1 / 12

Magnetic Property Measurement System MPMS XL

Magnetic Property Measurement System MPMS XL. Configurations MPMS XL is offered in 2 configurations MPMS XL-5 (5 Tesla) MPMS XL-7 (7 Tesla). Principal components of this measurement system. Temperature Control System ( 1.7 K to 400 K) Magnet Control System ( ± 5 T) Sample Handling System

Download Presentation

Magnetic Property Measurement System MPMS XL

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Magnetic Property Measurement SystemMPMS XL

  2. ConfigurationsMPMS XL is offered in 2 configurationsMPMS XL-5 (5 Tesla)MPMS XL-7 (7 Tesla)

  3. Principal components of this measurement system Temperature Control System ( 1.7 K to 400 K) Magnet Control System ( ± 5 T) Sample Handling System Computer Operating System Super conducting SQUID Amplifier System

  4. Sample rod • Sample rotator • Sample transport • Probe assembly • He level Sensor • Superconducting Solenoid • Flow Impedance • Squid capsule with magnetic shield • Superconducting pick-up coil

  5. 10. Dewar Isolation cabinet 11. Dewar 12. Printer 13. Magnetic power supply 14. Temp. controller 15. Console Cabinet 16. Power distribution unit 17.MPMS controller 18. Gas/Magnet control unit 19. Computer 20. Monitor

  6. Specifications of MPMS XL-5Field =5 TeslaTemperature=1.7K to 400 KSweeping rate=0.001K to 10K/Min

  7. Possible Measurements Magnetization vs. Temperature (M vs. T):Default settings: 300K-10K, Temperature interval 5KZero Field Cooling Field Cooling AC Susceptibility Magnetization vs. Field (M vs. H):Default settings: -5T to +5T step size 0.5T Hysteresis Hysteresis

  8. Features of MPMS MultiVu Software Simultaneous Access to all aspects of system operation: sequence setup,data analysis,data display Enables multiple documents and view to be visible and useable at the same time Simultaneous display of data files in multiple formats (e.g.. Graphs vs. texts) Automated Background Subtraction

  9. Limitations Measurement of hysteresis at more than two temperatures is considered as II sample Measurement below 10K also considered as II sample (because it is necessary for the instrument to be stable at 10K at least for 1 hour before going below 10K) Sample should be submitted only on Friday

  10. Total Number of Measurements Done • Total Number of Working days = • Total = • Physics = • SSCU = • MRC = • ISU = • Metallurgy = • Outside = • IPC =

More Related