Modifications to MODIS Aqua ocean color bands calibration for 2010 OBPG reprocessing

Modifications to MODIS Aqua ocean color bands calibration for 2010 OBPG reprocessing. Gerhard Meister a,b a: Futuretech Corp. b: OBPG (Ocean Biology Processing Group) 1/25/10 MODIS Science Team Meeting 2010, Washington, D.C. MODIS Calibration Workshop 2010, Greenbelt, MD. 1. 1.

rashida
Download Presentation

Modifications to MODIS Aqua ocean color bands calibration for 2010 OBPG reprocessing

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Modifications to MODIS Aqua ocean color bands calibration for 2010 OBPG reprocessing Gerhard Meistera,b a: Futuretech Corp. b: OBPG (Ocean Biology Processing Group) 1/25/10 MODIS Science Team Meeting 2010, Washington, D.C. MODIS Calibration Workshop 2010, Greenbelt, MD 1 1

  2. Methodology: • Baseline: MCST lunar and SD trending (coll. 6) • New MCST lunar analysis: time dependent NIR RVS • MODIS Aqua crosscalibrated to SeaWiFS (as for Terra in Kwiatkowska et al., Applied Optics, 2008) • Approach: Use SeaWiFS L3 nLw, bring to TOA, adjust MODIS cal. and pol., for every month of the mission (4-day L3) • Verify with analysis using only MODIS Aqua data: temporal trends (seasonal cycle removed) and ratio of L2/L3 versus scan angle 2 2

  3. Temporal issues: 412nm and red bands 3 3

  4. RVS issues: 412nm, red bands 4 4

  5. New lunar trending of bands 13-16 (667-869nm) Comparison of collection 5 LUT to coll. 6: 5

  6. Temporal issue resolved: 678nm / FLH 6

  7. Reduced temporal variation for 488-547nm 7

  8. Increased trend in 412nm, angstrom improved: 8

  9. Crosscalibration approach: Lm = M11*Lt + m12*Q + m13*U Lm: measured TOA radiance (MODIS) Lt: true TOA radiance (from SeaWiFS) Q, U : linear Stokes vector components, modeled from Rayleigh and glint M11, m12, m13 : fitted instrument characterization parameters (depend on band, MS, detector, scan angle) 9

  10. Crosscalibration results: Polarization (temporal) - Larger seasonal cycle than in MODIS Terra - No trend in polarization coefficient m12 until 2008, not clear if trend afterwards Band 8: 412nm Black, blue, red:detectors 1,5,10, MS 1 View angle: nadir 10 10

  11. Crosscalibration results: Polarization (temporal) - Cycle in m12 decreases with wavelength - All bands stable over time Band 10: 488nm Band 13: 667nm 11 11

  12. Crosscalibration results: Polarization (temporal) - Variability in m13 similar as in Terra - All bands stable over time, prelaunch values used Band 8: 412nm Band 13: 667nm 12 12

  13. Crosscalibration results: Polarization (detectors) • Crosscalibration results confirm detector trend from prelaunch measurements (not used before) - Absolute offset at BOS (low TOA deg. of pol.) Band 8: 412nm Band 12: 547nm 13 13

  14. Crosscalibration results: Calibration (m1 and RVS) • Implementation for 2010 reprocessing: Temporal correction for 412-443nm, constant correction for 488-678nm 14 14

  15. RVS issues resolved: 412nm, 443nm 15 15

  16. RVS issues resolved: red bands (not EOM) 16 16

  17. Temporal issue resolved: 412nm 17

  18. Chlorophyll trends: similar to SeaWiFS 18

  19. Summary • Principal changes for MODIS Aqua calibration and characterization:- New temporal NIR scan angle dependence (MCST) - New temporal trends for 412-443nm (xcal)- New scan angle dependence for 488nm-678nm (xcal, no time dependence)- New detector dependence of polarization sensitivity (prelaunch) • Resulting improvements to ocean color products:- FLH stable over mission in olig.- Rrs 412nm stable over mission, variability reduced for remaining bands- Large scan dependence at EOM removed for 412nm- Minor scan angle dependence removed for 443-547nm- Large scan angle dependence reduced for 667-678nm, but still present at EOM 19

  20. Backup 20