Investigation of Bubble Formation in Tuohy-Borst Adaptors

www.angiodynamics.com. http://www.laboratoryglassapparatus .cm/parts_bubbletrap.gif. Investigation of Bubble Formation in Tuohy-Borst Adaptors. The Problem. Videotape Analysis. Melanie Bernard † , Isaac Clements†, and Jason Hirshburg† Advisor: Dr. Ted Larson, III, M.D.‡

saber
Download Presentation

Investigation of Bubble Formation in Tuohy-Borst Adaptors

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. www.angiodynamics.com http://www.laboratoryglassapparatus .cm/parts_bubbletrap.gif Investigation of Bubble Formation in Tuohy-Borst Adaptors The Problem Videotape Analysis Melanie Bernard†, Isaac Clements†, and Jason Hirshburg† Advisor: Dr. Ted Larson, III, M.D.‡ † Vanderbilt University Department of Biomedical Engineering ‡Vanderbilt University Medical Center Department of Radiology and Radiological Sciences • A Tuohy-Borst Adaptor, or a Y-adaptor, is a standard device used in catheterization procedures. • Saline can be injected into the side arm of the adaptor, while a microcatheter can be inserted through the adaptor and body of the catheter into the patient. • Observations: • There are many bubbles forming! • Out of 4 minutes of footage, we recorded 15 bubble occurrences. • Bubbles appear to be coming from within the catheter • In 14 out of 15 cases, the bubbles appear under set conditions: • Saline is flowing forward into the catheter • The 3-way valve is opened and flow is reversed • Bubbles appear immediately and seem to be coming from within the catheter. Research and Analysis Experiments Conducted • Primary Experiments: Test equipment for connections/leaks • Observed many bubbles throughout system, but no leaks The problem is that when saline is shut off and flow reverses, bubbles of unknown origin appear within the Tuohy-Borst Adaptor. • Observed device in a medical procedure on 11/18/03 • Literature Search • Innovation WorkBench (revised 3/2) • Conducted experiments • Primary: tested equipment • Secondary: simulated pressures and temperatures of entire system • Tertiary: testing individual components to different conditions • Filmed procedures on 2/2/04 and 3/31/04; obtained new supplies • Performed calculations based on actual and worst-case scenario parameters • We have researched the following catheter clearing methods: • Wetting agents • Dynamic Bubble Traps (DBTs) • Degassing and heating saline before injection • FEMLAB automated fluid flow analysis • Fuzzy inference system analysis with MATLAB • Designsafe Analysis (4/20/04) • Bubbles in the bloodstream as small as .5 μL can get stuck in arteries, blocking the flow of blood and oxygen to the brain. • Bubbles can cause strokes in 0.1% of all neurovascular procedures, which is about 120 procedures each year [1]. • Other Problems • Bubbles are hard to remove; they stick to the inner surface of the adaptor. • Bubbles can be created from a combination of sources. • Secondary Experiments: Simulated OR system w/ glycerin • Blood pressure = 120/80 (mmHg) • Saline pressure = 250 mmHg • Blood temp = 37º C • Saline temp = 25º C • Found some bubbles in system FEMLAB Analysis Our Task • Find the cause of bubble formation and correct it! • Constraints • Solution must apply to all adaptor types • Any system changes must not significantly change adaptor cost or the catheterization procedure. • Must eliminate problem without introducing new ones • Limitations • Time – 6 months • Money – have only a reasonable budget. • Equipment – don’t have access to blood or ultrasound machines Market Potential • Target group: operating rooms worldwide • Will meet customer needs: • Catheterization procedures will be safer. • Is technologically feasible • Physical principles and/or methods can be changed to eliminate bubble formation • Cost Analysis • $45,000 per procedure • $25,000 devoted to Radiology • 4-6 minutes spent clearing bubbles • 3 hour procedure • Approximately $700 spent per procedure dealing with bubbles • Tertiary Experiments: Test components for bubbles • -Tested components for leaks = none • 35 degree catheter turn causes leak • Air solubility in saline under pressure (250 mmHg) = 3.9 ml air/L • Tested catheter for bubbles after flushing = few • Tested bubble occurrence after syringe pull-back =many • Initial Observations • Bubbles… • Up to 50% of tube diameter (50μl) • Stick to inner surface of adaptor • Appear at the edge of the Y-junction • Occur before microcatheter is inserted Pressure profile of the Tuohy-Borst adaptor Velocity profile of adaptor with attached catheter during partial closure of rotating hemostatic valve. FEMLAB software was used to create realistic flow models of adaptor/catheter setup. Saline is assumed to enter the side arm of the adaptor at 250 mmHg, and the blood pressure is assumed to be fixed at 100 mmHg. These results revealed no local pressure drops due to adaptor geometry. Possible Bubble Sources 1. Air leaking through 3-way valve, rotating hemostatic valve, or into the junction between the adapter and catheter. 2. Air introduced during adjustments, i.e. removing and replacing adaptor and 3-way valve 3. Air captured in connections when the adaptor/3-way valve/catheter system is assembled 4. Air injected with contrast agent 5. Angle of the adaptor side arm causes local pressure drop which causes gas to come out of solution 6. The temperature difference between blood and saline causes:             a. N2 to come out of blood             b. O2 to come out of blood             c. CO2 to come out of blood 7. The temperature difference between blood and saline causes air to come out of saline 8. Initial flushing process leaves bubbles in catheter 9. Initial flushing process leaves bubbles in saline tube 10. Pulling back on syringe during flushing causes pressure drop and bubbles to come out of solution 11. The pressure cuff around the saline bag pushes air in saline bag into system 12. The pressure drop causes air to come out of saline 13. The pressure drop causes:             a. N2 to come out of blood             b. O2 to come out of blood             c. CO2 to come out of blood • Conclusions • The bubbles are being caused primarily by the following mechanisms: • Pressure of saline • In the beginning of the procedure, saline bag pressure is 250 mmHg. When the hemostatic valve is open to the atm., the infused saline experiences a relative drop of 275 mmHg (including height above patient). This can cause 3.9 ml of air/L saline to come out of solution. Within 5 min., the bag pressure can drop to 150 mmHg, which explains why there is a decrease in bubbles later on in the procedure. • Temperature of saline • As saline interfaces with blood, it experiences an increase in temp – which decreases solubility by 2.3 ml air/L saline. • Residual bubbles in saline tube • Even though the saline tube is flushed initially with 275 mmHg of saline, approximately 0.1 ml clings to the inner surface and may knock loose during the procedure. Fuzzy Inference Analysis Using a MATLAB analysis technique involving Fuzzy Inference Systems, we were able to quantitatively rank these sources from least harmful to most harmful based on subjective qualities such as probability of occurrence (Not Likely, Likely, and Most Likely) and volume of bubbles produced (Few, Moderate, Many). Solutions V1 = Likelihood V2 = Bubble Volume Output = Degree of Harm • Degas the saline before infusion • Agitate saline tube during flushing • Heat the saline to 37°C before infusion • Reduce saline bag pressure to 125 mmHg • 1.9 ml air/L saline • Implement a bubble trap Acknowledgements: Dr. Ted Larson & staff, Dr. Paul King, Dr. Joan Walker, Dr. Robert Roselli, Dr. Todd Giorgio, Dr. Cynthia Paschal Dr. Rick Hazelton, Dr. Bob Galloway, and Matt Lytle [1] http://www.ajnr.org/cgi/content/full/22/4/709

More Related