Hovermatt Module Presented by: JonnaBobeck BSN, RN, CEN
Introduction • Used to assist with lateral transfers and repositioning • Radiolucent and artifact free • Fewer staff required for transfer and decrease in injuries • Air perforations on the underside reduce friction
Indications • Patients unable to assist in their own lateral transfer • Patients whose weight or girth poses a potential health risk for the caregivers responsible for repositioning or laterally transferring said patients
Contraindications • Patients who are experiencing thoracic, cervical or lumbar fractures who are diagnosed as unstable, unless in conjunction with a spinal board on top of the Hovermatt mattress
Precautions • Caregivers must verify that bed is securely locked prior to transfer • Additional caregivers are recommended when moving patients over 750lbs/340kg • Do not attempt to move a patient on an un-inflated mattress • Route the power cord as to not pose a tripping hazard • Avoid blocking air intakes of the air supply unit • NEVER leave the patient unattended on an inflated mattress • ALWAYS use at least two people during patient transfer
MRI Use • When using the Hovermatt in the MRI environment, a 25 foot specialty hose is required!!
Instructions for use • Patient should be horizontal • Place HoverMatt underneath patient • Plug air supply into outlet • Insert hose • Turn on air Supply • Grasp handles and pull at an angle • Ensure the patient is centered • Press standby
Cleaning • Mattress: • Reusable constructed with nylon twill • Between use wipe with Virex • May be laundered for heavy soiling • Air Supply: • In between patient uses, the Air Supply can be cleaned by wiping down using a damp cloth with soap and water or mild neutral detergent. Dry using a clean , dry cloth or disposable paper towel • DO NOT spray cleaners or liquids directly on the Air Supply
Preventative Maintenance • Visually inspect • Make sure all handles are intact • Inspect for tears or holes • Remove from use if any damage is found
Infection Control • Seams are heat sealed • Offers a stain and fluid proof surface • May use sheet on top of mattress • Employ same protocols for isolation patient
Reference • Hovertech International. (2012). Hovermatt air transfer system user manual. Bethlehem, PA: Hovertech International. • Youtube. (Performer) (2011). Hovermatt demonstration [Theater]. Available from http://www.youtube.com/watch? v=5ASVQODzwQA