1 / 31

WEAK MAGNETIC FIELD MEASUREMENT SYSTEM

WEAK MAGNETIC FIELD MEASUREMENT SYSTEM. Andrés G. Delannoy - 14949N SIST Program University of Puerto Rico at Mayagüez Fermi National Accelerator Laboratory August 5, 2008. A genda. Introduction Hall Probe Magnetic Measurement System Lock-In Amplifier LabVIEW Simulations

troylopez
Download Presentation

WEAK MAGNETIC FIELD MEASUREMENT SYSTEM

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. WEAK MAGNETIC FIELD MEASUREMENT SYSTEM Andrés G. Delannoy - 14949N SIST Program University of Puerto Rico at Mayagüez Fermi National Accelerator Laboratory August 5, 2008

  2. Agenda • Introduction • Hall Probe • Magnetic Measurement System • Lock-In Amplifier • LabVIEW Simulations • Lo-Pass & Curve Fitting Models • Results • Data Acquisition & Results • DC Probe Excitation • AC Probe Excitation • Probe Frequency Response • Summary & Conclusions • Future Work

  3. Introduction Hall Probe Magnetic Measurement System Lock-In Amplifier

  4. Introduction • HINS 8GeV linear accelerator R&D project • 7.2T @ 250A superconducting focusing magnets • Very low stray field (<0.10 G) requirement proximate to solenoid • Signal must be recovered from electromagnetic noise • Magnetic field measurement necessary at 300K and 4K

  5. Hall probe • Current through a Hall probe responds to external magnetic field by generating voltage. • The measured voltage is proportional to the magnetic field. V+ Copper winding (2 turns) i B Current Source V- I+ Reference Signal I-

  6. Magnetic measurement system • Setup: • Circular coil with two turns generating 0.201 G/A • Two probes in axis of coil • Cryomagnetics model HSP-T • Sensitivity > 1 mV/G • Nested magnetic shield • Probe excitation methods: • DC • AC • Measurement goals (0.01 G): • Probe Sensitivity (mV/G) • Noise Level (G)

  7. Coil’s Magnetic Field Calculation Biot and Savart Law: At 1A and 12.5cm coil circumference

  8. Lock-In Amplifier Signal Recovery Method

  9. Lock-In Amplifier • Recovers a periodic signal from a noisy environment • Mixes the noisy signal with a reference signal of the same frequency • Applies Low-Pass Filter to rectified signal

  10. Lock-In Amplifer

  11. LabView Simulations Lo-Pass & Curve Fitting Models Results

  12. LabView Simulation • Laboratory Virtual Instrumentation Engineering Workbench • Graphical programming language (named G) • Used for data acquisition, instrument control, and industrial automation • Based on Virtual Instruments (VI)

  13. Lock-In Simulation: Lo-Pass + Chopper Model 370 3000

  14. Lock-In Simulation: Curve-Fitting Model 25 300

  15. Lock-In Simulation:Results • Model approaches: • Chopper + Lo-Pass Model • Full Wave Rectifier + Lo-Pass Model • Fourier Transform Model • Curve-Fitting Model • The modeled filters recovered the simulated signal’s amplitude with 5-10% error (for 1:1 SNR). • A much higher level of certainty is required for our magnetic measurement system.

  16. Data Acquisition DC Probe Excitation AC Probe Excitation

  17. Data Acquisition:Power Supply & Preamp/Filter • SRS Module • SIM 911 BJT Preamp • SIM 965 Analog Filter • HP E3610A • Coil Current Source • Keithley 2400 • DC Probe Current Source

  18. Data Acquisition:Signal Recovery 7265 • DSP Lock-In Amplifier

  19. Data Acquisition: PXI • PCI eXtensions for Instrumentation (PXI) is a modular instrumentation platform used as a basis for building electronic test equipment or automation systems.

  20. DATA ACQUISITION: Data Recording • GPIB Connector & Analog Voltage Output readings were recorded using the NI interface. • Various LabView programs were written to record these data to file.

  21. DATA ACQUISITION: DC Probe Excitation- Sensitivity Probe 1 DC sensitivity ΔV/ΔB = 9.1289 V/G Probe Direct Current Source Low Noise Pre-Amp Coil Direct Current Source Hall Probe Low-Pass Filter Probe 2 DC sensitivity ΔV/ΔB = 6.4051 V/G LabVIEW Voltage Measurement

  22. DATA ACQUISITION:DC Probe Excitation - Noise Level • Probe 1 Noise Level • Probe 2 Noise Level

  23. DATA ACQUISITION:AC Probe Excitation - Sensitivity & Noise Level • Probe 1 Noise Level vs. Frequency • Probe 2 Noise Level vs. Frequency

  24. DATA ACQUISITION:AC Frequency Response - Probe 1 & 2 LabVIEW Voltage Measurement Lock-In Amplifier Lock-In Amplifier Input Probe Direct Current Source 10:1 Transformer Hall Probe

  25. DATA ACQUISITION:Transformer Noise Level vs. Frequency LabVIEW Voltage Measurement Lock-In Amplifier Lock-In Amplifier Input Probe Direct Current Source 10:1 Transformer Lock-In Voltage Without Transformer is Constant with Frequency

  26. Coil Current Source Noise Level Coil Direct Current Source LabVIEW Voltage Measurement 1.1W Shunt Resistor Coil Current Supply contributes 0.00006 G to the noise level

  27. Summary & Conclusions • Software based signal recovery results are inappropriate for the certainty required • Hall Probe sensitivity depends on excitation current (AC vs. DC). • Probe’s 1 & 2 sensitivity dissimilarity may be due to geometric or manufacturing inconsistencies • Lock-In amplifier offers lower noise figures, DC excitation must implement lo-pass filter. • Noise level may be further reduced by using higher precision equipment (power supplies, transformers, etc.)

  28. Future Work • Calculate the change in magnetic field as probe is moved off-center • Detailed noise level analysis at wider range of frequencies for AC probe excitation • Shielding method & effectiveness • Cryogenic temperature measurements

  29. Acknowledgements • Michael A. Tartaglia • Dmitri A. Sergatskov • Magnet Systems Department

  30. References • Griffiths, David J. Introduction to Electrodynamics. 3rd ed. New Jersey: John Wiley & Sons, 1999 p.215-218. • Reitz, John R., and Frederick J. Milford. Foundations of Electromagnetic Theory. London: Addison Wesley, 1960 p.149-159. • Jackson, John D. Classical Electrodynamics. 3rd ed. California: John Wiley & Sons, 1999 p.174-178. • R. Carcagno, et al. Superconducting Solenoid Magnet Test Results. ASC-06, Seattle, 2006. • I. Terechkine, et al. Focusing Solenoid for the Front End of a Linear RF Accelerator. PAQ-07, Albuquerque, 2007. • R. Carcagno, et al. Test Solenoids: Expected Performance and Test Results. TD-06-027, TD-06-028, TD-06-029, FNAL, July 2006.

  31. Questions The Answer to Life, the Universe, and Everything

More Related