Immucor Galileo Echo

Immucor Galileo Echo. ATTENTION SHIPPER: PLEASE READ!! PLEASE READ!! Important Unpacking Information Included. Immucor Galileo Echo. If you have questions about the information contained in this document, please contact Immucor Technical Support at 800.492.2583.

snowy
Download Presentation

Immucor Galileo Echo

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Immucor Galileo Echo ATTENTION SHIPPER: PLEASE READ!! PLEASE READ!! Important Unpacking Information Included. Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  2. Immucor Galileo Echo If you have questions about the information contained in this document, please contact Immucor Technical Support at 800.492.2583.  If you think you may have run into an issue that would prevent you from completing the delivery.... DO NOT LEAVE THE DELIVERY SITE call Matt Gaines at TECH TRANS at 1-800-952-3411-IMMEDIATELY Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  3. Immucor Galileo Echo Tech Trans System Installation Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  4. Immucor Galileo Echo Tools required • Flathead screwdriver, 5/32” Allen wrench, and a box cutter Or For Instrument Serial numbers M00252-M00291 and M00309 and above. • #2 Phillips screwdriver and a box cutter Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  5. Immucor Galileo Echo There are two (2) cartons that are labeled Main Unit and Fluidics • Tech Trans is responsible for unpacking the “Main Unit” and “Fluidics” cartons only. • There are additional cartons for the computer, monitor, printer, UPS, and accessories. These cartons will be unpacked by an Immucor Representative. These boxes should NOT be unpacked by Tech Trans personnel. Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  6. Immucor Galileo Echo Unpacking the Main Unit carton • Open the large box and remove the four foam pads. • Open the inner small box. Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  7. Immucor Galileo Echo Unpacking the Main Unit carton • Remove the cardboard insert and the plastic bag Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  8. Immucor Galileo Echo Unpacking the Main Unit carton • Grasp the instrument by the handles near the rear of the instrument • Grasp the back panel of the instrument for stability while lifting • Lift the instrument out of the box Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  9. Immucor Galileo Echo Unpacking the Main Unit carton • Place the instrument on the counter with the bottom plywood exposed • Using the screwdriver, unscrew the seven screws and remove the plywood from the bottom of the instrument Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  10. Immucor Galileo Echo Unpacking the Main Unit carton • Place the instrument upright on the right side of the counter Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  11. Immucor Galileo Echo Unpacking the Main Unit carton • Place the plywood, plastic bag and cardboard insert back into the box for removal • Place the foam padding on the four corners of the inner box for removal Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  12. Immucor Galileo Echo Unpacking the Fluidics carton • The Fluidics carton consists of the instrument cover for the Main unit and the Fluidics module • Open the large box and remove the four foam pads. • Open the inner box and remove the cardboard divider. The instrument cover is located on top. Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  13. Immucor Galileo Echo Unpacking the Fluidics carton • Remove the instrument cover from the box and remove the bubble wrap from the cover Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  14. Immucor Galileo Echo Unpacking the Fluidics carton • Place the instrument cover on the Main unit by sliding it onto the front of the instrument. Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  15. Immucor Galileo Echo Unpacking the Fluidics carton • Remove the cardboard divider. The Fluidics Module is in the bottom of the carton. • Lift the cardboard insert out of the box. Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  16. Immucor Galileo Echo Unpacking the Fluidics carton • Lift the Fluidics module out of the box. • Place on the counter with the bottom plywood exposed. Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  17. Immucor Galileo Echo Unpacking the Fluidics carton • Using the 5/32” Allen wrench or the Phillips screwdriver for fluidics modules F00252-F00291 and F00309 and above, remove the five screws and remove the plywood • Place the Fluidics module upright to the left of the main instrument Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  18. Immucor Galileo Echo Unpacking the Fluidics carton • Return the plywood, cardboard insert, cardboard dividers and bubble wrap inside the box for removal • Replace the foam padding onto the corners of the inner box for removal Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  19. Immucor Galileo Echo • Remove the empty Main unit and Fluidics cartons from the facility • The G3-Access and all additional cartons will be unpacked by Immucor • If requested by the customer to unpack all additional accessory boxes, call Matt Gaines at TECH TRANS at 1-800-952-3411-IMMEDIATELY Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT

  20. Immucor Galileo Echo Minimum space requirements for instrument placement: • Width – 43 1/8 inches • Height – 26 inches • Depth – 22 1/2 inches • 12 inches of permanent clearance in back of instrument for service and maintenance access • Verify level surface for instrument placement. Instrument must be placed on a stationary bench, not on an equipment cart. • Additional space required for printer: 14.6" W x 9.6" D x 9.5” H Technical Bulletin TB-08-005-06 Immucor, Inc. Department of Technical Services and Field Quality, Norcross, GA DRAFT