Basic Measurement of the Hamamatsu 10 inch PMT at –40 degree

Basic Measurement of the Hamamatsu 10 inch PMT at –40 degree Hiroko Miyamoto Dept. of Physics CHIBA University. The Uniformity measurement of the Hamamatsu 10 inch PMT for the IceCube Experiment.

selah
Download Presentation

Basic Measurement of the Hamamatsu 10 inch PMT at –40 degree

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Basic Measurement of the Hamamatsu 10 inch PMT at –40 degree Hiroko Miyamoto Dept. of Physics CHIBA University The Uniformity measurement of the Hamamatsu 10 inch PMTfor the IceCube Experiment Hiroko Miyamoto Dept. of PhysicsChiba University

  2. Our Work Chiba university group is working for both software and hardware part. • Software–develop the Java-based propagator of high energy cosmic neutrino in the earth and the ice. • Hardware–uniformity measurement of the PMT, gain measurement @-32℃, absolute QE, etc.

  3. PMT Measurements

  4. Signal〓QE×CE×Gain Gain Collection Efficiency (CE) Quantum Efficiency (QE) photoelectron photon

  5. UV LED (LIGHT SOURCE)

  6. PCI GPIB CAMAC Interface Function Generator CAMAC GPIB Interface Crate Controller 8ch ADC 50 nsec Sync. 200 nsec NIM 10 inch PMT Gate Generator UV LED Preamp TTL-NIM High Voltage Freezer 1750 V

  7. The Result of the Gain Measurement @ -32℃ See more results : http://www-ppl.s.chiba-u.jp/research/IceCube/pmt/ screening/FY2003/index.html

  8. 2D SCAN SYSTEM PMT UV LED R-guide Rotation-bed Motor

  9. 108Gain 75nsec

  10. The Measurement(of the first 14PMTs) The Light Source – UV (380 nm) Nichiya LED (220W driving) Collimated to less than 1mm diameter on the surface of PMT. Flashing at the every 1.3mm point on the R-guide. The Pulse is formed by the Agilent Func. Generator (100 nsec) 30photoelectrons/shots, corresponding the average of 3000 shots. [204 points/slice] PMT – Hamamatsu R7081 “ –02 tube. HV – 2000 V, 5E7~5E8 Gain Dark current ~ 404 mA @ 2000V (~363 mA @ 1800V) Noise Rate ~100 Hz ~2.5kHz with 50 mV threshold @ room temperature.

  11. Table 2 : Possible Error The total length on cathode : 26.6 cm

  12. Projection of length from on the X axis to the surface of PMT ( h → l ) PMT r θ R-guide rsinθ = h Step Number C ∝ h l l ∝ sin-1θ h X l = r sin-1(k/r・C)[cm]

  13. Improvements of measurement systems.

  14. Results • Calibration for data analysis • Effect of the geomagnetic field • Gain dependence • PMT by PMT difference

  15. Calibration 1: The fluctuation of LED 0 [deg] vs 360[deg] 0[deg] vs 180[deg]

  16. Calibration 2 : Reduction of the shape of crack of Magnetic Shield.

  17. Result 1: The effect of the geomagnetic field With B shield Without B shield

  18. Compare the average of the effect of B shield. Blue : with B shield Magenta : without B shield

  19. Summary 1 : the effect of geomagnetic field and B shield • The effect of the geomagnetic field can be seen at ~10% level. • B shield reduces the effect of geomagnetic field, ~5% still remains.

  20. Result 2: Gain dependence 2000V vs 1800V sf0050 azimuth 67.5[deg] sf0050 azimuth 123.75[deg]

  21. Averaged Difference 2000V vs 1800V

  22. Summary 2: Gain dependence • The gain difference of the efficiency is at 5% level, which is not significant except around the edge.

  23. Result 3: PMT by PMT difference SF0030 vs SF0050 SF0016 vs SF0050

  24. Averaged Difference sf0050 vs sf0030 and sf0050 vs sf0016

  25. Averaged Difference sf0004 vs flat and sf0050 vs flat

  26. Summary 3 : PMT by PMT difference • The difference of the collection efficiency from a PMT to PMT shows the variance by 10% at maximum in the average, but, point to point difference reach 20% level. • A single PMT cannot represent all PMTs.

  27. Discussion: 2 dimensional view of the cathode surface

  28. Executive Summary • Completed the 2D survey data for IceCube detector MC implementation. • Geomagnetic filed effects at ~5% level. • Gain vs Collection efficiency is not significant, less than ~5% . • PMT by PMT difference is remarkable which reach ~20% level.

  29. Outlook • Absolute Quantum Efficiency : PMT/DOM • Gain Scanning • Wave Form data taking • Angular Response • DOM simulation

  30. Preliminary

  31. Absolute QE measurement

  32. Monitor PMTs Laser Baffles Calibrated PMT Future plan • Now, we start developing new scattering box. • Easy to change PMTs  Useful to calibrate many PMTs • To exchange gas, we will use vacuum pump  We can easily control the gas quality.

  33. Signal〓QE×CE×Gain Gain Collection Efficiency (CE) Quantum Efficiency (QE) photoelectron photon

  34. The End