaf FNI 2C EM

  11. Red Blood Cells FNI 2C EM

  12. CVTC SEM FNI 2C EM

  13. Scanning Electron Microscope Components Console Column Display Chamber Control Panel FNI 2C EM

  14. SEM Theory of Operation • An SEM is usually operated under high vacuum of 10-4 Pa or 10-6 Torr. • There is a large voltage difference between the anode and the electron gun cathode typically 10 kV to 40 kV. • When current begins to flow through the filament then electrons begin to be emitted from the filament through field emission. • The electrons are accelerated down the column by the electric potential. • Electrons pass through a series of apertures and magnetic lenses. • The final magnetic lens brings the electron beam into focus on the surface of the sample. FNI 2C EM

  15. Scanning Electron Microscope Components Electron Gun Condenser Lens Electron Beam Scan Coils Final Aperture Final Lens/ Focus Sample Secondary Electron Detector FNI 2C EM

  16. SEM Theory of Operation • The scan coils raster the beam back and forth over the surface of the sample. • Secondary electrons are drawn into the secondary electron detector and amplified. • The signal is translated into bright and dark areas on the monitor. • Magnification is determined by the amount of area scanned. • Magnification can range from 10x to 500,000x • Resolution is approximately 1 nm. FNI 2C EM

  17. Electron Beam Specimen Interactions ~3 nm FNI 2C EM

  18. Inside an SEM chamber Final lens Secondary electron detector Backscattered electron detector FNI 2C EM

  19. Inside CVTC’s SEM chamber Final lens Secondary electron detector Backscattered electron detector FNI 2C EM

  20. Sample Preparation FNI 2C EM

  21. Specimen Preparation Specimen Solid Non-solid Surface preparation Dryable Non-dryable Biological specimen preparation Low moisture High moisture Non-conductive Conductive Metal coating High vacuum High accelerating voltage High Vacuum Low accelerating voltage Low vacuum High vacuum Cryo-technique FNI 2C EM

  22. Sample Preparation • Samples will usually need to be prepared in some way for viewing in an SEM. • Often a coating of gold, palladium or carbon is applied to the sample if the sample is not conductive. • Sometimes a cross sectional sample is required. • To prepare a cross section sample it is first mounted in epoxy. • After the epoxy has hardened the sample is polished on a polishing wheel with different grades of sand paper and grit. • Finally a slurry of sub micron particles can be used to prepare the surface of the sample. • The sample may then be treated with different chemicals to highlight different features of the sample. • These “stains” include HF and Wright etch for semiconductors. • 4% nitric acid in methanol for stainless steel • After the sample is stained then they will be coated with a conductive layer. FNI 2C EM

  23. Colloidal Graphite or Silver Carbon tape Specimen Stub Sample mounting Sample FNI 2C EM

  24. Compression mount sample in plastic block Mounted sample preparation Sand sample on successively finer sand papers. Rotate 90° between grits and inspect under a microscope to ensure previous scratches are polished out. Polish with successively finer polishing slurry. (5 micron to .1 micron) until no scratches are visible under a microscope. Sample may need to be stained or etched with acid after this. FNI 2C EM

  25. Sputter coating Electrically insulating samples must be sputter coated with a layer of metal usually Palladium or Gold. This reactor uses argon excited into a plasma. The argon ions smash into the metal target knocking atoms off like billiard balls. The atoms deposit onto the sample to form a continuous layer. This allows electrons from the electron beam to flow to ground. FNI 2C EM

  26. Biological Sample Preparation Obtain the sample. This may involving toming or obtaining thins sections with a razor blade Fix the sample with gluteraldehyde This is causes cross linking among proteins similar to embalming Displace the water with successively more concentrated ethanol. This may take several hours. Displace the ethanol with liquid CO2 and dry in the critical point dryer to avoid drying with a meniscus. FNI 2C EM

  27. Critical Point Dryer – Caution this operates under high pressure! FNI 2C EM

  28. Electron Gun FNI 2C EM

  29. Thermionic Emission • In thermionic emission a cathode is heated to a high temperature, typically over 1000 K, by flowing electrical current through the filament. This reduces the work function for removing an electron. • There are two main types of thermionic cathodes. • Tungsten (W) • Lanthanum hexaboride (LaB6) FNI 2C EM

  30. Tungsten Cathode • Tungsten Cathode – A filament of tungsten wire is bent into a point. Current is passed through the wire causing it to heat to 2800 K. • The work function for removing an electron from a tungsten cathode is Ew = 4.5 eV. • Tungsten cathodes are inexpensive but they tend to burn out eventually like a light bulb as tungsten atoms evaporates from the surface of the filament. • Temperature: 2800 K. • Work function: Ew = 4.5 eV. • Current density: 5 x 104 A/cm2 • Electron source size: 20 μm • Lifetime: 50-100 hours • Vacuum: 10-4 Pa FNI 2C EM

  31. LaB6 Cathode • Temperature: 1800 K. • Work function: Ew = 2.4 eV. • Current density: 2-3 x 105 A/cm2 • Electron source size: 10 μm • Lifetime: 300-500 hours • Vacuum: 10-2 Pa FNI 2C EM

  32. Cold Cathode Field Emission Gun • A very sharp tip (<100 nm) • The electric field at the tip is > 107 V/cm. • The potential barrier becomes very narrow as well as reduced in height. • Electron tunnel through the barrier and leave the cathode. • Brightness: 107 A/cm2 • Temperature: Room temperature • Brightness: 2-3 x 105 A/cm2 • Electron source size: 5-10 nm • Lifetime: One year or more • Vacuum: 10-8 Pa • High resolution • Can image nonconductive materials at low accelerating voltage (1500 eV) FNI 2C EM

  33. Electron Acceleration Example Consider an electron emitted from a 30 kV electron gun. Calculate how fast it is going (compared to light speed). Calculate its ultimate resolution. Assume a tungsten cathode KE = Eγ-w h = 6.626 x 10-34 m2kg/s KE = ½ mv2 c = 2.998x108 m/s 1 eV = 1.602 x 10-19 J me = 9.11x10-31 kg FNI 2C EM

  34. Sample Stage • The sample is typically placed on a five axis stage in the vacuum chamber. The five dimensions in which the sample can be moved are: • X • Y • Z • Rotation • Tilt FNI 2C EM

  35. A Five Axis Stage Z X Y Tilt Rotation FNI 2C EM

  36. Venting the chamber FNI 2C EM

  37. Imaging FNI 2C EM

  38. Brightness and Contrast FNI 2C EM

  39. Brightness and Contrast FNI 2C EM

  40. Spot size 70 FNI 2C EM

  41. Spot size 18 FNI 2C EM

  42. Spot size 35 FNI 2C EM

  43. Stigmatism FNI 2C EM

  44. Stigmatism Stig X Stig Y FNI 2C EM

  45. Secondary vs. Backscattered FNI 2C EM

  46. Backscattered Electron Images Scan mode 2 (too fast) FNI 2C EM

  47. Backscattered Electron Images Topographical Mode Compositional Mode FNI 2C EM

  48. Burr FNI 2C EM